Ambient neutron and photon dose equivalent H*(10) around a pencil beam scanning proton therapy facility

被引:1
|
作者
Sharma, Dayananda Shamurailatpam [1 ]
Patro, Kartikeswar C. H. [1 ]
Padannayel, Noufal Manthala [1 ]
Arjunan, Manikandan [1 ]
Krishnan, Ganapathy [1 ]
Thiyagarajan, Rajesh [1 ]
Chilukuri, Srinivas [2 ]
Jalali, Rakesh [2 ]
机构
[1] Apollo Proton Canc Ctr, Dept Med Phys, 100 Feet Rd Tharamani, Chennai, Tamil Nadu, India
[2] Apollo Proton Canc Ctr, Dept Radiat Oncol, 100 Feet Rd Tharamani, Chennai, Tamil Nadu, India
来源
BRITISH JOURNAL OF RADIOLOGY | 2019年 / 92卷 / 1102期
关键词
D O I
10.1259/bjr.20190382
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To measure leakage ambient dose equivalent H*(10) from stray secondary neutron and photon radiation around proton therapy (PT) facility and evaluate adequacy of shielding design. Methods and materials: H*(10) measurement were carried out at 149 locations around cyclotron vault (CV), beam transport system (BTS) and first treatment room (GTR3) of a multiroom PT facility using WENDI-11 and Smartlon survey meter. Measurement were performed under extreme case scenarios wherein maximum secondary neutrons and photons were produced around CV, BTS and GTR3 by stopping 230MeV proton of 300nA on beam degrader, end of BTS and isocenter of GTR3. Weekly time average dose rate (TADR) were calculated from EI*(10) value measured at selective hot spots by irradiating actual treatment plans of mix clinical sites. Results: The maximum total H*(10) were within 2 mu Sy/hr around CV, 5 mu Sv/hr around outer wall of BTS which increases up to 62 mu Sv/hr at the end of inside BTS corridor. Maximum H*(10) of 20.8 mu Sv/hr in treatment control console (P125), 23.4 mu Sv/hr behind the common wall between GTR3 and GTR2 (P132) and 25.7 mu Sv/hr above isocenter (P99) were observed around GTR3. Reduction of beam current from 6 to 3 nA and 1 nA at nozzle exit lead to decrease in total H*(10) at P125 from 20.8 to 11.35 and 4.62 mu Sv/hr. In comparison to extreme case scenario, H*(10) value at P125, P132 and P99 from clinically relevant irradiation parameters were reduce by a factor ranging from 8.6 for high range cube to 46.4 for brain clinical plan. The maximum weekly TADR per fraction was highest for large volume, sacral chordoma patient at 8.5 mu Sv/hr compare to 0.3 mu Sv/hr for brain patient. The calculated weekly TADR for 30 mix clinical cases and 15 fractions of 1 L cube resulted total weekly TADR of 83-84 mu Sv/hr at P125, P132 and P99. The maximum annual dose level at these hot spots were estimated at 4.37 mSv/Yr. Conclusion: We have carried out an extensive measurement of H*(10) under different conditions. The shielding thickness of our PT facility is adequate to limit the dose to occupational worker and general public within the permissible stipulated limit. The data reported here can bridge the knowledge gap in ambient dose around PT facility and can also be used as a reference for any new and existing proton facility for intercomparison and validation. Advances in knowledge: First extensive investigation of neutron and photon H*(10) around PT facility and can bridge the knowledge gap on ambient dose.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Neutron Dose Equivalent Evaluation for Pencil Beam Scanning Proton Therapy with Apertures
    Geng, C.
    Schuemann, J.
    Moteabbed, M.
    Paganetti, H.
    MEDICAL PHYSICS, 2015, 42 (06) : 3466 - 3466
  • [2] Measurement of Neutron Spectrum and Ambient Dose Equivalent Around a Mini-Phantom at a Proton Therapy Facility
    Wang, Z.
    Kry, S.
    Burgett, E.
    Howell, R.
    Tailor, R.
    Oliver, J.
    Followill, D.
    Smith, A.
    Salehpour, M.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [3] Secondary neutron ambient dose equivalent measurement of the wobbling system of a proton beam radiotherapy facility
    Liao, Ying-Lan
    Chen, Hsien-Hsin
    Chen, Hsin-Yu
    Huang, Hsiao-Chieh
    Yeh, Chien-Yi
    Tsai, Hui-Yu
    RADIATION PHYSICS AND CHEMISTRY, 2017, 137 : 116 - 120
  • [4] Evaluation of Secondary Neutron Dose Equivalent to a Point Representing a Fetus During Pencil Beam Scanning Proton Therapy Treatment of the Mother
    Hecksel, D.
    Pankuch, M.
    Gao, M.
    MEDICAL PHYSICS, 2018, 45 (06) : E495 - E496
  • [5] Monte Carlo Simulations of Neutron Ambient Dose Equivalent in a Novel Proton Therapy Facility Design
    Titt, Uwe
    Pera, Enzo
    Gillin, Michael T.
    INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2020, 6 (04) : 29 - 37
  • [6] Determination of surface dose in pencil beam scanning proton therapy
    Kern, A.
    Baeumer, C.
    Kroeninger, K.
    Mertens, L.
    Timmermann, B.
    Walbersloh, J.
    Wulff, J.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S483 - S483
  • [7] Determination of surface dose in pencil beam scanning proton therapy
    Kern, A.
    Baeumer, C.
    Kroeninger, K.
    Mertens, L.
    Timmermann, B.
    Walbersloh, J.
    Wulff, J.
    MEDICAL PHYSICS, 2020, 47 (05) : 2277 - 2288
  • [8] The Effect of Proton Therapy Beam Scanning Pattern Size On Secondary Neutron Dose Equivalent
    Hecksel, D.
    Anferov, V.
    Shahnazi, K.
    Fitzek, M.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [9] Secondary neutron dose contribution from pencil beam scanning, scattered and spatially fractionated proton therapy
    Leite, A. M. M.
    Ronga, M. G.
    Giorgi, M.
    Ristic, Y.
    Perrot, Y.
    Trompier, F.
    Prezado, Y.
    Crehange, G.
    De Marzi, L.
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (22):
  • [10] Monte Carlo Dose Verification of Pencil Beam Scanning Proton Therapy
    Giantsoudi, D.
    Clasie, B.
    Grassberger, C.
    Dowdell, S.
    Depauw, N.
    Madden, T.
    Kooy, H.
    Paganetti, H.
    MEDICAL PHYSICS, 2013, 40 (06)