This paper describes measurements aimed to develop the high-temperature scale for high-temperature standard platinum resistance thermometers (HT-SPRTs). A furnace designed to facilitate comparison of a radiation thermometer with a HT-SPRT was developed for this work. The measurement uncertainty of the comparison is first reported, and then the comparison results for eight HT-SPRTs for temperatures from 660 °C up to 1040 °C are presented. Two different methods were adopted to analyze the comparison results. Upon discussion of the comparison results and adopting the second method to analyze the data, a correction term to the reference function of HT-SPRTs for temperatures above the current high-temperature limit, the Ag freezing point (961.78 °C), is suggested, as a preliminary result upon seeking the possibility to propose the extension of the platinum resistance thermometer scale. The two irregular cases for discussion of the criteria for interpolating HT-SPRTs of the current International Temperature Scale of 1990 (ITS-90) are also discussed.
引用
收藏
页码:86 / 105
页数:19
相关论文
共 50 条
  • [1] Comparison Measurements for the Development of High-Temperature Platinum Resistance Scale
    Yamazawa, K.
    Widiatmo, J. V.
    Tamba, J.
    Arai, M.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2011, 32 (1-2) : 86 - 105
  • [2] A NEW HIGH-TEMPERATURE PLATINUM RESISTANCE THERMOMETER
    LI, XM
    ZHANG, JD
    SU, JR
    CHEN, DM
    METROLOGIA, 1982, 18 (04) : 203 - 208
  • [3] High-temperature furnace controlled by a platinum resistance thermometer
    Arai, M
    Yamazawa, K
    SICE 2004 ANNUAL CONFERENCE, VOLS 1-3, 2004, : 1172 - 1175
  • [4] Improved High-Temperature Standard Platinum Resistance Thermometer
    Zhao, M.
    Chen, D.
    Newman, M.
    Ding, R.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2010, 31 (8-9) : 1477 - 1483
  • [5] Improved High-Temperature Standard Platinum Resistance Thermometer
    M. Zhao
    D. Chen
    M. Newman
    R. Ding
    International Journal of Thermophysics, 2010, 31 : 1477 - 1483
  • [6] A Challenge to Improve High-Temperature Platinum Resistance Thermometer
    Y. Tanaka
    J. V. Widiatmo
    K. Harada
    T. Kobayashi
    K. Yamazawa
    International Journal of Thermophysics, 2017, 38
  • [7] A Challenge to Improve High-Temperature Platinum Resistance Thermometer
    Tanaka, Y.
    Widiatmo, J. V.
    Harada, K.
    Kobayashi, T.
    Yamazawa, K.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2017, 38 (05)
  • [8] EVALUATION OF SOME HIGH-TEMPERATURE PLATINUM RESISTANCE THERMOMETERS
    EVANS, JP
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1984, 89 (05): : 349 - 373
  • [9] Comparison of temperature scale maintained on platinum resistance thermometers with the NPL radiance temperature scale
    McEvoy, HC
    Chattle, MV
    Butler, J
    METROLOGIA, 1996, 33 (04) : 353 - 362
  • [10] Characterization of High-Temperature Platinum Resistance Thermometers at Silver Point
    Widiatmo, J. V.
    Harada, K.
    Yamazawa, K.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2017, 38 (03)