Design of a Real-time Self-adjusting Calibration Algorithm to Improve the Accuracy of Continuous Blood Glucose Monitoring

被引:0
|
作者
Ziru Jia
Lijuan Huang
Hongying Liu
Yonghong Huang
Wang Li
Xitian Pi
Xiaolin Zheng
机构
[1] Chongqing University,Key Laboratory for National Defense Science and Technology of Innovative Micro
[2] Chongqing Engineering Research Center of Medical Electronics,nano Devices and System Technology
[3] Chongqing University,Key Laboratory of Biorheological Science and Technology, Ministry of Education, College of Bioengineering
来源
关键词
Glucose sensor; Dynamic blood glucose monitoring; Calibration algorithm; Signal drift; Sensitivity attenuation;
D O I
暂无
中图分类号
学科分类号
摘要
The aim of this study is to establish a real-time self-adjusting calibration algorithm to compensate for signal drift and sensitivity attenuation of subcutaneous implantable glucose sensors. A real-time self-adjusting in vivo calibration method was designed based on the one-point calibration model. The current signal was compensated in real-time and the sensitivity was calibrated regularly. The least squares method was used to fit the initial parameters of the model, and then, the in vivo monitored current data was calibrated. Comparing with the mean absolute relative difference (MARD) of the blood glucose concentration by the traditional one-point calibration model (22.85 ± 5.76%), the MARD of the blood glucose concentration calibrated by the real-time self-adjusting in vivo calibration method was 6.28 ± 2.31%. The accuracy of the dynamic blood glucose monitoring was effectively improved. This calibration algorithm could compensate the signal drift in real time and correct sensitivity regularly to improve the accuracy of dynamic glucose monitoring, thus significantly enhancing diabetic self-management.
引用
收藏
页码:1163 / 1176
页数:13
相关论文
共 50 条
  • [1] Design of a Real-time Self-adjusting Calibration Algorithm to Improve the Accuracy of Continuous Blood Glucose Monitoring
    Jia, Ziru
    Huang, Lijuan
    Liu, Hongying
    Huang, Yonghong
    Li, Wang
    Pi, Xitian
    Zheng, Xiaolin
    [J]. APPLIED BIOCHEMISTRY AND BIOTECHNOLOGY, 2020, 190 (04) : 1163 - 1176
  • [2] Accuracy of a new real-time continuous glucose monitoring algorithm
    Navas de Solis, M. S.
    [J]. AVANCES EN DIABETOLOGIA, 2010, 26 (02): : 129 - 130
  • [3] Accuracy Evaluation of a New Real-Time Continuous Glucose Monitoring Algorithm in Hypoglycemia
    Mahmoudi, Zeinab
    Jensen, Morten Hasselstrom
    Johansen, Mette Dencker
    Christensen, Toke Folke
    Tarnow, Lise
    Christiansen, Jens Sandahl
    Hejlesen, Ole
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2014, 16 (10) : 667 - 678
  • [4] Real-Time Improvement of Continuous Glucose Monitoring Accuracy
    Facchinetti, Andrea
    Sparacino, Giovanni
    Guerra, Stefania
    Luijf, Yoeri M.
    DeVries, J. Hans
    Mader, Julia K.
    Ellmerer, Martin
    Benesch, Carsten
    Heinemann, Lutz
    Bruttomesso, Daniela
    Avogaro, Angelo
    Cobelli, Claudio
    [J]. DIABETES CARE, 2013, 36 (04) : 793 - 800
  • [5] A real-time self-adjusting scheduling algorithm for tandem precedence structure in cluster computing
    Alhamdan, A
    Ammar, R
    El Dessouky, A
    [J]. Proceedings of the 46th IEEE International Midwest Symposium on Circuits & Systems, Vols 1-3, 2003, : 1608 - 1611
  • [6] Postprandial instead of preprandial calibration allows better accuracy of real-time continuous glucose monitoring (CGM)
    Capel, I.
    Pons, B.
    Hernando, M. E.
    Garcia, G.
    Brugues, E.
    de Leiva, A.
    Gomez, E.
    Prados, A.
    Erdozain, L.
    Rigla, M.
    [J]. DIABETOLOGIA, 2007, 50 : S419 - S419
  • [7] Accuracy of real-time continuous glucose monitoring in the MiniMed paradigm system
    Mastrototaro, John
    Soundararajan, Gopikrishnan
    Cooper, Ken
    Shah, Rajiv
    [J]. DIABETES, 2007, 56 : A112 - A112
  • [8] Accuracy of real-time glucose values using continuous glucose monitoring:: the Guardian™ II continuous glucose monitoring system
    Schwartz, SL
    Buchea, O
    Shin, J
    Hoss, U
    Shah, R
    [J]. DIABETOLOGIA, 2004, 47 : A315 - A315
  • [9] Design and implementation of real-time continuous glucose monitoring instrument
    实时动态血糖监测器的设计与实现
    [J]. Liu, Hongying (liu_hongying@163.com), 2017, West China Hospital, Sichuan Institute of Biomedical Engineering (34):
  • [10] Real-time continuous glucose monitoring in the ICU
    J Gios
    B Manuel-y-Keenoy
    P Rogiers
    [J]. Critical Care, 19 (Suppl 1):