Routine breath-hold gradient echo MRI-derived right ventricular mass, volumes and function: accuracy, reproducibility and coherence study

被引:0
|
作者
Farzin Beygui
Alain Furber
Stéphane Delépine
Gérard Helft
Jean-Philippe Metzger
Philippe Geslin
Jean Jacques Le Jeune
机构
[1] University Hospital of Angers,Nuclear Medicine and Biophysics Laboratory
[2] La Pitié-Salpêtrière University Hospital,Department of Cardiology, Cardiology Institute
[3] University Hospital of Angers,Department of Cardiology
关键词
ejection fraction; magnetic resonance imaging; right ventricle; ventricular volumes;
D O I
暂无
中图分类号
学科分类号
摘要
Right ventricular (RV) dysfunction is a predictor of poor outcome in patients with heart disease. Conventional imaging modalities fail to assess RV volumes accurately. We sought to assess the accuracy and reproducibility of routine breath-hold gradient echo magnetic resonance imaging (MRI)-derived RV mass, volumes and function. We assessed (1) The accuracy of in vivo MRI-derived RV mass in comparison to the RV weight in 9 minipigs. (2) Intra- and inter-observer reproducibility of RV mass, end-diastolic (EDV) and end-systolic (ESV) volumes and ejection fraction (EF) in 15 normal volunteers and 10 patients with heart disease. (3) Inter-study reproducibility of the former parameters in 25 coronary artery disease patients. (4) The correlation between right and left ventricular stroke volumes in the total population. Strong statistically significant correlations were found between: (1) MRI-derived RV mass and RV weight (r = 0.98, bias = 2.5 g), (2) Intra-observer measurements of RV mass (r = 0.96, bias = 0.5 g), EDV (r = 0.99, bias = −1.5 ml), ESV (r = 0.98, bias = 0.1 ml) and EF (r = 0.92, bias = −1.4%), (3) Inter-observer measurements of RV mass (r = 0.95, bias = 1.1 g), EDV (r = 0.98, bias = −1.1 ml), ESV (r = 0.98, bias = 1.2 ml) and EF (r = 0.87, bias = −1.9%), (4) Inter-study measurements of RV mass (r = 0.91, bias = −0.1 g), EDV (r = 0.96, bias = 3.8 ml), ESV (r = 0.98, bias = 0.3 ml) and EF (r = 0.90, bias = 0.9%), (5) MRI-derived right and left ventricular stroke volumes (r = 0.87). The assessment of the RV mass, volumes and function by routine breath-hold gradient echo MRI is accurate and highly reproducible. The correlation between left and RV MRI-derived stroke volumes indicates excellent coherence of simultaneous bi-ventricular volume measurements.
引用
收藏
页码:509 / 516
页数:7
相关论文
共 7 条
  • [1] Routine breath-hold gradient echo MRI-derived right ventricular mass, volumes and function:: accuracy, reproducibility and coherence study
    Beygui, F
    Furber, A
    Delépine, S
    Helft, G
    Metzger, JP
    Geslin, P
    Le Jeune, JJ
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2004, 20 (06): : 509 - 516
  • [2] REPRODUCIBILITY OF MRI-DERIVED MEASUREMENTS OF RIGHT-VENTRICULAR VOLUMES AND MYOCARDIAL MASS
    PATTYNAMA, PMT
    LAMB, HJ
    VANDERVELDE, EA
    VANDERGEEST, RJ
    VANDERWALL, EE
    DEROOS, A
    MAGNETIC RESONANCE IMAGING, 1995, 13 (01) : 53 - 63
  • [3] Accuracy of right- and left-ventricular heart volume and left-ventricular muscle mass determinations by cine MRI in breath-hold technique.
    Rominger, MB
    Bachmann, GF
    Geuer, M
    Puzik, M
    Boedeker, RH
    Ricken, WW
    Rau, WS
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1999, 170 (01): : 54 - 60
  • [4] Reproducibility study of left ventricular measurements with breath-hold cine MRI using a semiautomated volumetric image analysis program
    Butler, SP
    McKay, E
    Paszkowski, AL
    Quinn, RJ
    Shnier, RC
    Donovan, JT
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (02): : 467 - 472
  • [5] Accuracy of single breath-hold cine MRI analyzed by guide-point modeling for the assessment of Left Ventricular Function
    Christina Heilmaier
    Kai Nassenstein
    Sonia Nielles-Vallespin
    Sven Zuehlsdorff
    Peter Hunold
    Joerg Barkhausen
    Journal of Cardiovascular Magnetic Resonance, 11 (Suppl 1)
  • [6] Reproducibility of left ventricular mass measurement using a half-fourier black-blood single-shot fast spin-echo sequence within a single breath hold: Comparison with a conventional multiple breath-hold gradient echo technique segmented in patients
    Vignaux, O
    Duboc, D
    Coste, J
    Argaud, C
    Carlier, P
    Le Roux, P
    Weber, S
    Legmann, P
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 15 (06) : 654 - 660
  • [7] Rapid evaluation of right and left ventricular function and mass using real-time True-FISP Cine MR imaging without breath-hold: Comparison with segmented True-FISP Cine MR imaging with breath-hold
    Hori, Y
    Yamada, N
    Higashi, M
    Hirai, N
    Nakatani, S
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2003, 5 (03) : 439 - 450