Accuracy of proton stopping power estimation of silicone breast implants with single and dual-energy CT calibration techniques

被引:3
|
作者
Chacko, Michael S. [1 ,2 ]
Grewal, Hardev S. [1 ,3 ]
Wu, Dee [2 ]
Sonnad, Jagadeesh R. [2 ]
机构
[1] Oklahoma Proton Ctr, 5901 West Mem Rd, Oklahoma City, OK 73142 USA
[2] Univ Oklahoma, Dept Radiol Sci, Ctr Hlth Sci, Oklahoma City, OK USA
[3] Univ Oklahoma, Dept Radiat Oncol, Ctr Hlth Sci, Oklahoma City, OK USA
来源
关键词
breast; dual energy; proton therapy; stopping power; PREDICTION; DESIGN; GEL;
D O I
10.1002/acm2.13358
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A major contributing factor to proton range uncertainty is the conversion of computed tomography (CT) Hounsfield Units (HU) to proton relative stopping power (RSP). This uncertainty is elevated with implanted devices, such as silicone breast implants when computed with single energy CT (SECT). In recent years, manufacturers have introduced implants with variations in gel cohesivity. Deriving the RSP for these implants from dual-energy CT (DECT) can result in a marked reduction of the error associated with SECT. In this study, we investigate the validity of DECT calibration of HU to RSP on silicone breast implants of varying cohesivity levels. A DECT capable scanner was calibrated using the stoichiometric method of Bourque et al for SECT and DECT using a tissue substitute phantom. Three silicone breast implants of increasing gel cohesivity were measured in a proton beam of clinical energy to determine ground-truth RSP and water equivalent thickness (WET). These were compared to SECT-derived RSP at three CT spectrum energies and DECT with two energy pairs (80/140 kVp and 100/140 kVp) as obtained from scans with and without an anthropomorphic phantom. The RSP derived from parameters estimates from CT vendor-specific software (syngo.via) was compared. The WET estimates from SECT deviated from MLIC ground truth approximately +11%-19%, which would result in overpenetration if used clinically. Both the Bourque calibration and syngo.via WET estimates from DECT yielded error <= 0.5% from ground truth; no significant difference was found between models of varying gel cohesivity levels. WET estimates without the anthropomorphic phantom were significantly different than ground truth for the Bourque calibration. From these results, gel cohesivity had no effect on proton RSP. User-generated DECT calibration can yield comparably accurate RSP estimates for silicone breast implants to vendor software methods. However, care must be taken to account for beam hardening effects.
引用
收藏
页码:159 / 170
页数:12
相关论文
共 50 条
  • [1] Dual-energy CT for the evaluation of silicone breast implants
    Thorsten R. C. Johnson
    Isabelle Himsl
    Karin Hellerhoff
    Doris Mayr
    Dorothea Rjosk-Dendorfer
    Nina Ditsch
    Bernhard Krauss
    Klaus Friese
    Maximilian F. Reiser
    Miriam S. Lenhard
    European Radiology, 2013, 23 : 991 - 996
  • [2] Dual-energy CT for the evaluation of silicone breast implants
    Johnson, Thorsten R. C.
    Himsl, Isabelle
    Hellerhoff, Karin
    Mayr, Doris
    Rjosk-Dendorfer, Dorothea
    Ditsch, Nina
    Krauss, Bernhard
    Friese, Klaus
    Reiser, Maximilian F.
    Lenhard, Miriam S.
    EUROPEAN RADIOLOGY, 2013, 23 (04) : 991 - 996
  • [3] Dosimetric comparison of stopping power calibration with dual-energy CT and single-energy CT in proton therapy treatment planning
    Zhu, Jiahua
    Penfold, Scott N.
    MEDICAL PHYSICS, 2016, 43 (06) : 2845 - 2854
  • [4] Quantitative Dual-Energy CT Imaging for Proton Stopping Power Computation
    Han, D.
    Siebers, J.
    Williamson, J.
    MEDICAL PHYSICS, 2014, 41 (06) : 496 - 496
  • [5] Proton Stopping Power Ratio Estimation for Different Phantom Sizes Using Dual-Energy CT Scanners
    Taasti, V.
    Michalak, G.
    Deisher, A.
    Kruse, J.
    Muren, L.
    Petersen, J.
    Hansen, D.
    McCollough, C.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [6] Comparison of Integrated and Post-Reconstruction Dual-Energy CT Proton Stopping Power Ratio Estimation Approaches
    Zhang, S.
    Han, D.
    Politte, D.
    Williamson, J.
    O'Sullivan, J.
    MEDICAL PHYSICS, 2017, 44 (06) : 3004 - 3005
  • [7] A Deep Learning Approach for Dual-Energy CT-Based Proton Stopping-Power-Ratio Estimation
    Lee, H.
    Duan, X.
    Jia, X.
    Yang, M.
    MEDICAL PHYSICS, 2019, 46 (06) : E451 - E451
  • [8] The Optimization of Spectra Gap for Proton Stopping Power Estimation Using Dual-Energy CT Image Domain Method
    Han, D.
    Zhang, S.
    Ding, K.
    MEDICAL PHYSICS, 2020, 47 (06) : E664 - E664
  • [9] IDENTIFICATION OF LEAKING SILICONE IN BREAST IMPLANTS WITH DUAL-ENERGY RADIOGRAPHY
    DOBBINS, JT
    KORNGUTH, PJ
    ERGUN, DL
    RADIOLOGY, 1992, 185 : 250 - 250
  • [10] Dual Energy CT Proton Stopping Power Ratio Calibration and Validation with Animal Tissues
    Xie, Y.
    Yin, L.
    Ainsley, C.
    McDonough, J.
    Solberg, T.
    Lin, A.
    Teo, B.
    MEDICAL PHYSICS, 2016, 43 (06) : 3756 - 3756