Complication rates and reduction potential of palmar versus dorsal locking plate osteosynthesis for the treatment of distal radius fractures

被引:58
|
作者
Wichlas F. [1 ]
Haas N.P. [1 ]
Disch A. [1 ]
Machó D. [1 ]
Tsitsilonis S. [1 ,2 ]
机构
[1] Center for Musculoskeletal Surgery, Charité,Universitätsmedizin Berlin, Augustenburger Platz 1, Berlin
[2] Berlin-Brandenburg Center of Regenerative Therapies, Charité,Universitätsmedizin Berlin, Berlin
关键词
Approach; Complication; Distal radius fracture; Locking plate;
D O I
10.1007/s10195-014-0306-y
中图分类号
学科分类号
摘要
Background: The aim of this study was to evaluate the complication rates of volar versus dorsal locking plates and postoperative reduction potential after distal radius fractures.; Materials and methods: For this study 285 distal radius fractures (280 patients/59.4 % female) treated with locked plating were retrospectively evaluated. The mean age of the patients was 54.6 years (SD 17.4) and the mean follow-up was 33.2 months (SD 17.2). The palmar approach was used in 225 cases and the dorsal approach in 60 cases (95 % type C fractures).; Results: Adequate reduction was achieved with both approaches, regardless of fracture severity. In the dorsal group, the complications and implant removal rates were significantly higher and the operative time was also longer.; Conclusions: Based on these facts, we advocate the palmar locking plate for the vast majority of fractures. In cases of complex multifragmentary articular fractures where no compromise in reduction is acceptable, and with the biomechanical equality of palmar and dorsal plating remaining unproven, dorsal plating may still be considered.; Level of evidence: Therapeutic level IV. © 2014, The Author(s).
引用
收藏
页码:259 / 264
页数:5
相关论文
共 50 条
  • [1] Minimally invasive plate osteosynthesis for distal radius fractures with a palmar locking plate
    N. Takada
    T. Otsuka
    K. Yamada
    H. Suzuki
    T. Hasuo
    A. Kondo
    M. Fukuta
    European Journal of Trauma and Emergency Surgery, 2012, 38 : 627 - 632
  • [2] Minimally invasive plate osteosynthesis for distal radius fractures with a palmar locking plate
    Takada, N.
    Otsuka, T.
    Yamada, K.
    Suzuki, H.
    Hasuo, T.
    Kondo, A.
    Fukuta, M.
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2012, 38 (06) : 627 - 632
  • [3] Palmar locking plate for treatment of unstable dorsal dislocated distal radius fractures
    R. Arora
    M. Lutz
    D. Fritz
    R. Zimmermann
    J. Oberladstätter
    M. Gabl
    Archives of Orthopaedic and Trauma Surgery, 2005, 125 : 399 - 404
  • [4] Palmar locking plate for treatment of unstable dorsal dislocated distal radius fractures
    Arora, R
    Lutz, M
    Fritz, D
    Zimmermann, R
    Oberladstätter, J
    Gabl, M
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2005, 125 (06) : 399 - 404
  • [5] Combined dorsal and palmar plate osteosynthesis for intraarticular distal radius fractures
    Can, Ue
    Lattmann, T.
    Crook, D.
    Trentz, O.
    Platz, A.
    UNFALLCHIRURG, 2008, 111 (08): : 607 - 612
  • [6] Limits of palmar locking-plate osteosynthesis of unstable distal radius fractures
    Arora, R.
    Lutz, M.
    Zimmermann, R.
    Krappinger, D.
    Gabl, M.
    Pechlaner, S.
    HANDCHIRURGIE MIKROCHIRURGIE PLASTISCHE CHIRURGIE, 2007, 39 (01) : 34 - 41
  • [7] Indications for locking plate osteosynthesis in the treatment of distal radius fractures
    Woltmann, A.
    Hierholzer, C.
    Buehren, V
    TRAUMA UND BERUFSKRANKHEIT, 2007, 9 (01) : 6 - 13
  • [8] Locking Versus Nonlocking Palmar Plate Fixation of Distal Radius Fractures
    Osti, Michael
    Mittler, Christoph
    Zinnecker, Richard
    Westreicher, Christoph
    Allhoff, Clemens
    Benedetto, Karl Peter
    ORTHOPEDICS, 2012, 35 (11) : E1613 - E1617
  • [9] Comparison between fixation with dorsal T plate and palmar locking plate in the treatment of unstable displaced distal radius fractures
    Nalbantoglu, Ufuk
    Gereli, Arel
    Ucar, Yavuz
    Turkmen, Metin
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2008, 42 (05) : 365 - 372
  • [10] Distal radius fractures: Minimally invasive plate osteosynthesis with dorsal bicolumnar locking plates fixation