Parallax error in long-axial field-of-view PET scanners-a simulation study

被引:49
|
作者
Schmall, Jeffrey P. [1 ]
Karp, Joel S. [1 ,2 ]
Werner, Matt [1 ]
Surti, Suleman [1 ]
机构
[1] Univ Penn, Dept Radiol, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Phys & Astron, Philadelphia, PA 19104 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2016年 / 61卷 / 14期
基金
美国国家卫生研究院;
关键词
positron emission tomography; total-body imaging; depth-of-interaction; IMAGE-RECONSTRUCTION; 3D PET; PERFORMANCE; FLIGHT; SENSITIVITY; DETECTOR; SCINTILLATOR; DESIGN;
D O I
10.1088/0031-9155/61/14/5443
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There is a growing interest in the design and construction of a PET scanner with a very long axial extent. One critical design challenge is the impact of the long axial extent on the scanner spatial resolution properties. In this work, we characterize the effect of parallax error in PET system designs having an axial field-of-view (FOV) of 198 cm (total-body PET scanner) using fully-3D Monte Carlo simulations. Two different scintillation materials were studied: LSO and LaBr3. The crystal size in both cases was 4 x 4 x 20 mm(3). Several different depth-of-interaction (DOI) encoding techniques were investigated to characterize the improvement in spatial resolution when using a DOI capable detector. To measure spatial resolution we simulated point sources in a warm background in the center of the imaging FOV, where the effects of axial parallax are largest, and at several positions radially offset from the center. Using a line-of-response based ordered-subset expectation maximization reconstruction algorithm we found that the axial resolution in an LSO scanner degrades from 4.8 mm to 5.7 mm (full width at half max) at the center of the imaging FOV when extending the axial acceptance angle (alpha) from +/- 12 degrees (corresponding to an axial FOV of 18 cm) to the maximum of +/- 67 degrees-a similar result was obtained with LaBr3, in which the axial resolution degraded from 5.3 mm to 6.1 mm. For comparison we also measured the degradation due to radial parallax error in the transverse imaging FOV; the transverse resolution, averaging radial and tangential directions, of an LSO scanner was degraded from 4.9 mm to 7.7 mm, for a measurement at the center of the scanner compared to a measurement with a radial offset of 23 cm. Simulations of a DOI detector design improved the spatial resolution in all dimensions. The axial resolution in the LSO-based scanner, with alpha = +/- 67 degrees, was improved from 5.7 mm to 5.0 mm by incorporating a two-layer DOI detector. These results characterize the maximum axial blurring for a fully open 2 m long PET scanner and demonstrate that large sensitivity gains are possible with a modest reduction in resolution when using current clinical detector technology with no DOI capability.
引用
下载
收藏
页码:5443 / 5455
页数:13
相关论文
共 50 条
  • [1] Performance Characteristics of Long Axial Field-of-View PET Scanners With Axial Gaps
    Daube-Witherspoon, Margaret E.
    Viswanath, Varsha
    Werner, Matthew E.
    Karp, Joel S.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2021, 5 (03) : 322 - 330
  • [2] Performance Characteristics of Long Axial Field-of-View PET Scanners with Axial Gaps
    Daube-Witherspoon, Margaret E.
    Viswanath, Varsha
    Werner, Matthew E.
    Karp, Joel S.
    IEEE Transactions on Radiation and Plasma Medical Sciences, 2021, 5 (03): : 322 - 330
  • [3] Long-axial field-of-view PET/CT improves radiomics feature reliability
    Ian L. Alberts
    Song Xue
    Hasan Sari
    Lara Cavinato
    George Prenosil
    Ali Afshar-Oromieh
    Clemens Mingels
    Kuangyu Shi
    Federico Caobelli
    Arman Rahmim
    Thomas Pyka
    Axel Rominger
    European Journal of Nuclear Medicine and Molecular Imaging, 2025, 52 (3) : 1004 - 1016
  • [4] Reconstruction performance for long axial field-of-view PET scanners with large axial gaps
    Daube-Witherspoon, Margaret E.
    Viswanath, Varsha
    Surti, Suleman
    Karp, Joel S.
    15TH INTERNATIONAL MEETING ON FULLY THREE-DIMENSIONAL IMAGE RECONSTRUCTION IN RADIOLOGY AND NUCLEAR MEDICINE, 2019, 11072
  • [5] Clinical applications of long axial field-of-view PET/CT scanners in oncology
    Urso, Luca
    Frantellizzi, Viviana
    De Vincentis, Giuseppe
    Schillaci, Orazio
    Filippi, Luca
    Evangelista, Laura
    CLINICAL AND TRANSLATIONAL IMAGING, 2023, 11 (04) : 365 - 380
  • [6] Clinical applications of long axial field-of-view PET/CT scanners in oncology
    Luca Urso
    Viviana Frantellizzi
    Giuseppe De Vincentis
    Orazio Schillaci
    Luca Filippi
    Laura Evangelista
    Clinical and Translational Imaging, 2023, 11 : 365 - 380
  • [7] Characterization of the scatter component in large axial field-of-view PET scanners: a Monte Carlo simulation study
    Ghabrial, Amir
    Franklin, Daniel
    Zaidi, Habib
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [8] A Monte Carlo simulation study of scatter fraction and the impact of patient BMI on scatter in long axial field-of-view PET scanners
    Ghabrial, Amir
    Franklin, Daniel R.
    Zaidi, Habib
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2021, 31 (03): : 305 - 315
  • [9] Evaluation of cost-effective system designs for long axial field-of-view PET scanners
    Surti, Suleman
    Werner, Matthew E.
    Karp, Joel S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (10):
  • [10] Long-axial field-of-view PET/CT scanners improve the detection of inflamed coronary artery plaques with 68Ga-DOTA-TOC imaging
    Mingels, Clemens
    Alberts, Ian
    Rominger, Axel
    Caobelli, Federico
    JOURNAL OF NUCLEAR MEDICINE, 2023, 64