Comprehensive characterisation of the IBA myQA SRS for SRS and SBRT patient specific quality assurance

被引:0
|
作者
Izabela Junis
Yousif Yousif
Rachel Stensmyr
Jeffrey Barber
机构
[1] Westmead Hospital,Sydney West Radiation Oncology Network, Crown Princess Mary Cancer Centre
[2] University of Sydney,Institute of Medical Physics
关键词
Radiotherapy; Stereotactic Radiation Therapy; Stereotactic Body Radiation Therapy; PSQA; CMOS detector array;
D O I
暂无
中图分类号
学科分类号
摘要
The myQA SRS (IBA) is a new to market 2D complementary metal oxide semiconductor detector array with an active area 140 × 120 mm2 and 0.4 mm resolution, making it a potential real-time dosimetry alternative to radiochromic film for stereotactic plan verification. Characterisation of the device was completed to assess performance. The dosimetric properties of the device were assessed for 6FF and 6FFF beams from a Varian TrueBeam STx with high definition multileaf collimator. Clinical suitability of the device for Patient Specific Quality Assurance was verified using ten SRS/SBRT plans, compared against other detectors, as well as multi leaf collimator (MLC) tests including picket fence and chair. Gamma analysis was performed using myQA software with criteria of 4%/1 mm. The device demonstrated compliance with recommended specifications for basic tests. After the required warm-up period, the maximum deviation in detector signal from initial readings was 0.2%. Short-term and long-term reproducibility was 0.1% (6FF) and 1.0% (6FFF), respectively. Dose linearity was within 0.3% (6FF) and 0.7% (6FFF) and dose-rate dependence within 1.7% (6FF) and 2.9% (6FFF) and were verified with a Farmer type ionization chamber (PTW 30013). Angular dependence was quantified for coplanar and non-coplanar situations. Output factors and beam profiles measured on the device showed agreement within 1% of baseline RAZOR diode (IBA) and CC04 ionisation chamber (IBA) measurements for field sizes 1 × 1 to 10 × 10 cm2. The minimum gamma (4%/1 mm) pass rates for MLC-pattern tests were 96.5% and 98.1% for the myQA SRS and film, respectively. The average gamma (4%/1 mm) pass rates for SBRT and SRS plans were 98.8% and 99.8% respectively. This work represents one of the first studies performed on the commissioning and performance characterisation of this novel device, demonstrating its accuracy and reliability, making it highly useful as a film alternative in stereotactic treatment plan verification.
引用
收藏
页码:327 / 337
页数:10
相关论文
共 50 条
  • [1] Comprehensive characterisation of the IBA myQA SRS for SRS and SBRT patient specific quality assurance
    Junis, Izabela
    Yousif, Yousif
    Stensmyr, Rachel
    Barber, Jeffrey
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (01) : 327 - 337
  • [2] IBA myQA SRS Detector for CyberKnife Robotic Radiosurgery Quality Assurance
    Padelli, Francesco
    Aquino, Domenico
    Fariselli, Laura
    De Martin, Elena
    [J]. APPLIED SCIENCES-BASEL, 2022, 12 (15):
  • [3] Safety and Quality Assurance in SRS and SBRT
    Jaffray, D.
    Benedict, S.
    Solberg, T.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [4] Improvement in SRS Commission and Patient-Specific Quality Assurance
    Ding, M.
    Heard, D.
    Ma, C.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [5] Portal Dosimetry for SRS/SBRT Patient Specific Plan Quality Control
    Zakikhani, R.
    Lee, H.
    Able, C.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2955 - 2955
  • [6] Comparison of IMRT Quality Assurance for SRS Plans Using MapCheck2 in High-Density Mode Versus MyQA SRS
    Thropp, C.
    Buchanan, L.
    Yang, Y.
    Cardarelli, G.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E954 - E955
  • [7] Patient Specific Quality Assurance for Multiple Tumor SRS Treatment with a Single Isocenter
    Jiang, H.
    Badkul, R. K.
    Demez, N.
    Kauweloa, K.
    Saleh, H.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E494 - E494
  • [8] SRS/SBRT Patient-Specific QA Device Comparison
    Epstein, D.
    Shacham, T.
    Levin, D.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E948 - E948
  • [9] Use of a plastic scintillator detector for patient-specific quality assurance of VMAT SRS
    Snyder, Jesse D.
    Sullivan, Rodney J.
    Wu, Xingen
    Covington, Elizabeth L.
    Popple, Richard A.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (09): : 143 - 148
  • [10] Small Field SRS/SBRT Patient Specific QA at Extended SSD
    Stathakis, S.
    Mavroidis, P.
    Esquivel, C.
    Myers, P.
    Li, Y.
    Crownover, R.
    Papanikolaou, N.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)