Sample concentration and impedance detection on a microfluidic polymer chip

被引:0
|
作者
Poorya Sabounchi
Alfredo M. Morales
Pierre Ponce
Luke P. Lee
Blake A. Simmons
Rafael V. Davalos
机构
[1] Sandia National Laboratories,Energy Systems Department
[2] University of California,Berkeley Sensor and Actuator Center, Department of Bioengineering
[3] Sandia National Laboratories,Materials Chemistry Department
[4] Virginia Tech-Wake Forest University,School of Biomedical Engineering and Sciences
来源
Biomedical Microdevices | 2008年 / 10卷
关键词
Dielectrophoresis; Insulator-based dielectrophoresis; Biochip; Sample enrichment; Pathogen monitoring; BioMEMS;
D O I
暂无
中图分类号
学科分类号
摘要
We present an on-chip microfluidic sample concentrator and detection triggering system for microparticles based on a combination of insulator-based dielectrophoresis (iDEP) and electrical impedance measurement. This platform operates by first using iDEP to selectively concentrate microparticles of interest based on their electrical and physiological characteristics in a primary fluidic channel; the concentrated microparticles are then directed into a side channel configured for particle detection using electrical impedance measurements with embedded electrodes. This is the first study showing iDEP concentration with subsequent sample diversion down an analysis channel and is the first to demonstrate iDEP in the presence of pressure driven flow. Experimental results demonstrating the capabilities of this platform were obtained using polystyrene microspheres and Bacillus subtilis spores. The feasibility of selective iDEP trapping and impedance detection of these particles was demonstrated. The system is intended for use as a front-end unit that can be easily paired with multiple biodetection/bioidentification systems. This platform is envisioned to act as a decision-making component to determine if confirmatory downstream identification assays are required. Without a front end component that triggers downstream analysis only when necessary, bio-identification systems (based on current analytical technologies such as PCR and immunoassays) may incur prohibitively high costs to operate due to continuous consumption of expensive reagents.
引用
收藏
相关论文
共 50 条
  • [1] Sample concentration and impedance detection on a microfluidic polymer chip
    Sabounchi, Poorya
    Morales, Alfredo M.
    Ponce, Pierre
    Lee, Luke P.
    Simmons, Blake A.
    Davalos, Rafael V.
    BIOMEDICAL MICRODEVICES, 2008, 10 (05) : 661 - 670
  • [2] Microfluidic evaporator for on-chip sample concentration
    Casadevall i Solvas, Xavier
    Turek, Vladimir
    Prodromakis, Themistoklis
    Edel, Joshua B.
    LAB ON A CHIP, 2012, 12 (20) : 4049 - 4054
  • [3] Microfluidic distillation chip for methanol concentration detection
    Wang, Yao-Nan
    Liu, Chan-Chiung
    Yang, Ruey-Jen
    Ju, Wei-Jhong
    Fu, Lung-Ming
    ANALYTICA CHIMICA ACTA, 2016, 912 : 97 - 104
  • [4] Polymer microfluidic chip with embedded CNTs for DNA sample preparation
    Khuntontong, P.
    Chan, J. X.
    Gong, M.
    Wang, Z. P.
    NANOTECHNOLOGY 2012, VOL 2: ELECTRONICS, DEVICES, FABRICATION, MEMS, FLUIDICS AND COMPUTATIONAL, 2012, : 369 - 371
  • [5] A PDMS microfluidic chip with nanostructures for bacteria concentration and fast detection
    Yu, J. J.
    Xiao, L. D.
    Yang, M.
    MEMS 2008: 21ST IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 2008, : 272 - 275
  • [6] Rapid detection of Azotobacter chroococcum concentration based on microfluidic chip
    Guo, Jianjiang
    Zhang, Rongbiao
    Yang, Ning
    Xu, Peifeng
    Nongye Jixie Xuebao/Transactions of the Chinese Society for Agricultural Machinery, 2015, 46 (01): : 155 - 159
  • [7] A compact microfluidic chip with integrated impedance biosensor for protein preconcentration and detection
    Tuan Vu Quoc
    Wu, Meng-Syuan
    Tung Thanh Bui
    Trinh Chu Duc
    Jen, Chun-Ping
    BIOMICROFLUIDICS, 2017, 11 (05):
  • [8] A Radiate Microstructure MALDI Chip for Sample Concentration and Detection
    Chen, Shun-Yuan
    Li, Kun-In
    Yu, Chih-Sheng
    Wang, Jun-Sheng
    Hu, Yi-Chiuen
    Lai, Chien-Chen
    ANALYTICAL CHEMISTRY, 2010, 82 (14) : 5951 - 5957
  • [9] Hemoglobin Detection on a Microfluidic Sensor Chip with a Partially Conjugated Polymer
    Eo, Soo-Han
    Won, Kwang-Jae
    Song, Simon
    Yoon, Bora
    Kim, Jong-Man
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2010, 31 (02) : 467 - 469
  • [10] An Enclosed Paper Microfluidic Chip as a Sample Preconcentrator Based on Ion Concentration Polarization
    Liu, Ning
    Dinh-Tuan Phan
    Lew, Wen Siang
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2017, 11 (06) : 1392 - 1399