Q-Method for High-Resolution, Whole-Cell Patch-Clamp Impedance Measurements Using Square Wave Stimulation

被引:0
|
作者
Pavel Novák
Ivan Zahradník
机构
[1] Slovak Academy of Sciences,Institute of Molecular Physiology and Genetics
[2] Slovak Academy of Sciences,Institute of Molecular Physiology and Genetics
来源
关键词
Cell impedance; Membrane capacitance; Membrane resistance; Gating charge; Impedance resolution;
D O I
暂无
中图分类号
学科分类号
摘要
High-resolution measurements of cell impedance provide invaluable information on various cellular processes such as exocytosis, ion channel gating, or fertilization. The best recent techniques, although achieving impedance resolution at theoretical limits, have limited applicability due to their inherent constrains and high complexity. We report here a simple method of high-resolution impedance measurement, dubbed as the Q-method, based on measurement of a charge by integrating the cell current during square wave stimulation and on its decomposition into specific components related to segments of the voltage stimulus. Simple relations were derived allowing very fast and direct estimation of cell impedance parameters. The major advantages of the Q-method are its inherently low sensitivity to low-pass filtering, rejection of periodic interference signals, automatic on-the-fly adjustment of the stimulation frequency for the highest capacitance resolution, and simultaneous high-resolution low-crosstalk monitoring of membrane resistance, series resistance and parasitic capacitance in addition to membrane capacitance. Implementation of the Q-method is straightforward with any patch-clamp setup and any cell type. Theoretical grounds of the Q-method, including its resolution and the noise of individual parameters, are developed and experimentally verified.
引用
收藏
页码:1201 / 1212
页数:11
相关论文
共 32 条
  • [1] Q-method for high-resolution, whole-cell patch-clamp impedance measurements using square wave stimulation
    Novák, Pavel
    Zahradník, Ivan
    ANNALS OF BIOMEDICAL ENGINEERING, 2006, 34 (07) : 1201 - 1212
  • [2] Robust, high-resolution, whole cell patch-clamp capacitance measurements using square wave stimulation
    Thompson, RE
    Lindau, M
    Webb, WW
    BIOPHYSICAL JOURNAL, 2001, 81 (02) : 937 - 948
  • [3] PATCH-CLAMP WHOLE-CELL PULSE PROTOCOL MEASUREMENTS USING A MICROCOMPUTER
    SMITH, PM
    JOURNAL OF PHYSIOLOGY-LONDON, 1992, 446 : P72 - P72
  • [4] Evaluation of Synaptic Multiplicity Using Whole-cell Patch-clamp Electrophysiology
    Sunstrum, Julia K.
    Inoue, Wataru
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2019, (146):
  • [5] An integrated patch-clamp amplifier for high-density whole-cell recordings
    Weerakoon, Pujitha
    Klemic, Kate
    Sigworth, Fred J.
    Culurciello, Eugenio
    2007 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOLS 1-11, 2007, : 1205 - 1208
  • [6] STIMULATION OF OUTWARD CURRENT BY AUXIN IN WHOLE-CELL PATCH-CLAMP PROTOPLASTS FROM AVENA COLEOPTILES
    KELLER, CP
    VANVOLKENBURGH, E
    PLANT PHYSIOLOGY, 1993, 102 (01) : 105 - 105
  • [7] High-resolution scanning patch-clamp: new insights into cell function
    Gu, YC
    Gorelik, J
    Spohr, HA
    Shevchuk, A
    Lab, MJ
    Harding, SE
    Vodyanoy, I
    Klenerman, D
    Korchev, YE
    FASEB JOURNAL, 2002, 16 (03): : 748 - +
  • [8] Automated, in-vivo, whole-cell electrophysiology using an integrated patch-clamp amplifier
    Ilya Kolb
    Gregory Holst
    Brian Goldstein
    Suhasa B Kodandaramaiah
    Edward S Boyden
    Eugenio Culurciello
    Craig R Forest
    BMC Neuroscience, 14 (Suppl 1)
  • [9] WHOLE-CELL RECORDED K CURRENTS USING THE PATCH-CLAMP TECHNIQUE ARE INFLUENCED BY THE GLASS OF THE PATCH PIPETTE
    COTA, G
    ARMSTRONG, CM
    BIOPHYSICAL JOURNAL, 1987, 51 (02) : A49 - A49
  • [10] ACTION-POTENTIALS OF CULTURED HUMAN OAT CELLS - WHOLE-CELL MEASUREMENTS WITH THE PATCH-CLAMP TECHNIQUE
    JOHANSSON, S
    RYDQVIST, B
    SWERUP, C
    HEILBRONN, E
    ARHEM, P
    ACTA PHYSIOLOGICA SCANDINAVICA, 1989, 135 (04): : 573 - 578