Neural Firing Mechanism Underlying Two-Electrode Discrimination by 3D Transcutaneous Electrical Nerve Stimulation Computational Model

被引:2
|
作者
Ye S. [1 ]
Zhu K. [1 ]
Li P. [1 ]
Sui X. [1 ]
机构
[1] School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai
关键词
A; artificial tactile sensation; computational model; R; 318.18; transcutaneous electrical nerve stimulation (TENS); two-electrode discrimination distance (TEDD);
D O I
10.1007/s12204-019-2134-y
中图分类号
学科分类号
摘要
Transcutaneous electrical nerve stimulation (TENS) has been widely used for sensory feedback which is a key consideration of improving the performance of prosthetic hands. Two-electrode discriminability is the key to realize high-spatial-resolution TENS, but the neural firing mechanism is not clear yet. The goal of this research is to investigate the neural firing patterns under two-electrode stimulation and to reveal the potential mechanisms. A three-dimensional (3D) model is established by incorporating Aβ fiber neuron clusters into a layered forearm structure. The diameters of the stimulating electrodes are selected as 5, 7, 9 and 12 mm, and the two-electrode discrimination distance (TEDD) is quantified. It is found that a distant TEDD is obtained for a relatively large electrode size, and 7mm is suggested to be the optimal diameter of stimulating electrodes. The present study reveals the neural firing patterns under two-electrode stimulation by the 3D TENS model. In orderto discriminate individual electrodes under simultaneous stimulation, no crosstalk of activated Aβ fibers exists between two electrodes. This research can further guide the optimization of the electrode-array floorplan. © 2019, Shanghai Jiao Tong University and Springer-Verlag GmbH Germany, part of Springer Nature.
引用
收藏
页码:716 / 722
页数:6
相关论文
共 48 条
  • [1] Neural Firing Mechanism Underlying Two-Electrode Discrimination by 3D Transcutaneous Electrical Nerve Stimulation Computational Model
    叶栓
    祝凯华
    李鹏
    隋晓红
    Journal of Shanghai Jiaotong University(Science), 2019, 24 (06) : 716 - 722
  • [2] A 3D Computational Model of Transcutaneous Electrical Nerve Stimulation for Estimating Aβ Tactile Nerve Fiber Excitability
    Zhu, Kaihua
    Li, Liming
    Wei, Xuyong
    Sui, Xiaohong
    FRONTIERS IN NEUROSCIENCE, 2017, 11
  • [3] Novel Two-electrode Excitation Mode for 3D Electrical Capacitance Tomography
    Wang, Yutao
    Li, Ya
    Zhao, Yaping
    PROCEEDINGS OF THE 2019 31ST CHINESE CONTROL AND DECISION CONFERENCE (CCDC 2019), 2019, : 3206 - 3210
  • [4] Thermal Model of Spiked Electrode in Transcutaneous Electrical Nerve Stimulation (TENS)
    Chen, Luyao
    Lo, Yi-Kai
    Wang, Yushan
    Liu, Wentai
    2017 8TH INTERNATIONAL IEEE/EMBS CONFERENCE ON NEURAL ENGINEERING (NER), 2017, : 219 - 222
  • [5] 3D splint prototype system for applications in muscular rehab by transcutaneous electrical nerve stimulation (TENS)
    Saldana-Martinez, M. I.
    Guzman-Gonzalez, J. V.
    Barajas-Gonzalez, O. G.
    Guzman-Ramos, V.
    Garcia-Garza, A. K.
    Gonzalez-Garcia, R. B.
    Garcia-Ramirez, M. A.
    DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES X, 2017, 10056
  • [6] Numerical modeling of percutaneous auricular vagus nerve stimulation: a realistic 3D model to evaluate sensitivity of neural activation to electrode position
    Samoudi, Amine M.
    Kampusch, Stefan
    Tanghe, Emmeric
    Szeles, Jozsef C.
    Martens, Luc
    Kaniusas, Eugenijus
    Joseph, Wout
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2017, 55 (10) : 1763 - 1772
  • [7] Numerical modeling of percutaneous auricular vagus nerve stimulation: a realistic 3D model to evaluate sensitivity of neural activation to electrode position
    Amine M. Samoudi
    Stefan Kampusch
    Emmeric Tanghe
    Jozsef C. Széles
    Luc Martens
    Eugenijus Kaniusas
    Wout Joseph
    Medical & Biological Engineering & Computing, 2017, 55 : 1763 - 1772
  • [8] Effects of Electrode Position on Spatiotemporal Auditory Nerve Fiber Responses: A 3D Computational Model Study
    Kang, Soojin
    Chwodhury, Tanmoy
    Moon, Il Joon
    Hong, Sung Hwa
    Yang, Hyejin
    Won, Jong Ho
    Woo, Jihwan
    COMPUTATIONAL AND MATHEMATICAL METHODS IN MEDICINE, 2015, 2015
  • [9] A foldable neural electrode for 3D stimulation of deep brain cavities
    Kil, Dries
    De Vloo, Philippe
    Nuttin, Bart
    Puers, Robert
    PROCEEDINGS OF THE 30TH ANNIVERSARY EUROSENSORS CONFERENCE - EUROSENSORS 2016, 2016, 168 : 137 - 142
  • [10] The Effects of Electrical Stimulation on a 3D Osteoblast Cell Model
    Mahadeo, Crystal O.
    Shahin-Shamsabadi, Alireza
    Khodamoradi, Maedeh
    Fahnestock, Margaret
    Selvaganapathy, Ponnambalam Ravi
    CELLS, 2025, 14 (06)