Automated Real-time Needle-Guide Tracking for Fast 3-T MR-guided Transrectal Prostate Biopsy: A Feasibility Study

被引:23
|
作者
Zamecnik, Patrik [1 ]
Schouten, Martijn G. [1 ]
Krafft, Axel J. [2 ]
Maier, Florian [3 ]
Schlemmer, Heinz-Peter [4 ]
Barentsz, Jelle O. [1 ]
Bock, Michael [5 ]
Fuetterer, Jurgen J. [1 ,6 ]
机构
[1] St Radboud Univ Nijmegen, Med Ctr, Dept Radiol, NL-6500 HB Nijmegen, Netherlands
[2] St Jude Childrens Res Hosp, Dept Radiol Sci, Memphis, TN 38105 USA
[3] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
[4] German Canc Res Ctr, Dept Radiol, Heidelberg, Germany
[5] Univ Freiburg, Dept Radiol, Div Med Phys, D-79106 Freiburg, Germany
[6] Univ Twente, MIRA Inst Biomed Technol & Tech Med, NL-7500 AE Enschede, Netherlands
关键词
CANCER; INTERVENTIONS; PARAMETERS; SYSTEM; MEN;
D O I
10.1148/radiol.14132067
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the feasibility of automatic needle-guide tracking by using a real-time phase-only cross correlation (POCC) algorithm-based sequence for transrectal 3-T inbore magnetic resonance (MR)-guided prostate biopsies. Materials and Methods: This study was approved by the ethics review board, and written informed consent was obtained from all patients. Eleven patients with a prostate-specific antigen level of at least 4 ng/mL (4 mu g/L) and at least one transrectal ultrasonography-guided biopsy session with negative findings were enrolled. Regions suspicious for cancer were identified on 3-T multiparametric MR images. During a subsequent MR-guided biopsy, the regions suspicious for cancer were reidentified and targeted by using the POCC-based tracking sequence. Besides testing a general technical feasibility of the biopsy procedure by using the POCC-based tracking sequence, the procedure times were measured, and a pathologic analysis of the biopsy cores was performed. Results: Thirty-eight core samples were obtained from 25 regions suspicious for cancer. It was technically feasible to perform the POCC-based biopsies in all regions suspicious for cancer in each patient, with adequate biopsy samples obtained with each biopsy attempt. The median size of the region suspicious for cancer was 8 mm (range, 4-13 mm). In each region suspicious for cancer (median number per patient, two; range, 1-4), a median of one core sample per region was obtained (range, 1-3). The median time for guidance per target was 1.5 minutes (range, 0.7-5 minutes). Nineteen of 38 core biopsy samples contained cancer. Conclusion: This study shows that it is feasible to perform transrectal 3-T MR-guided biopsies by using a POCC algorithm-based real-time tracking sequence. (C) RSNA, 2014
引用
收藏
页码:879 / 886
页数:8
相关论文
共 34 条
  • [1] Fast 3-T MR-guided transrectal prostate biopsy using an in-room tablet device for needle guide alignment: a feasibility study
    Christiaan G. Overduin
    Jan Heidkamp
    Eva Rothgang
    Jelle O. Barentsz
    Frank de Lange
    Jurgen J. Fütterer
    European Radiology, 2018, 28 : 4824 - 4831
  • [2] Fast 3-T MR-guided transrectal prostate biopsy using an in-room tablet device for needle guide alignment: a feasibility study
    Overduin, Christiaan G.
    Heidkamp, Jan
    Rothgang, Eva
    Barentsz, Jelle O.
    de Lange, Frank
    Futterer, Jurgen J.
    EUROPEAN RADIOLOGY, 2018, 28 (11) : 4824 - 4831
  • [3] MR-guided cholecystostomy: Assessment of biplanar, real-time needle tracking in three pigs
    Susanne C. Göhde
    Thomas Pfammatter
    Paul Steiner
    Peter Erhart
    Benjamin J. Romanowski
    Jörg F. Debatin
    CardioVascular and Interventional Radiology, 1997, 20 : 295 - 299
  • [4] MR-guided cholecystostomy: Assessment of biplanar, real-time needle tracking in three pigs
    Gohde, SC
    Pfammatter, T
    Steiner, P
    Erhart, P
    Romanowski, BJ
    Debatin, JF
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1997, 20 (04) : 295 - 299
  • [5] 3-T Multiparametric MRI Followed by In-Bore MR-Guided Biopsy for Detecting Clinically Significant Prostate Cancer After Prior Negative Transrectal Ultrasound-Guided Biopsy
    Hosseiny, Melina
    Shakeri, Sepideh
    Felker, Ely R.
    Lu, David
    Sayre, James
    Ahuja, Preeti
    Raman, Steven S.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 215 (03) : 660 - 666
  • [6] 3-T in-bore MR-guided prostate biopsy based on a scoring system for target lesions characterization
    Quentin, Michael
    Schimmoeller, Lars
    Arsov, Christian
    Rabenalt, Robert
    Antoch, Gerald
    Albers, Peter
    Blondin, Dirk
    ACTA RADIOLOGICA, 2013, 54 (10) : 1224 - 1229
  • [7] Real-time MR-guided Endourological Upper Urinary Tract Interventions: Feasibility Study
    Popeneciu, V
    Hadaschik, B.
    Pahernik, S.
    Homagk, A.
    Umathum, R.
    Galmbacher, R.
    Hallscheidt, P.
    Hohenfellner, M.
    UROLOGY, 2012, 80 (03) : S232 - S233
  • [8] In-Bore 3-T MR-guided Transrectal Targeted Prostate Biopsy: Prostate Imaging Reporting and Data System Version 2-based Diagnostic Performance for Detection of Prostate Cancer
    Tan, Nelly
    Lin, Wei-Chan
    Khoshnoodi, Pooria
    Asvadi, Nazanin H.
    Yoshida, Jeffrey
    Margolis, Daniel J. A.
    Lu, David S. K.
    Wu, Holden
    Sung, Kyung Hyun
    Lu, David Y.
    Huang, Jaioti
    Raman, Steven S.
    RADIOLOGY, 2017, 283 (01) : 130 - 139
  • [9] Fiber Bragg gratings-based sensing for real-time needle tracking during MR-guided brachytherapy
    de Battisti, Maxence Borot
    de Senneville, Baudouin Denis
    Maenhout, Metha
    Lagendijk, Jan J. W.
    van Vulpen, Marco
    Hautvast, Gilion
    Binnekamp, Dirk
    Moerland, Marinus A.
    MEDICAL PHYSICS, 2016, 43 (10) : 5288 - 5297
  • [10] Real-time MR-guided brain biopsy using 1.0-T open MRI scanner
    Xiangmeng He
    Ming Liu
    Chao Liu
    Jing Fang
    Yujun Xu
    Ligang Wang
    Jianfeng Xiang
    Roberto Blanco Sequeiros
    Chengli Li
    European Radiology, 2019, 29 : 85 - 92