Evaluation of Cadmium-Zinc-Telluride Detector-based Single-Photon Emission Computed Tomography for Nuclear Cardiology: a Comparison with Conventional Anger Single-Photon Emission Computed Tomography

被引:15
|
作者
Niimi T. [1 ]
Nanasato M. [2 ]
Sugimoto M. [1 ]
Maeda H. [3 ]
机构
[1] Department of Radiological Technology, Nagoya Daini Red Cross Hospital, Takanaga Niimi, 2-9 Myouken-cho, Showa-ku, Nagoya
[2] Cardiovascular Center, Nagoya Daini Red Cross Hospital, 2-9 Myouken-cho, Showa-ku, Nagoya
[3] Department of Radiological Technology, Nagoya University School of Health Sciences, 1-1-20 Daiko-minami, Higashi-ku, Nagoya
关键词
Anger SPECT; Collimation system; CZT camera; D-SPECT; Nuclear cardiology;
D O I
10.1007/s13139-017-0474-9
中图分类号
学科分类号
摘要
Purpose: The differences in performance between the cadmium-zinc-telluride (CZT) camera or collimation systems and conventional Anger single-photon emission computed tomography (A-SPECT) remain insufficient from the viewpoint of the user. We evaluated the performance of the D-SPECT (Spectrum Dynamics, Israel) system to provide more information to the cardiologist or radiological technologist about its use in the clinical field. Materials and Methods: This study evaluated the performance of the D-SPECT system in terms of energy resolution, detector sensitivity, spatial resolution, modulation transfer function (MTF), and collimator resolution in comparison with that of A-SPECT (Bright-View, Philips, Japan). Energy resolution and detector sensitivity were measured for Tc-99m, I-123, and Tl-201. The SPECT images produced by both systems were evaluated visually using the anthropomorphic torso phantom. Results: The energy resolution of D-SPECT with Tc-99m and I-123 was approximately two times higher than that of A-SPECT. The detector sensitivity of D-SPECT was higher than that of A-SPECT (Tc-99m: 4.2 times, I-123: 2.2 times, and Tl-201: 5.9 times). The mean spatial resolution of D-SPECT was two times higher than that of A-SPECT. The MTF of D-SPECT was superior to that of the A-SPECT system for all frequencies. The collimator resolution of D-SPECT was lower than that of A-SPECT; however, the D-SPECT images clearly indicated better spatial resolution than the A-SPECT images. Conclusion: The energy resolution, detector sensitivity, spatial resolution, and MTF of D-SPECT were superior to those of A-SPECT. Although the collimator resolution was lower than that of A-SPECT, the D-SPECT images were clearly of better quality. © 2017, Korean Society of Nuclear Medicine.
引用
收藏
页码:331 / 337
页数:6
相关论文
共 50 条
  • [1] Myocardial perfusion imaging using single-photon emission computed tomography with cadmium-zinc-telluride technology
    Konsek-Komorowska, Sonia J.
    Peczkowska, Mariola
    Cwikla, Jaroslaw B.
    NUCLEAR MEDICINE REVIEW, 2022, 25 (02) : 119 - 126
  • [2] Is Single-Photon Emission Computed Tomography/Computed Tomography Superior to Single-Photon Emission Computed Tomography in Assessing Unilateral Condylar Hyperplasia?
    Liu, Pingan
    Shi, Jun
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2019, 77 (06) : 1279.e1 - 1279.e7
  • [3] Single-photon emission computed tomography/computed tomography in endocrinology
    Krausz, Yodphat
    Israel, Ora
    SEMINARS IN NUCLEAR MEDICINE, 2006, 36 (04) : 267 - 274
  • [4] SINGLE-PHOTON TRACERS FOR SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY
    KUNG, HF
    ACADEMIC RADIOLOGY, 1995, 2 : S94 - S95
  • [5] Myocardial blood flow evaluation with dynamic cadmium-zinc-telluride single-photon emission computed tomography: Bright and dark sides
    D'Antonio, Adriana
    Assante, Roberta
    Zampella, Emilia
    Mannarino, Teresa
    Buongiorno, Pietro
    Cuocolo, Alberto
    Acampa, Wanda
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2023, 104 (7-8) : 323 - 329
  • [6] SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY
    KNOLL, GF
    PROCEEDINGS OF THE IEEE, 1983, 71 (03) : 320 - 329
  • [7] Single-photon emission computed tomography in neurotherapeutics
    Devous Sr. M.D.
    NeuroRX, 2005, 2 (2): : 237 - 249
  • [8] SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY
    UREN, RF
    MAGISTRETTI, PL
    ROYAL, HD
    PARKER, JA
    FRONT, D
    HILL, TC
    HOLMAN, BL
    JONES, AG
    KOLODNY, GM
    MEDICAL JOURNAL OF AUSTRALIA, 1983, 1 (09) : 411 - 413
  • [9] Single-photon emission computed tomography/computed tomography in brain tumors
    Schillaci, Orazio
    Filippi, Luca
    Manni, Carlo
    Santoni, Riccardo
    SEMINARS IN NUCLEAR MEDICINE, 2007, 37 (01) : 34 - 47
  • [10] Single-photon emission computed tomography/computed tomography in abdominal diseases
    Schillaci, Orazio
    Filippi, Luca
    Danieli, Roberta
    Simonetti, Giovanni
    SEMINARS IN NUCLEAR MEDICINE, 2007, 37 (01) : 48 - 61