Measuring Systems for Thermometer Calibration in Low-Temperature Range

被引:0
|
作者
A. Szmyrka-Grzebyk
L. Lipiński
H. Manuszkiewicz
A. Kowal
A. Grykałowska
D. Jancewicz
机构
[1] Institute of Low Temperature and Structure Research,
来源
关键词
Calibration; Temperature scale; Thermometry;
D O I
暂无
中图分类号
学科分类号
摘要
The national temperature standard for the low-temperature range between 13.8033 K and 273.16 K has been established in Poland at the Institute of Low Temperature and Structure Research (INTiBS). The standard consists of sealed cells for realization of six fixed points of the International Temperature Scale of 1990 (ITS-90) in the low-temperature range, an adiabatic cryostat and Isotech water and mercury triple-point baths, capsule standard resistance thermometers (CSPRT), and AC and DC bridges with standard resistors for thermometers resistance measurements. INTiBS calibrates CSPRTs at the low-temperature fixed points with uncertainties less than 1 mK. In lower temperature range—between 2.5 K and about 25 K — rhodium–iron (RhFe) resistance thermometers are calibrated by comparison with a standard which participated in the EURAMET.T-K1.1 comparison. INTiBS offers a calibration service for industrial platinum resistance thermometers and for digital thermometers between 77 K and 273 K. These types of thermometers may be calibrated at INTiBS also in a higher temperature range up to 550°C. The Laboratory of Temperature Standard at INTiBS acquired an accreditation from the Polish Centre for Accreditation. A management system according to EN ISO/IEC 17025:2005 was established at the Laboratory and presented on EURAMET QSM Forum.
引用
收藏
页码:2466 / 2476
页数:10
相关论文
共 50 条
  • [1] Measuring Systems for Thermometer Calibration in Low-Temperature Range
    Szmyrka-Grzebyk, A.
    Lipinski, L.
    Manuszkiewicz, H.
    Kowal, A.
    Grykalowska, A.
    Jancewicz, D.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2011, 32 (11-12) : 2466 - 2476
  • [2] DEVELOPMENT OF RADIATION THERMOMETER FOR LOW-TEMPERATURE RANGE
    MANABE, C
    AKAMATSU, M
    YANAI, S
    SUZUKI, N
    TERAI, K
    FUJIMOTO, S
    TETSU TO HAGANE-JOURNAL OF THE IRON AND STEEL INSTITUTE OF JAPAN, 1993, 79 (07): : 759 - 764
  • [3] LOW-TEMPERATURE THERMOMETER
    不详
    NATURE, 1958, 182 (4638) : 770 - 770
  • [4] Processing calibration data of low-temperature thermometer based on clustering algorithm
    Liao, Yi
    Zhang, Yu
    Zha, Kuifan
    Liu, Xuming
    Pan, Changzhao
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2024, 95 (08):
  • [5] LOW-TEMPERATURE INDIUM RESISTANCE THERMOMETER
    LOGVINENKO, SP
    MIKHINA, GF
    MENDE, GT
    CRYOGENICS, 1984, 24 (08) : 421 - 422
  • [6] LOW-TEMPERATURE THERMAL NOISE THERMOMETER
    PATRONIS, ET
    MARSHAK, H
    REYNOLDS, CA
    SAILOR, VL
    SHORE, FJ
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1959, 30 (07): : 578 - 580
  • [7] LOW-TEMPERATURE SEMICONDUCTOR RESISTANCE THERMOMETER
    LOGVINENKO, SP
    ROSSOSHANSKII, OA
    OPRISHCHENKO, LA
    MEASUREMENT TECHNIQUES USSR, 1988, 31 (11): : 1110 - 1112
  • [8] TIN, A CANDIDATE FOR LOW-TEMPERATURE NMR THERMOMETER
    AHOLA, H
    EHNHOLM, GJ
    ISLANDER, ST
    OSTMAN, P
    RANTALA, B
    CRYOGENICS, 1980, 20 (05) : 277 - 282
  • [9] ACOUSTICAL THERMOMETER ADVANCES LOW-TEMPERATURE MEASUREMENT
    不详
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1964, 36 (12): : 2420 - &
  • [10] CARBONACEOUS MICROCOMPOSITE LOW-TEMPERATURE RESISTANCE THERMOMETER
    DULNEV, GN
    MESHKOVSKII, IK
    NOVIKOV, AF
    SHAVKUNOVA, VA
    CHERNOPOLSKII, AD
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1983, 26 (01) : 230 - 232