Visualization of Volatile Substances in Different Organelles with an Atmospheric-Pressure Mass Microscope

被引:106
|
作者
Harada, Takahiro [3 ]
Yuba-Kubo, Akiko [2 ]
Sugiura, Yuki [1 ,4 ]
Zaima, Nobuhiro [1 ]
Hayasaka, Takahiro [1 ]
Goto-Inoue, Naoko [1 ]
Wakui, Masatoshi [2 ]
Suematsu, Makoto [2 ]
Takeshita, Kengo [3 ]
Ogawa, Kiyoshi [3 ]
Yoshida, Yoshikazu [3 ]
Setou, Mitsutoshi [1 ]
机构
[1] Hamamatsu Univ Sch Med, Dept Mol Anat, Mol Imaging Frontier Res Ctr, Higashi Ku, Shizuoka 4313192, Japan
[2] Keio Univ, Sch Med, Dept Biochem & Integrat Med Biol, Shinjuku Ku, Tokyo 1608582, Japan
[3] Shimadzu Co Ltd, Technol Res Lab, Kyoto 6190237, Japan
[4] Tokyo Inst Technol, Dept Biosci & Biotechnol, Midori Ku, Kanagawa 2268501, Japan
基金
日本科学技术振兴机构;
关键词
GINGER ZINGIBER-OFFICINALE; HIGH-SPATIAL-RESOLUTION; TISSUE-SECTIONS; SMALL MOLECULES; RAT-BRAIN; SPECTROMETRY; DESORPTION/IONIZATION; METABOLITES; IDENTIFICATION; METABOLOMICS;
D O I
10.1021/ac901872n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We have developed a mass microscope (mass spectrometry imager with spatial resolution higher than the naked eye) equipped with an atmospheric pressure ion-source chamber for laser desorption/ionization (AP-LDI) and a quadrupole ion trap time-of-flight (QIT-TOF) analyzer. The optical microscope combined with the mass spectrometer permitted us to precisely determine the relevant tissue region prior to performing imaging mass spectrometry (IMS). An ultraviolet laser tightly focused with a triplet lens was used to achieve high spatial resolution. An atmospheric pressure ion-source chamber enables us to analyze fresh samples with minimal loss of intrinsic water or volatile compounds. Mass-microscopic AP-LDI imaging of freshly cut ginger rhizome sections revealed that 6-gingerol ([M + K](+) at m/z 333.15, positive mode; [M - H](-) at m/z 293.17, negative mode) and the monoterpene ([M + Kr](-) at m/z 191.09), which are the compounds related to pungency and flavor, respectively, were localized in oil drop-containing organelles. AP-LDI-tandem MS/MS analyses were applied to compare authentic signals from freshly cut ginger directly with the standard reagent. Thus, our atmosphere-imaging mass spectrometer enabled us to monitor a quality of plants at the organelle level.
引用
收藏
页码:9153 / 9157
页数:5
相关论文
共 50 条
  • [1] Visualization of biomolecules in the eyestalk of the blue swimming crab, Portunus pelagicus, by imaging mass spectrometry using the atmospheric-pressure mass microscope
    Sroyraya, Morakot
    Goto-Inoue, Naoko
    Zaima, Nobuhiro
    Hayasaka, Takahiro
    Chansela, Piyachat
    Tanasawet, Supita
    Shrivas, Kamlesh
    Sobhon, Prasert
    Setou, Mitsutoshi
    SURFACE AND INTERFACE ANALYSIS, 2010, 42 (10-11) : 1589 - 1592
  • [2] Atmospheric-pressure Penning ionization mass spectrometry
    Hiraoka, K
    Fujimaki, S
    Kambara, S
    Furuya, H
    Okazaki, S
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (19) : 2323 - 2330
  • [3] ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY
    MITCHUM, RK
    KORFMACHER, WA
    ANALYTICAL CHEMISTRY, 1983, 55 (14) : 1485 - +
  • [4] ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY
    PROCTOR, CJ
    TODD, JFJ
    ORGANIC MASS SPECTROMETRY, 1983, 18 (12): : 509 - 516
  • [5] Mass spectrometry of atmospheric-pressure ball plasmoids
    Dubowsky, Scott E.
    Friday, David M.
    Peters, Kevin C.
    Zhao, Zhangji
    Perry, Richard H.
    McCall, Benjamin J.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2015, 376 : 39 - 45
  • [6] ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY
    CARROLL, DI
    DZIDIC, I
    HORNING, EC
    STILLWELL, RN
    APPLIED SPECTROSCOPY REVIEWS, 1981, 17 (03) : 337 - 406
  • [8] ION EXTRACTION FROM SOLUTIONS AT ATMOSPHERIC-PRESSURE - A METHOD OF MASS-SPECTROMETRIC ANALYSIS OF BIOORGANIC SUBSTANCES
    ALEXANDROV, ML
    GALL, LN
    KRASNOV, NV
    NIKOLAEV, VI
    PAVLENKO, VA
    SHKUROV, VA
    DOKLADY AKADEMII NAUK SSSR, 1984, 277 (02): : 379 - 383
  • [9] ATMOSPHERIC-PRESSURE ION EVAPORATION MASS-SPECTROMETRY
    IRIBARNE, JV
    DZIEDZIC, PJ
    THOMSON, BA
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1983, 50 (03): : 331 - 347
  • [10] DETECTION MASS BIAS IN ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY
    ZOOK, DR
    GRIMSRUD, EP
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1991, 2 (03) : 232 - 239