A clinical evaluation of the quantitative accuracy of simultaneous emission/transmission scanning in whole-body positron emission tomography

被引:9
|
作者
Lodge, MA [1 ]
Badawi, RD [1 ]
Marsden, PK [1 ]
机构
[1] United Med & Dent Sch Guys & St Thomas Hosp, Clin PET Ctr, London SE1 9RT, England
关键词
whole-body positron emission tomography; simultaneous emission/transmission; quantification;
D O I
10.1007/s002590050240
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present a clinical evaluation of the quantitative bias which is introduced during simultaneous emission/transmission (SET) acquisition for the application of whole-body positron emission tomography (PET) with fluorine-18 2-fluoro-2-deoxy-D-glucose. The quantitative accuracy of the SET technique was assessed by means of a clinical study involving 28 patients and a realistic phantom experiment. In the clinical study, SET overestimated the activity concentration in the tumours by a factor of approximately 1.10, but in the phantom study, where the tumours were smaller, the bias was found to increase to a value of 1.39. The bias in the soft tissue regions of the patient studies varied between 1.03 and 1.36, and close agreement was observed with the corresponding phantom results. The extent of the bias increased as the local activity concentration decreased and we attribute the effect to scattered photons from the transmission source which are detected in the emission window during SET.
引用
收藏
页码:417 / 423
页数:7
相关论文
共 50 条
  • [1] A clinical evaluation of the quantitative accuracy of simultaneous emission/transmission scanning in whole-body positron emission tomography
    Martin A. Lodge
    Ramsey D. Badawi
    Paul K. Marsden
    European Journal of Nuclear Medicine, 1998, 25 : 417 - 423
  • [2] Whole-body positron emission tomography and positron emission tomography/computed tomography in gynecologic oncology
    Kim, EE
    INTERNATIONAL JOURNAL OF GYNECOLOGICAL CANCER, 2004, 14 (01) : 12 - 22
  • [3] WHOLE-BODY TRANSMISSION-EMISSION SCANNING WITH WHOLE-BODY CAMERA
    SUZUKI, Y
    MORI, H
    MICHIGISHI, T
    HISADA, K
    MATSUDAIRA, M
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1975, 125 (04) : 978 - 980
  • [4] WHOLE-BODY POSITRON EMISSION TOMOGRAPHY IN BREAST-CANCER
    NITZSCHE, EU
    HOH, CK
    DALBOHM, NM
    GLASPY, JA
    PHELPS, ME
    MOSER, EA
    HAWKINS, RA
    FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1993, 158 (04): : 293 - 298
  • [5] SIMULTANEOUS TRANSMISSION AND EMISSION SCANS IN POSITRON EMISSION TOMOGRAPHY
    THOMPSON, CJ
    RANGER, NT
    EVANS, AC
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) : 1011 - 1016
  • [6] Whole-Body Simultaneous Positron Emission Tomography (PET)-MR: Optimization and Adaptation of MRI Sequences
    Fowler, Kathryn J.
    McConathy, Jon
    Narra, Vamsi R.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2014, 39 (02) : 259 - 268
  • [7] Clinical evaluation of whole-body 18F-fluordeoxyglucose positron emission tomography in the detection of liver metastases
    Hustinx, R
    Paulus, P
    Jacquet, N
    Jerusalem, G
    Bury, T
    Rigo, P
    ANNALS OF ONCOLOGY, 1998, 9 (04) : 397 - 401
  • [8] THE ROLE OF POSITRON EMISSION TOMOGRAPHY IN ONCOLOGY AND OTHER WHOLE-BODY APPLICATIONS
    HAWKINS, RA
    HOH, C
    GLASPY, J
    CHOI, Y
    DAHLBOM, M
    REGE, S
    MESSA, C
    NIETSZCHE, E
    HOFFMAN, E
    SEEGER, L
    MADDAHI, J
    PHELPS, ME
    SEMINARS IN NUCLEAR MEDICINE, 1992, 22 (04) : 268 - 284
  • [9] Whole-body tracking of single cells via positron emission tomography
    Jung, Kyung Oh
    Kim, Tae Jin
    Yu, Jung Ho
    Rhee, Siyeon
    Zhao, Wei
    Ha, Byunghang
    Red-Horse, Kristy
    Gambhir, Sanjiv Sam
    Pratx, Guillem
    NATURE BIOMEDICAL ENGINEERING, 2020, 4 (08) : 835 - 844
  • [10] Detection of lymphoma in bone marrow by whole-body positron emission tomography
    Carr, R
    Barrington, SF
    Madan, B
    O'Doherty, MJ
    Saunders, CAB
    van der Walt, J
    Timothy, AR
    BLOOD, 1998, 91 (09) : 3340 - 3346