摘要
Background: Linear Accelerator-based Intensity Modulated Radiation Therapy (IMRT), either as step-and shoot (SS) or in dynamic mode, is now considered routine in the definitive management of head and neck squamous cell carcinoma (HNSCC). Helical TomoTherapy (HT) is a new platform to deliver IMRT. This study aims to compare step-and-shoot Intensity Modulated Radiation Therapy (SS IMRT) with dynamic Helical TomoTherapy (HT) dosimetrically in patients with head and neck squamous cell carcinoma (HNSCC). Materials and Methods: Twelve patients with HNSCC, previously treated with SS IMRT, were re-planned on HT using the same CT dataset. Plans were compared for target coverage and organs-at-risk (OARs) sparing. Sparing of parotids was assessed after stratifying for side (contralateral vs. ipsilateral) and site of disease (laryngopharynx vs. oropharynx). Normal tissue complication probabilities (NTCP) were also compared for the parotid glands. Results: All HT plans showed improvement in target coverage and homogeneity, and reduction in OAR doses as compared to SS IMRT plans. For PTV 66, the mean V(99) improved by 14.65% (P = 0.02). Dose Homogeneity (D(10-90)) was significantly better in the HT plans (mean 2.07Gy as compared to 4.5Gy in the SS IMRT plans, P = 0.02). HT resulted in an average reduction of mean parotid dose of 12.66Gy and 18.28Gy for the contralateral and ipsilateral glands (P = 0.003) respectively. This translated into a 24.09% and 35.22% reduction in Normal Tissue Complication Probability (NTCP) for the contralateral and ipsilateral parotids respectively (P < 0.01). Site of disease (laryngopharynx vs. oropharynx) did not have any significant impact on parotid sparing between SS IMRT and HT. The maximum dose to the spinal cord showed a mean reduction of 12.07Gy in HT plans (P = 0.02). Conclusion: Helical Tomotherapy achieved better target coverage with improved OAR sparing as compared to SS IMRT. The significant reduction in mean parotid doses translated into meaningful reduction in NTCP, with potential clinical implications in terms of reduction in Xerostomia and improved quality of life in patients with HNSCC.
引用
收藏
页码:194 / 198
页数:5
相关论文
共 50 条
  • [1] A comparison of the dosimetric effects of intrafraction motion on step-and-shoot, compensator, and helical tomotherapy-based IMRT
    Waghorn, Ben J.
    Staton, Robert J.
    Rineer, Justin M.
    Meeks, Sanford L.
    Langen, Katja M.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (03): : 121 - 132
  • [2] The Dosimetric Effects of Intrafraction Motion on Step-and-Shoot, Compensator and Helical Tomotherapy based IMRT
    Waghorn, B. J.
    Staton, R. J.
    Rineer, J. M.
    Meeks, S. L.
    Ngwa, W. F.
    Langen, K. M.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S809 - S809
  • [3] Comparing step-and-shoot IMRT with dynamic helical tomotherapy IMRT plans for head-and-neck cancer
    Van Vulpen, M
    Field, C
    Raaijmakers, CPJ
    Parliament, MB
    Terhaard, CHJ
    MacKenzie, MA
    Scrimger, R
    Lagendijk, JJW
    Fallone, BG
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (05): : 1535 - 1539
  • [4] IMRT dosimetry for prostate, breast and head-and-neck: comparing biologically based step-and-shoot IMRT with dynamic helical tomotherapy
    Ploquin, N.
    Belec, J.
    Clark, B.
    [J]. MEDICAL PHYSICS, 2009, 36 (09) : 4311 - 4311
  • [5] Volumetric Modulated Arc Radiotherapy for Nasopharyngeal Carcinoma: A Dosimetric Comparison with Tomotherapy and Step-and-shoot IMRT
    Lu, S.
    Wang, C.
    Chen, L.
    Lee, J.
    Wen, S.
    Wu, J.
    Lin, Y.
    Cheng, J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S816 - S816
  • [6] Volumetric modulated arc therapy for nasopharyngeal carcinoma: A dosimetric comparison with TomoTherapy and step-and-shoot IMRT
    Lu, Szu-Huai
    Cheng, Jason Chia-Hsien
    Kuo, Sung-Hsin
    Lee, Jason Jeun-Shenn
    Chen, Liang-Hsin
    Wu, Jian-Kuen
    Chen, Yu-Lisuan
    Chen, Wan-Yu
    Wen, Shu-Yu
    Chong, Fok-Ching
    Wu, Chien-Jang
    Wang, Chun-Wei
    [J]. RADIOTHERAPY AND ONCOLOGY, 2012, 104 (03) : 324 - 330
  • [7] Comparison of IMRT Techniques in the Radiotherapeutic Management of Head and Neck Cancer: Is Tomotherapy "Better" than Step-and-Shoot IMRT?
    Chen, A. M.
    Marsano, J.
    Perks, J.
    Farwell, G.
    Luu, Q.
    Donald, P. J.
    Purdy, J. A.
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2011, 10 (02) : 171 - 177
  • [8] Intensity-modulated radiation therapy (IMRT) dosimetry of the head and neck: A comparison of treatment plans using linear accelerator-based IMRT and helical tomotherapy
    Sheng, Ke
    Molloy, Janelle A.
    Read, Paul W.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (03): : 917 - 923
  • [9] A comparison of helical tomotherapy IMRT, conformal radiation therapy, and step-and-shoot IMRT for esophageal cancer
    Chen, Y
    Liu, A
    Han, C
    Tsai, PT
    Schultheiss, TE
    Wong, JYC
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 78 : S81 - S81
  • [10] Dosimetric comparisons of helical tomotherapy and step-and-shoot intensity-modulated radiotherapy in nasopharyngeal carcinoma
    Lee, Tsair-Fwu
    Fang, Fu-Min
    Chao, Pei-Ju
    Su, T. -J.
    Wang, Luke K.
    Leung, Stephen W.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 89 (01) : 89 - 96