A preliminary test for the application of the k0 standardization method of neutron activation analysis at the Radiochemistry and Spectroscopy Laboratory of the Istituto Nazionale di Ricerca Metrologica

被引:0
|
作者
M. Di Luzio
L. Bergamaschi
M. Oddone
M. Prata
A. Salvini
G. D’Agostino
机构
[1] c/o Department of Chemistry – University of Pavia,Istituto Nazionale di Ricerca Metrologica (INRIM), Unit of Radiochemistry and Spectroscopy
[2] Department of Chemistry – University of Pavia,Laboratorio di Energia Nucleare Applicata (LENA)
[3] University of Pavia,undefined
关键词
-method; TRIGA Mark II; Instrumental neutron activation analysis;
D O I
暂无
中图分类号
学科分类号
摘要
The application of the k0-standardization method of neutron activation analysis offers the great advantage to perform multi-elemental analyses using a single element as a comparator. On the other hand, some knowledge of the neutron flux energy distribution is mandatory. Recently, neutron flux parameters f and α in three irradiation channels of the TRIGA Mark II nuclear reactor in Pavia were measured for the first time to allow the application of the k0-standardization method at the Radiochemistry and Spectroscopy Laboratory of the Istituto Nazionale di Ricerca Metrologica. In this framework, a preliminary test was performed by analyzing a set of samples irradiated in Central Channel, including a number of elements prepared with mono-elemental solutions and a reference material. The results of all the selected elements were in agreement with the expected value and confirmed the applicability of the k0-standardization method in Central Channel.
引用
收藏
页码:723 / 729
页数:6
相关论文
共 49 条
  • [1] A preliminary test for the application of the k0 standardization method of neutron activation analysis at the Radiochemistry and Spectroscopy Laboratory of the Istituto Nazionale di Ricerca Metrologica
    Di Luzio, M.
    Bergamaschi, L.
    Oddone, M.
    Prata, M.
    Salvini, A.
    D'Agostino, G.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2018, 315 (03) : 723 - 729
  • [2] Application of the k0 method in neutron activation analysis and in prompt gamma activation analysis
    Revay, Zs
    Kennedy, G.
    RADIOCHIMICA ACTA, 2012, 100 (8-9) : 687 - 698
  • [3] Validation of k0 Standardization Method in Neutron Activation Analysis Using Kayzero Program for Windows
    Kubesova, Marie
    Kucera, Jan
    CHEMICKE LISTY, 2011, 105 (04): : 261 - 268
  • [4] Elemental analysis of herbal preparations for traditional medicines by neutron activation analysis with the k0 standardization method
    S. B. Sarmani
    I. Abugassa
    A. Hamzah
    M. D. Yahya
    Biological Trace Element Research, 1999, 71-72 : 365 - 376
  • [5] Elemental analysis of herbal preparations for traditional medicines by neutron activation analysis with the k0 standardization method
    Sarmani, SB
    Abugassa, I
    Hamzah, A
    Yahya, MD
    BIOLOGICAL TRACE ELEMENT RESEARCH, 1999, 71-2 (1) : 365 - 376
  • [7] Recommended nuclear data for use in the k0 standardization of neutron activation analysis
    De Corte, F
    Simonits, AS
    ATOMIC DATA AND NUCLEAR DATA TABLES, 2003, 85 (01) : 47 - 67
  • [8] APPLICATION OF THE K(0) STANDARDIZATION METHOD IN NEUTRON-ACTIVATION ANALYSIS OF CERAMIC MATERIALS
    DARDENNE, C
    DESCHAMPS, N
    REVEL, G
    RIGLET, G
    HERVIOU, C
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1993, 168 (02): : 417 - 427
  • [9] Comparison of Kayzero for Windows and k0-IAEA software packages for k0 standardization in neutron activation analysis
    Kubesova, Marie
    Kucera, Jan
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 654 (01): : 206 - 212
  • [10] PANORAMIC INSTRUMENTAL NEUTRON-ACTIVATION ANALYSIS OF SYNTHETIC AND NATURAL SILICA - A PERFORMANCE REVIEW OF THE K0 STANDARDIZATION METHOD
    DECORTE, F
    DEWISPELAERE, A
    VANDENBOER, M
    BOSSUS, D
    VANSLUIJS, R
    ANALYTICA CHIMICA ACTA, 1991, 254 (1-2) : 127 - 143