Quantitative sample injection for capillary electrophoresis

被引:20
|
作者
Yao, CY
Gao, RY
Yan, C
机构
[1] Unimicro Technol Inc, Pleasanton, CA 94566 USA
[2] Nankai Univ, Inst Elementoorgan Chem, State Key Lab, Tianjin 300071, Peoples R China
[3] Chinese Acad Sci, Dalian Inst Chem Phys, Dept Biotechnol, Dalian 116011, Peoples R China
来源
JOURNAL OF SEPARATION SCIENCE | 2003年 / 26卷 / 1-2期
关键词
quantitative sample injection; capillary electrophoresis; capillary electrochromatography; pressurized CEC; micro-HPLC;
D O I
10.1002/jssc.200390011
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An apparatus including a rotary-type injector was designed for quantitative sample injection in capillary electrophoresis (CE), in which both pressurized flow and electroosmotic flow were used to drive the background electrolyte solution. A relative standard deviation of peak area of lower than 1% was achieved by using this apparatus. The effects of back-pressure regulator, restrictor, and applied voltage on separation efficiency and resolution were investigated. The utility of this apparatus in both micro-HPLC and pressurized capillary electrochromatography (pCEC) was also demonstrated.