Fabrication of a MEMS retinal tack and long-term in vivo biocompatibility evaluation for retinal prostheses

被引:0
|
作者
Sangmin Lee
Hyoungho Ko
机构
[1] Kyung Hee University,Department of Biomedical Engineering
[2] Chungnam National University,Department of Electronics Engineering
来源
Microsystem Technologies | 2017年 / 23卷
关键词
PMMA; Barb; Retinitis Pigmentosa; Ganglion Cell Layer; Microelectrode Array;
D O I
暂无
中图分类号
学科分类号
摘要
A silicon and PMMA (poly-(methyl methacrylate)) retinal tack with barbs is developed for retinal prostheses. The MEMS retinal tack presented in this paper has barbs, which are intended for the fixation of the epiretinal-type retina-stimulating microelectrode array. Although conventional retinal tacks made of titanium are available, they are expensive, invasive, and restricted in shape. The silicon retinal tack is fabricated using silicon micromachining, and the PMMA retinal tack is fabricated by X-ray lithography for accurate micro-patterning, which is very scalable and is suitable for inexpensive and repeatable mass production. In this paper, long-term in vivo biocompatibility evaluation of silicon and PMMA retinal tacks is presented. The MEMS retina tack is implanted with epiretinal microelectrodes onto the retinal surface of the rabbit eye and harvested after 1 year for the silicon retinal tack and 6 months for the PMMA retinal tack. Through histological examination, no inflammatory infiltrates in retina, choroid, nor sclera are revealed. Scanning electron microscope (SEM) images show no chemical reaction trace nor mechanical fracture of the implanted device. The results indicate that both silicon and PMMA retinal tacks are suitable for mechanical fixation of microelectrode arrays. Also, the long-term evaluation of the durability in vivo shows acceptable biocompatibility for further use in retinal prostheses.
引用
收藏
页码:483 / 489
页数:6
相关论文
共 50 条
  • [1] Fabrication of a MEMS retinal tack and long-term in vivo biocompatibility evaluation for retinal prostheses
    Lee, Sangmin
    Ko, Hyoungho
    MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2017, 23 (02): : 483 - 489
  • [2] Long-term In vivo Biocompatibility Evaluation of MEMS Retinal Tack for Retinal Prosthesis
    Lee, S.
    Koo, K. -, I
    Ban, J. -W.
    Jeong, H. -Y
    Park, H.
    Seo, J. -M.
    Chung, H.
    Cho, D. -I.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 11: BIOMEDICAL ENGINEERING FOR AUDIOLOGY, OPHTHALMOLOGY, EMERGENCY AND DENTAL MEDICINE, 2009, 25 (11): : 194 - +
  • [3] A Long-Term in vivo Biocompatibility Evaluation of PMMA Retinal Tack for Retinal Prosthesis
    Lee, S.
    Koo, K. -I.
    Ban, J-W.
    Jeong, H. -Y.
    Park, H.
    Hong, S. -J.
    Bae, S. H.
    Seo, J. -M.
    Chung, H.
    Cho, D. D.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [4] LONG-TERM HISTOLOGIC AND ELECTROPHYSIOLOGIC EVALUATION OF THE ALLOY RETINAL TACK
    OHIRA, A
    DEJUAN, E
    TSAI, MS
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1991, 229 (01) : 95 - 98
  • [5] DEVELOPMENT OF A SILICON RETINAL IMPLANT - LONG-TERM BIOCOMPATIBILITY OF MATERIALS
    MILLER, S
    RIZZO, JF
    WYATT, JL
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1995, 36 (04) : S746 - S746
  • [6] Long-term biocompatibility of an epi-retinal electrode array implant
    D'Anna, SA
    Humayun, MS
    Weiland, JD
    Suzuki, S
    Majji, AB
    de Juan, E
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S732 - S732
  • [7] In vivo evaluations of retinal prostheses
    Salzmann, J
    Linderholm, P
    Paques, M
    Simonutti, M
    Chiappore, JA
    Safran, AB
    Renaud, P
    Sahel, J
    Picaud, S
    2004 IEEE INTERNATIONAL JOINT CONFERENCE ON NEURAL NETWORKS, VOLS 1-4, PROCEEDINGS, 2004, : 573 - 573
  • [8] A suprachoroidal electrical retinal stimulator design for long-term animal experiments and in vivo assessment of its feasibility and biocompatibility in rabbits
    Zhou, J. A.
    Woo, S. J.
    Park, S. I.
    Kim, E. T.
    Seo, J. M.
    Chung, H.
    Kim, S. J.
    JOURNAL OF BIOMEDICINE AND BIOTECHNOLOGY, 2008,
  • [9] Long-term juxtafoveal retinal telangiectasia
    Watzke, Robert C.
    Klein, Michael L.
    Folk, James C.
    Farmer, Samuel G.
    Munsen, Richard S.
    Champfer, Robert J.
    Sletten, K. R.
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2005, 25 (06): : 727 - 735
  • [10] LONG-TERM EVALUATION OF PATELLAR PROSTHESES
    LEVITT, RL
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1973, (97) : 153 - 157