MRI of the female pelvis at 3T compared to 1.5T: Evaluation on high-resolution T2-weighted and HASTE images

被引:45
|
作者
Kataoka, Masako
Kido, Aki
Koyama, Takashi
Isoda, Hiroyoshi
Umeoka, Shigeaki
Tamai, Ken
Nakamoto, Yuji
Maetani, Yoji
Morisawa, Nobuko
Saga, Tsuneo
Togashi, Kaori
机构
[1] Univ Cambridge, Addenbrookes Hosp, Dept Radiol, Cambridge CB2 2QQ, England
[2] Kyoto Univ, Dept Diagnost Imaging & Nucl Med, Grad Sch Med, Kyoto, Japan
[3] Kyoto Univ, Dept Radiol, Grad Sch Med, Kyoto, Japan
关键词
MRI; ultrahigh magnetic field; pelvis; uterus; ovary;
D O I
10.1002/jmri.20842
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the feasibility of MRI of the female pelvis using high-resolution T2-weighted imaging (T2WI) and the half-Fourier acquisition single-shot turbo spin-echo (HASTE) technique at 3 Tesla (T) compared to 1.5T, while focusing on the uterine body and cervical anatomy. Materials and Methods: A total of 19 healthy women underwent pelvic MR scans on 3T and 1.5T scanners. Axial and sagittal T2W (voxel size of 0.6 × 0.8 × 2 mm) and sagittal HASTE images were obtained. The images were evaluated qualitatively for overall image quality, contrast in the uterine zonal appearance and cervical structure, image inhomogeneity, and artifacts. A quantitative evaluation was performed regarding zonal contrast and image inhomogeneity. Results: On T2WI, the image contrast in the uterine cervix and vagina were significantly higher at 3T than at 1.5T, although there was no significant difference in the overall image quality or contrast in the uterine zonal appearance. Image inhomogeneity was more prominent at 3T, and motion artifact was more severe at 1.5T. Conclusion: Our results suggest that MRI of the female pelvis at 3T may potentially provide excellent images of the uterine cervix on high-resolution T2WI. New techniques to reduce inhomogeneity are thus called for. © 2007 Wiley-Liss, Inc.
引用
收藏
页码:527 / 534
页数:8
相关论文
共 50 条
  • [1] Signal intensity of motor and sensory cortices on T2-weighted and FLAIR images: intraindividual comparison of 1.5T and 3T MRI
    Kamada, Koji
    Kakeda, Shingo
    Ohnari, Norihiro
    Moriya, Junji
    Sato, Toru
    Korogi, Yukunori
    [J]. EUROPEAN RADIOLOGY, 2008, 18 (12) : 2949 - 2955
  • [2] Signal intensity of motor and sensory cortices on T2-weighted and FLAIR images: intraindividual comparison of 1.5T and 3T MRI
    Koji Kamada
    Shingo Kakeda
    Norihiro Ohnari
    Junji Moriya
    Toru Sato
    Yukunori Korogi
    [J]. European Radiology, 2008, 18 : 2949 - 2955
  • [3] T1 at 1.5T and 3T compared with conventional T2*at 1.5T for cardiac siderosis
    Alam, Mohammed H.
    Auger, Dominique
    Smith, Gillian C.
    He, Taigang
    Vassiliou, Vassilis
    Baksi, A. John
    Wage, Rick
    Drivas, Peter
    Feng, Yanqiu
    Firmin, David N.
    Pennell, Dudley J.
    [J]. JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2015, 17
  • [4] T1 at 1.5T and 3T compared with conventional T2* at 1.5T for cardiac siderosis
    Mohammed H. Alam
    Dominique Auger
    Gillian C. Smith
    Taigang He
    Vassilis Vassiliou
    A. John Baksi
    Rick Wage
    Peter Drivas
    Yanqiu Feng
    David N. Firmin
    Dudley J. Pennell
    [J]. Journal of Cardiovascular Magnetic Resonance, 17
  • [5] T2-Weighted Liver MRI Using the MultiVane Technique at 3T: Comparison with Conventional T2-Weighted MRI
    Kang, Kyung A.
    Kim, Young Kon
    Kim, EunJu
    Jeong, Woo Kyoung
    Choi, Dongil
    Lee, Won Jae
    Jung, Sin-Ho
    Baek, Sun-Young
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2015, 16 (05) : 1038 - 1046
  • [6] Short Breast MRI with High-Resolution T2-Weighted and Dynamic Contrast Enhanced T1-Weighted Images
    Ma, J.
    Son, J.
    Arun, B.
    Hazle, J.
    Hwang, K.
    Wang, K.
    Bayram, E.
    Madewell, J.
    Yang, W.
    Dogan, B.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3390 - 3390
  • [7] Carotid MRI Detection of Intraplaque Hemorrhage at 3T and 1.5T
    McNally, J. Scott
    Yoon, Hyo-Chun
    Kim, Seong-Eun
    Narra, Krishna K.
    McLaughlin, Michael S.
    Parker, Dennis L.
    Treiman, Gerald S.
    [J]. JOURNAL OF NEUROIMAGING, 2015, 25 (03) : 390 - 396
  • [8] Comparison of 1.5T and 3T MRI scanners in evaluation of acute bone stress in the foot
    Sormaala, Markus J.
    Ruohola, Juha-Petri
    Mattila, Ville M.
    Koskinen, Seppo K.
    Pihlajamaki, Harri K.
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2011, 12
  • [9] Comparison of 1.5T and 3T MRI scanners in evaluation of acute bone stress in the foot
    Markus J Sormaala
    Juha-Petri Ruohola
    Ville M Mattila
    Seppo K Koskinen
    Harri K Pihlajamäki
    [J]. BMC Musculoskeletal Disorders, 12
  • [10] HIGH-RESOLUTION MRI IMAGING AT 1.5T USING SURFACE COILS
    BLINDER, RA
    HERFKENS, RJ
    COLEMAN, RE
    JOHNSON, GA
    SCHENCK, JF
    HART, HR
    FOSTER, TH
    EDELSTEIN, WA
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1985, 26 (05) : P8 - P8