A comparison of automatic and manual compensation methods for the calculation of tube currents during off-centered patient positioning with a noise-based automatic exposure control system in computed tomography

被引:0
|
作者
Yasushi Furukawa
Kosuke Matsubara
Yoshinori Tsutsumi
机构
[1] Nagoya University Hospital,Department of Radiological Technology
[2] Kanazawa University,Department of Quantum Medical Technology, Graduate Course of Medical Science and Technology, Division of Health Sciences, Graduate School of Medical Science
[3] Kanazawa University,Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences
关键词
Computed tomography; Automatic exposure control; Off-centering; Table height; Localizer radiograph; Noise-based AEC;
D O I
暂无
中图分类号
学科分类号
摘要
Automatic exposure control (AEC) is used to optimize the X-ray tube output during computed tomography (CT) scans. However, calculation of the tube current by AEC can be affected when a patient is not aligned with the rotational center of the X-ray tube. An automatic couch height-positioning compensation mechanism provides a corrective function when the patient is off-center. In this study, we aimed to (a) evaluate the performance characteristics of the positioning compensation mechanism and (b) confirm whether our proposed compensation method can be properly applied to a noise-based AEC system even if the CT device is not equipped with a positioning compensation mechanism. An elliptical phantom was scanned at various table heights on systems without/with the positioning compensation mechanism. Expressions describing the offset from the gantry’s isocenter and adjusted standard deviation settings were derived and used in our proposed compensation method. A phantom was scanned at various table heights with our proposed compensation method, and volume CT dose index (CTDIvol) and image noise levels were obtained. An anthropomorphic chest phantom was also scanned using the proposed compensation method to verify its accuracy. When the positioning compensation mechanism was used, it yielded a constant CTDIvol and image noise levels at various table heights tested. A comparison between our proposed method and the positioning compensation mechanism for both the elliptical and chest phantoms yielded similar CTDIvol. Therefore, both automatic and manual positioning compensation methods are useful for avoiding AEC miscalculations in off-centered patient positioning cases.
引用
收藏
页码:823 / 832
页数:9
相关论文
共 1 条