pH Gradient as a Tool for the Separation of Ionizable Analytes in Reversed-Phase High-Performance Chromatography

被引:25
|
作者
Wiczling, Pawel [1 ]
Kaliszan, Roman [1 ]
机构
[1] Med Univ Gdansk, Dept Biopharmaceut & Pharmacodynam, PL-80416 Gdansk, Poland
关键词
LIQUID-CHROMATOGRAPHY; PEAK COMPRESSION; RP-HPLC; ELUTION; PK(A); RETENTION;
D O I
10.1021/ac100023z
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The aim of this work was to propose a general scheme of optimizing separation of ionizable analytes and to determine conditions of maximal peak compression in pH-gradient reversed-phase high-performance liquid chromatography (RP HPLC). The approximated explicit equation of the linear pH gradient has been developed. It allows predicting retention times for a given organic modifier content, initial value of pH, and the start and steepness of the pH gradient. Also the formula for calculating maximal peak compression is provided. The developed theory was compared with experimental data on the example of a weak acid (ketoprofen) and a weak base (papaverine). Five parameters characterizing analyte retention (log k(w) and S of the ionized and nonionized forms along with pK(a,chrom)) were determined in a series of isocratic experiments carried out at different pH values and with different methanol contents in the eluent. Next, a series of pH gradients of different pH-gradient steepness and of different pH-gradient starting time has been obtained and used to test the validity of our theoretical approach. The conditions of maximal peak width compression have been found. The derived theory was proved to be in a good agreement with the experimental data. The pH-gradient method led to peak compression of up to 0.2, and minimized peak tailing was obtained for the tested analytes. Since the majority of analytical separations are done in an isocratic mode we proposed a means to transfer an isocratic method to a pH-gradient method.
引用
收藏
页码:3692 / 3698
页数:7
相关论文
共 50 条
  • [1] Gradient reversed-phase high-performance chromatography of ionogenic analytes
    Kaliszan, Roman
    Wiczling, Pawel
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2011, 30 (09) : 1372 - 1381
  • [2] pH gradient reversed-phase liquid chromatography as a fractionation tool for the separation of peptides
    Baczek, Tomasz
    Walijewski, Lukasz
    Kaliszan, Roman
    TALANTA, 2008, 75 (01) : 76 - 82
  • [3] pH-Gradient Reversed-Phase Liquid Chromatography of Ionogenic Analytes Revisited
    Nikitas, P.
    Pappa-Louisi, A.
    Zisi, Ch.
    ANALYTICAL CHEMISTRY, 2012, 84 (15) : 6611 - 6618
  • [4] The application of gradient reversed-phase high-performance liquid chromatography to the pKa and log kw determination of polyprotic analytes
    Wiczling, Pawel
    Waszczuk-Jankowska, Malgorzata
    Markuszewski, Michal Jan
    Kaliszan, Roman
    JOURNAL OF CHROMATOGRAPHY A, 2008, 1214 (1-2) : 109 - 114
  • [5] SEPARATION OF CHLORINATED PHENOLS BY REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AT AN ALKALINE PH
    MCLEOD, HA
    LAVER, G
    JOURNAL OF CHROMATOGRAPHY, 1982, 244 (02): : 385 - 390
  • [6] Preparation and separation of metallobacteriochlorophyll a by reversed-phase high-performance liquid chromatography
    Inoue, H
    Nonomura, Y
    Ashizawa, A
    Tajima, I
    Nomura, M
    Nakamura, N
    Yoshioka, N
    CHROMATOGRAPHIA, 1996, 43 (7-8) : 361 - 364
  • [7] Retention behaviour of analytes in reversed-phase high-performance liquid chromatography-A review
    Ganesh, V
    Basuri, P. Poorna
    Sahini, K.
    Nalini, C. N.
    BIOMEDICAL CHROMATOGRAPHY, 2023, 37 (07)
  • [8] SEPARATION OF DESULFOGLUCOSINOLATES BY REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    MINCHINTON, I
    SANG, J
    BURKE, D
    TRUSCOTT, RJW
    JOURNAL OF CHROMATOGRAPHY, 1982, 247 (01): : 141 - 148
  • [9] SEPARATION OF PEPTIDES BY REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    HERMODSON, M
    MAHONEY, WC
    METHODS IN ENZYMOLOGY, 1983, 91 : 352 - 359
  • [10] SEPARATION OF FLAVONOIDS BY REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    VADECASTEELE, K
    GEIGER, H
    VANSUMERE, CF
    JOURNAL OF CHROMATOGRAPHY, 1982, 240 (01): : 81 - 94