Development of a micro-total analysis system (μ-TAS) for the determination of catecholamines

被引:0
|
作者
Cafer Cakal
Jerome P. Ferrance
James P. Landers
Perihan Caglar
机构
[1] Hacettepe University,The Institute for Graduate Studies in Science and Engineering, Bioengineering Subdivision
[2] University of Virginia,Department of Chemistry
[3] Hacettepe University,Faculty of Science, Department of Chemistry
来源
关键词
Micro-total analysis system; Microchip electrophoresis; Microchip SPE; Catecholamines; Boronic acid functional monolith;
D O I
暂无
中图分类号
学科分类号
摘要
In this study, the first micro-total analysis system (μ-TAS) for catecholamines (dopamine, epinephrine, and norepinephrine) analysis in which preconcentration, separation, and determination steps were integrated on a microchip was developed. Electrophoresis microchips in a variety of channel lengths and designs were produced in borofloat glass for the μ-TAS studies. Chambers for the preparation of monolithic disks were formed in the microchips at the intersection of the injection and separation channels. Vinyl phenylboronic acid–ethylene glycol dimethacrylate polymers were prepared as monolithic disks in these chambers with a depth of 0.05 mm and a diameter of 2.1 mm. The microchips could be used more than 50 times if mechanical problems such as plugging or fracturing did not occur. Adsorption and elution of catecholamines were realized electrokinetically, with catecholamines determined via laser-induced native fluorescence detection following elution and electrophoretic separation. The most promising results were obtained with 100 mM phosphate buffer (pH 2) for elution with 25% propanol added to the separation buffer (100 mM phosphate, pH 3).
引用
收藏
页码:1909 / 1917
页数:8
相关论文
共 50 条
  • [1] Development of a micro-total analysis system (μ-TAS) for the determination of catecholamines
    Cakal, Cafer
    Ferrance, Jerome P.
    Landers, James P.
    Caglar, Perihan
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 398 (05) : 1909 - 1917
  • [2] Integrated micro-total analysis system (μTAS) for biophotonic enzymatic detections
    Chandrasekaran, Arvind
    Packirisamy, Muthukumaran
    ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS VIII, 2010, 7555
  • [3] Screen Printed Electromechanical Micro-Total Analysis System (μtas) For Sensitive and Rapid Detection of Infectious Diseases
    Nordin, Anis Nurashikin
    Zainuddin, Ahmad Anwar
    Ab Rahim, Rosminazuin
    Voiculescu, Ioana
    Mak, Wing Cheung
    BIOSENSORS 2016, 2017, 27 : 100 - 101
  • [4] An integrated micro optical and micro fluidic system for micro-total analysis system
    Yang, R
    Wu, MX
    Jin, GF
    MICROFLUIDIC DEVICES AND SYSTEMS III, 2000, 4177 : 208 - 211
  • [5] Micro Total Analysis System (μTAS)
    Shoji, S
    ELECTRONICS AND COMMUNICATIONS IN JAPAN PART II-ELECTRONICS, 1999, 82 (02): : 21 - 29
  • [6] Water glass bonding for micro-total analysis system
    Ito, T
    Sobue, K
    Ohya, S
    SENSORS AND ACTUATORS B-CHEMICAL, 2002, 81 (2-3) : 187 - 195
  • [7] Micro-total analysis system in undergraduate biochemistry and analytical chemistry
    Harwood, John
    Boles, Jeffrey O.
    Hunley, Ryan
    Mancuso, Matt
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2010, 239
  • [8] Characterisation of a micro-total analytical system for the determination of nitrite with spectrophotometric detection
    Greenway, GM
    Haswell, SJ
    Petsul, PH
    ANALYTICA CHIMICA ACTA, 1999, 387 (01) : 1 - 10
  • [9] Micro total analysis system (μ-TAS) in biotechnology
    S. J. Lee
    S. Y. Lee
    Applied Microbiology and Biotechnology, 2004, 64 : 289 - 299
  • [10] Micro total analysis system (μ-TAS) in biotechnology
    Lee, SJ
    Lee, SY
    APPLIED MICROBIOLOGY AND BIOTECHNOLOGY, 2004, 64 (03) : 289 - 299