Image quality in low-dose coronary computed tomography angiography with a new high-definition CT scanner

被引:0
|
作者
Egle Kazakauskaite
Lars Husmann
Julia Stehli
Tobias Fuchs
Michael Fiechter
Bernd Klaeser
Jelena R. Ghadri
Catherine Gebhard
Oliver Gaemperli
Philipp A. Kaufmann
机构
[1] University Hospital Zurich,Cardiac Imaging
[2] University of Zurich,Zurich Center of Integrative Human Physiology (ZIHP)
关键词
High-definition computed tomography; Coronary angiography; Image quality; Radiation dose;
D O I
暂无
中图分类号
学科分类号
摘要
A new generation of high definition computed tomography (HDCT) 64-slice devices complemented by a new iterative image reconstruction algorithm—adaptive statistical iterative reconstruction, offer substantially higher resolution compared to standard definition CT (SDCT) scanners. As high resolution confers higher noise we have compared image quality and radiation dose of coronary computed tomography angiography (CCTA) from HDCT versus SDCT. Consecutive patients (n = 93) underwent HDCT, and were compared to 93 patients who had previously undergone CCTA with SDCT matched for heart rate (HR), HR variability and body mass index (BMI). Tube voltage and current were adapted to the patient’s BMI, using identical protocols in both groups. The image quality of all CCTA scans was evaluated by two independent readers in all coronary segments using a 4-point scale (1, excellent image quality; 2, blurring of the vessel wall; 3, image with artefacts but evaluative; 4, non-evaluative). Effective radiation dose was calculated from DLP multiplied by a conversion factor (0.014 mSv/mGy × cm). The mean image quality score from HDCT versus SDCT was comparable (2.02 ± 0.68 vs. 2.00 ± 0.76). Mean effective radiation dose did not significantly differ between HDCT (1.7 ± 0.6 mSv, range 1.0–3.7 mSv) and SDCT (1.9 ± 0.8 mSv, range 0.8–5.5 mSv; P = n.s.). HDCT scanners allow low-dose 64-slice CCTA scanning with higher resolution than SDCT but maintained image quality and equally low radiation dose. Whether this will translate into higher accuracy of HDCT for CAD detection remains to be evaluated.
引用
收藏
页码:471 / 477
页数:6
相关论文
共 50 条
  • [1] Image quality in low-dose coronary computed tomography angiography with a new high-definition CT scanner
    Kazakauskaite, Egle
    Husmann, Lars
    Stehli, Julia
    Fuchs, Tobias
    Fiechter, Michael
    Klaeser, Bernd
    Ghadri, Jelena R.
    Gebhard, Catherine
    Gaemperli, Oliver
    Kaufmann, Philipp A.
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 (02): : 471 - 477
  • [2] Image quality of low-dose CCTA in obese patients: impact of high-definition computed tomography and adaptive statistical iterative reconstruction
    Cathérine Gebhard
    Tobias A. Fuchs
    Michael Fiechter
    Julia Stehli
    Barbara E. Stähli
    Oliver Gaemperli
    Philipp A. Kaufmann
    [J]. The International Journal of Cardiovascular Imaging, 2013, 29 : 1565 - 1574
  • [3] Image quality of low-dose CCTA in obese patients: impact of high-definition computed tomography and adaptive statistical iterative reconstruction
    Gebhard, Catherine
    Fuchs, Tobias A.
    Fiechter, Michael
    Stehli, Julia
    Staehli, Barbara E.
    Gaemperli, Oliver
    Kaufmann, Philipp A.
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 (07): : 1565 - 1574
  • [4] Improving image quality and resolution of coronary arteries in coronary computed tomography angiography by using high-definition scans and deep learning image reconstruction
    Wang, Yiming
    Wang, Geliang
    Huang, Xin
    Zhao, Wenzhe
    Zeng, Qiang
    Li, Yanshou
    Guo, Jianxin
    [J]. QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (05) : 2933 - 2940
  • [5] Low-dose CT coronary angiography matchs regular-dose in image quality
    Bayol, A. P.
    Vallejos, J. A.
    Peloso, R. E.
    Aguero, M. A.
    Obregon, R.
    Zarza, A. C.
    Collante Bohle, M. A.
    Sandoval, D. H.
    Pozzer, P. A.
    Parras, J. I.
    [J]. EUROPEAN HEART JOURNAL, 2009, 30 : 489 - 489
  • [6] Coronary Angiography with Low-Dose Computed Tomography at 1.4 mSv
    Lars Husmann
    Ines Valenta
    Philipp A. Kaufmann
    [J]. Herz Kardiovaskuläre Erkrankungen, 2008, 33 : 75 - 75
  • [7] Coronary angiography with low-dose computed tomography at 1.4 mSv
    Husmann, Lars
    Valenta, Ines
    Kaufmann, Philipp A.
    [J]. HERZ, 2008, 33 (01) : 75 - 75
  • [8] Feasibility of ultra low-dose coronary computed tomography angiography
    McGraw, Sloane
    Carlson, Claire
    Grant, Katharine
    Nijjar, Prabhjot S.
    [J]. INDIAN HEART JOURNAL, 2018, 70 (03) : 443 - 445
  • [9] Diagnostic Performance of High-Definition Coronary Computed Tomography Angiography Performed With Multiple Radiation Dose Reduction Strategies
    Heydari, Bobak
    Leipsic, Jonathon
    Mancini, G. B. John
    Min, James K.
    LaBounty, Troy
    Taylor, C.
    Freue, Gabriela V. Cohen
    Heilbron, Brett
    [J]. CANADIAN JOURNAL OF CARDIOLOGY, 2011, 27 (05) : 606 - 612
  • [10] Image Quality in Computed Tomography Coronary Angiography and Radiation Dose Reduction
    Golubickas, Domas
    Lukosevicius, Saulius
    Tamakauskas, Vytenis
    Dobrovolskiene, Laima
    Baseviciene, Ingrida
    Grib, Liviu
    Ragaisyte, Nijole
    Leonavicius, Rytis
    Medvedev, Viktor
    Veikutis, Vincentas
    [J]. INFORMATICA, 2021, 32 (04) : 741 - 757