Large-volume multi-tined expandable RF ablation in pig livers: comparison of 2D and volumetric measurements of the ablation zone

被引:7
|
作者
Bangard, Christopher [1 ]
Roesgen, Silvia [1 ]
Wahba, Roger [2 ]
Wiemker, Rafael [3 ]
Hellmich, Martin [4 ]
Reiter, Hannah [5 ]
Fischer, Jurgen H. [5 ]
Stippel, Dirk L. [2 ]
Lackner, Klaus J. [1 ]
机构
[1] Univ Cologne, Dept Radiol, D-50931 Cologne, Germany
[2] Univ Cologne, Dept Visceral & Vasc Surg, D-50931 Cologne, Germany
[3] Philips Forschungslaboratorien, Hamburg, Germany
[4] Univ Cologne, Inst Med Stat Informat & Epidemiol, D-50931 Cologne, Germany
[5] Univ Cologne, Inst Expt Med, D-50931 Cologne, Germany
关键词
Radiofrequency ablation; Computed tomography; Pig liver; Volume; 3D; 2D; PERCUTANEOUS RADIOFREQUENCY ABLATION; SMALL HEPATOCELLULAR-CARCINOMA; THERMAL ABLATION; SIZE; METASTASES; MONOPOLAR; NECROSIS; CREATION; LESIONS; SYSTEM;
D O I
10.1007/s00330-009-1639-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To compare two-dimensional (2D) and three-dimensional (3D) computed tomography (CT) measurements of ablation zones (AZs) related to the shaft of two different large-volume monopolar multi-tined expandable electrodes. Percutaneous radiofrequency (RF) ablation was performed in 12 pigs (81.6 +/- 7.8 kg) using two electrodes (LeVeen 5 cm, Rita XL 5 cm; n = 6 in each group). Contrast-enhanced CT with the electrode shaft in place evaluated the AZ. The largest sphere centred on the electrode shaft within the AZ was calculated (1) based on the 2D axial CT image in the plane of the shaft assuming rotational symmetry of the AZ and (2) using prototype software and the 3D volume data of the AZ measured with CT. The mean largest diameter of a sphere centred on the electrode shaft was always smaller using the 3D data of the AZ than using 2D CT measurements assuming rotational symmetry of the AZ (3D vs 2D): LeVeen 18.2 +/- 4.8 mm; 24.5 +/- 3.1 mm; p = 0.001; Rita XL 20.0 +/- 3.7 mm; 28.8 +/- 4.9 mm; p = 0.0002. All AZ showed indentations around the tines. Two-dimensional CT measurements assuming rotational symmetry of the AZ overestimate the largest ablated sphere centred on the electrode shaft compared with 3D CT measurements.
引用
收藏
页码:1073 / 1078
页数:6
相关论文
共 1 条
  • [1] Large-volume multi-tined expandable RF ablation in pig livers: comparison of 2D and volumetric measurements of the ablation zone
    Christopher Bangard
    Silvia Rösgen
    Roger Wahba
    Rafael Wiemker
    Martin Hellmich
    Hannah Reiter
    Jürgen H. Fischer
    Dirk L. Stippel
    Klaus J. Lackner
    European Radiology, 2010, 20 : 1073 - 1078