Excursion of Oscillating Saw Blades in Total Knee Arthroplasty

被引:23
|
作者
Dimitris, Kirk [1 ]
Taylor, Benjamin C. [1 ]
Steensen, Robert N. [1 ]
机构
[1] Mt Carmel Med Ctr, Dept Orthopaed Surg, Columbus, OH 43222 USA
来源
JOURNAL OF ARTHROPLASTY | 2010年 / 25卷 / 01期
关键词
saw blades; total knee arthroplasty; chamfer; medial collateral ligament injury; TOTAL CONDYLAR-III; MEDIAL COLLATERAL LIGAMENT; COMPLETE TEARS; PROSTHESIS; REPAIR; FOLLOW;
D O I
10.1016/j.arth.2008.09.021
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Intraoperative disruption of the medial collateral ligament during total knee arthroplasty is an uncommon complication that can be avoided by retractor placement as well as by careful cutting of the femur and tibia. This study evaluated the excursion of a small and large oscillating saw blade and compared the data against the widths of both the medial and lateral femoral condyle cuts. We discovered that the large saw blade had a statistically significantly larger excursion than the medial and lateral condyle width in women, as well as the lateral condyle width in men. The small saw blade excursion did not exceed any condyle width. We conclude that the smaller saw blade should be considered when making these cuts because the excursion of the large saw blade may exceed the width of cut needed and endanger important structures such as collateral ligaments.
引用
收藏
页码:158 / 160
页数:3
相关论文
共 50 条
  • [1] Demonstration of Saw Blade Accuracy and Excursion A Cadaveric Comparison Study of Blade Types Used in Total Knee Arthroplasty
    Wetzel, Robert J.
    Shah, Ritesh R.
    Puri, Lalit
    JOURNAL OF ARTHROPLASTY, 2013, 28 (06): : 985 - 987
  • [2] Saw-tooth fracture in total knee arthroplasty
    Tang, WM
    Chiu, K
    ORTHOPEDICS, 2004, 27 (03) : 325 - 326
  • [3] Durability of oscillating bone saw blades
    Ark, Timmon W.
    Thacker, John G.
    McGregor, Walter
    Rodeheaver, George T.
    Edlich, Richard F.
    1997, Begell House Inc Publ, New York, NY, United States (07) : 3 - 4
  • [4] Innovations in oscillating bone saw blades
    Ark, TW
    Thacker, JG
    McGregor, W
    Rodeheaver, GT
    Edlich, RF
    JOURNAL OF LONG-TERM EFFECTS OF MEDICAL IMPLANTS, 1997, 7 (3-4) : 279 - 286
  • [5] Durability of oscillating bone saw blades
    Ark, TW
    Thacker, JG
    McGregor, W
    Rodeheaver, GT
    Edlich, RF
    JOURNAL OF LONG-TERM EFFECTS OF MEDICAL IMPLANTS, 1997, 7 (3-4) : 271 - 278
  • [6] Soft-tissue penetration of the oscillating saw during tibial resection in total knee arthroplasty A CADAVERIC STUDY
    Herregodts, S.
    Verhaeghe, M.
    Paridaens, R.
    Herregodts, J.
    Vermue, H.
    Arnout, N.
    De Baets, P.
    Victor, J.
    BONE & JOINT JOURNAL, 2020, 102B (10): : 1324 - 1330
  • [7] Comparison of Tip- Versus Hub-Oscillating Saw Blade Control in a Total Knee Arthroplasty Model
    Christensen, David D.
    Beynnon, Bruce D.
    Daniels, Stephen D.
    Salinas, Cris J.
    Blankstein, Michael
    Nelms, Nathaniel J.
    JOURNAL OF ARTHROPLASTY, 2020, 35 (11): : 3300 - 3304
  • [8] Total Knee Arthroplasty With a Computer-navigated Saw: A Pilot Study
    Garvin, Kevin L.
    Barrera, Andres
    Mahoney, Craig R.
    Hartman, Curtis W.
    Haider, Hani
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (01) : 155 - 161
  • [9] SOFT TISSUE DAMAGE BY OSCILLATING SAW BLADE DURING TOTAL KNEE ARTHROPALSTY
    Kwon, Y.
    Kim, R.
    Botimer, G.
    Inceoglu, S.
    Williams, P.
    JOURNAL OF INVESTIGATIVE MEDICINE, 2012, 60 (01) : 166 - 166
  • [10] A technique for quantifying the performance of oscillating bone saw blades
    Ark, TW
    Thacker, JG
    McGregor, W
    Rodeheaver, GT
    Edlich, RF