Electromagnetic interference by transcutaneous neuromuscular electrical stimulation in patients with bipolar sensing implantable cardioverter defibrillators: A pilot safety study

被引:20
|
作者
Crevenna, R
Stix, G
Pleiner, J
Pezawas, T
Schmidinger, H
Quittan, M
Wolzt, M
机构
[1] AKH Wien, Dept Phys Therapy & Rehabil, A-1090 Vienna, Austria
[2] Allgemeines Krankenhaus Wien, Dept Cardiol, Vienna, Austria
[3] Allgemeines Krankenhaus Wien, Clin Pharmacol, Vienna, Austria
来源
关键词
electromagnetic interference; ICD; muscle stimulation;
D O I
10.1046/j.1460-9592.2003.00104.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Neuromuscular electrical stimulation (NMES) is a frequently applied therapy for the treatment of pain and a therapeutic option to increase thigh muscle strength and endurance capacity in patients with heart failure. Electromagnetic interference (EMI) by the signals with sensing of ICDs is possible. Eight patients with subpectoral ICD systems and different transvenous bipolar sensing leads were subjected to electrical stimulation of the neck and shoulder and of the thighs using different stimulation algorithms. EMI with ventricular sensing was detectable in three of eight subjects. EMI occurred during stimulation of the neck (n = 2) and thigh (n = 2). EMI by NMES with atrial sensing was seen in two of four subjects with dual chamber ICDs. The safety of peripheral NMES has to be individually tested as EMI can also occur in ICD patients with bipolar sensing.
引用
收藏
页码:626 / 629
页数:4
相关论文
共 50 条
  • [1] Long-term transcutaneous neuromuscular electrical stimulation in patients with bipolar sensing implantable cardioverter defibrillators: A pilot safety study
    Crevenna, R
    Wolzt, M
    Fialka-Moser, V
    Keilani, M
    Nuhr, M
    Paternostro-Sluga, T
    Pacher, R
    Mayr, W
    Quittan, M
    ARTIFICIAL ORGANS, 2004, 28 (01) : 99 - 102
  • [2] Neuromuscular electrical stimulation of the thighs in cardiac patients with implantable cardioverter defibrillators
    Cenik, Fadime
    Schoberwalter, Dieter
    Keilani, Mohammad
    Maehr, Bruno
    Wolzt, Michael
    Marhold, Maximilian
    Crevenna, Richard
    WIENER KLINISCHE WOCHENSCHRIFT, 2016, 128 (21-22) : 802 - 808
  • [3] Neuromuscular electrical stimulation of the thighs in cardiac patients with implantable cardioverter defibrillators
    Fadime Cenik
    Dieter Schoberwalter
    Mohammad Keilani
    Bruno Maehr
    Michael Wolzt
    Maximilian Marhold
    Richard Crevenna
    Wiener klinische Wochenschrift, 2016, 128 : 802 - 808
  • [4] Safety of neuromuscular electrical stimulation in patients implanted with cardioverter defibrillators
    Kamiya, Kentaro
    Satoh, Akira
    Niwano, Shinichi
    Tanaka, Shinya
    Miida, Kazumasa
    Hamazaki, Nobuaki
    Maekawa, Emi
    Matsuzawa, Ryota
    Nozaki, Kohei
    Masuda, Takashi
    Ako, Junya
    JOURNAL OF ELECTROCARDIOLOGY, 2016, 49 (01) : 99 - 101
  • [5] Risk of occurrence of electromagnetic interference from the application of transcutaneous electrical nerve stimulation on the sensing function of implantable defibrillators
    Arain, Saba Suhail
    Cretnik, Anja
    Huemer, Martin
    Attanasio, Philipp
    Nagel, Patrick
    Landmesser, Ulf
    Hardt, Juliane
    Sidhu, Kiran
    Tscholl, Verena
    Roser, Mattias
    EUROPACE, 2023, 25 (07):
  • [6] Risk of interference from transcutaneous electrical nerve stimulation on the sensing function of implantable defibrillators
    Holmgren, Christina
    Carlsson, Tobias
    Mannheimer, Clas
    Edvardsson, Nils
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (02): : 151 - 158
  • [7] Interference of programmed electromagnetic stimulation with pacemakers and automatic implantable cardioverter defibrillators
    Gwechenberger, Marianne
    Rauscha, Friedrich
    Stix, Guenter
    Schmid, Gernot
    Strametz-Juranek, Jeanette
    BIOELECTROMAGNETICS, 2006, 27 (05) : 365 - 377
  • [8] Intraoperative Electrosurgical Electromagnetic Interference in Patients with Implantable Cardioverter Defibrillators Is It Safe?
    Rooke, G. Alec
    ANESTHESIOLOGY, 2019, 130 (04) : 523 - 525
  • [9] Transcutaneous electrical nerve stimulation electromagnetic interference in an implantable loop recorder
    Suarez-Fuster, Laiden
    Oh, Christopher
    Baranchuk, Adrian
    JOURNAL OF ARRHYTHMIA, 2018, 34 (01) : 96 - 97
  • [10] Safety of Electromagnetic Articulography in Patients With Pacemakers and Implantable Cardioverter-Defibrillators
    Joglar, Jose A.
    Nguyen, Carol
    Garst, Diane M.
    Katz, William F.
    JOURNAL OF SPEECH LANGUAGE AND HEARING RESEARCH, 2009, 52 (04): : 1082 - 1087