Secondary shielding barriers for diagnostic X-ray facilities: Scatter and leakage revisited

被引:30
|
作者
Simpkin, DJ
Dixon, RL
机构
[1] Wake Forest Univ, Bowman Gray Sch Med, Winston Salem, NC 27157 USA
[2] St Lukes Med Ctr, Dept Radiol, Milwaukee, WI 53201 USA
来源
HEALTH PHYSICS | 1998年 / 74卷 / 03期
关键词
radiation protection; scattering; shielding; x rays;
D O I
10.1097/00004032-199803000-00008
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Methods are presented for determining the thickness requirements for barriers against scatter and leakage radiation generated in a diagnostic x-ray facility. Equations are developed that express the shielded doses due to scatter and leakage with explicit dependence on the operating potential used in clinical settings. The 1972 scatter experiment of Trout and Kelley is revisited, with suggested values for the scatter fraction that are somewhat different from those used in Report Number 49 of the NCRP, The dose from leakage radiation due to operation at clinical kVp settings is seen to be orders of magnitude below that predicted by the model in NCRP49. The unshielded secondary radiation dose in air at unit distance and workload is presented as a function of operating potential and for clinical workload distributions. The net transmission of secondary radiations through common shielding materials is presented. Compared to that for the primary beam at the same potential, the transmission of secondary radiation is more penetrating due to the inclusion of leakage radiation that has been hardened by the tube housing. Sample shielding calculations illustrate the utility of this scheme.
引用
收藏
页码:350 / 365
页数:16
相关论文
共 50 条
  • [1] Primary shielding barriers for diagnostic x-ray facilities: A new model
    Dixon, RL
    Simpkin, DJ
    HEALTH PHYSICS, 1998, 74 (02): : 181 - 189
  • [2] A model for calculating shielding requirements in diagnostic X-ray facilities
    Tsalafoutas, IA
    Yakoumakis, E
    Sandilos, P
    BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (910): : 731 - 737
  • [3] Investigation of scatter fractions for estimating leakage dose in medical X-ray imaging facilities
    Noto K.
    Koshida K.
    Iida H.
    Yamamoto T.
    Kobayashi I.
    Kawabata C.
    Radiological Physics and Technology, 2009, 2 (2) : 138 - 144
  • [4] A conversion method of air kerma from the primary, scatter, and leakage radiations to effective dose for calculating x-ray shielding barriers in mammography
    Kharrati, H
    MEDICAL PHYSICS, 2005, 32 (05) : 1367 - 1374
  • [5] HISTORY OF THE SHIELDING OF DIAGNOSTIC-X-RAY FACILITIES
    ARCHER, BR
    HEALTH PHYSICS, 1995, 69 (05): : 750 - 758
  • [6] Monte Carlo simulation of x-ray buildup factors of lead and its applications in shielding of diagnostic x-ray facilities
    Kharrati, Hedi
    Agrebi, Amel
    Karaoui, Mohamed-Karim
    MEDICAL PHYSICS, 2007, 34 (04) : 1398 - 1404
  • [7] Numerical models for the determination of primary structural barriers for diagnostic x-ray facilities
    Okunade, AA
    MEDICAL PHYSICS, 2004, 31 (03) : 513 - 520
  • [8] X-RAY SCATTER DATA FOR DIAGNOSTIC RADIOLOGY
    DICK, CE
    SOARES, CG
    MOTZ, JW
    PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (06): : 1076 - 1085
  • [9] A study of scatter in diagnostic X-ray rooms
    McVey, G
    Weatherburn, H
    BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (913): : 28 - 38
  • [10] An evaluation of the effectiveness of current shielding methodologies for diagnostic X-ray and radiation therapy facilities
    Shearer, D
    Donovan, G
    North, D
    Cardarelli, G
    MEDICAL PHYSICS, 2003, 30 (06) : 1399 - 1399