Analysis of quantitative [I-123] mIBG SPECT/CT in a phantom and in patients with neuroblastoma

被引:11
|
作者
Brady, Samuel L. [1 ]
Shulkin, Barry L. [2 ]
机构
[1] Univ Cincinnati, Cincinnati Childrens Hosp Med Ctr, Dept Radiol, Cincinnati, OH 45229 USA
[2] St Jude Childrens Res Hosp, Dept Diagnost Imaging, MS 220,262 Danny Thomas Pl, Memphis, TN 38105 USA
关键词
mIBG; Nuclear medicine; Neuroblastoma; Pediatrics; SCATTER CORRECTION; MONTE-CARLO; ATTENUATION; ACCURACY; COMPENSATION; VALIDATION; RECOVERY;
D O I
10.1186/s40658-019-0267-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To determine the accuracy of quantitative SPECT, intersystem and interpatient standardized uptake value (SUV) calculation consistency for a manufacturer-independent quantitative SPECT/CT reconstruction algorithm, and the range of SUVs of normal and neoplastic tissue. Methods A NEMA body phantom with 6 spheres (ranging 10-37 mm) was filled with a known activity-to-volume ratio and used to determine the contrast recovery coefficient (CRC) for each visible sphere, and the measured SUV accuracy of those spheres and background water solution. One hundred eleven 123I-metaiodobenzylguanidine ([I-123] mIBG) SPECT/CT examinations from 43 patients were reconstructed using SUV SPECT (R) (HERMES Medical Solutions Inc.); 42 examinations were acquired using a GE Infinia Hawkeye 4 SPECT/CT, and 69 were acquired on a Siemens Symbia Intevo SPECT/CT. Inter scanner SUV analysis of 9 regions of normal [I-123] mIBG tissue uptake was conducted. Intrapatient mean SUV variability was calculated by measuring normal liver uptake within patients scanned on both cameras. The intensity of uptake by neoplastic tissue in the images was quantified using maximum SUV and, if present, compared over time. Results The phantom results of the visible spheres and background resulted in accuracy calculations better than 5-10% with CRC correction. Interscanner SUV variability showed no statistical difference (average p value 0.559; range 0.066-1.0) among the 9 normal tissues analyzed. Intrapatient liver mean SUV varied <= 16% as calculated for 28 patients (87 examinations) studied on both scanners. In one patient, a thoracic tumor evaluated over 10 time points (18 months) underwent a 74% (3.1/12.0) reduction in maximum SUV with treatment. Conclusion The results demonstrate quantitative accuracy to better than 10%, and both consistent SUV calculation between 2 different SPECT/CT scanners for 9 tissues, and low intrapatient measurement variability for quantitative SPECT/CT analysis in a pediatric population with neuroblastoma. Quantitative SPECT/CT offers the opportunity for objective analysis of tumor response using [I-123] mIBG by normalizing the uptake to injected dose and patient weight, as is done for PET.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Analysis of quantitative [I-123] mIBG SPECT/CT in a phantom and in patients with neuroblastoma
    Samuel L. Brady
    Barry L. Shulkin
    [J]. EJNMMI Physics, 6
  • [2] Quantitative 123I-mIBG SPECT/CT in patients with neuroblastoma
    Brady, Samuel
    Shulkin, Barry
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [3] Additional value of I-123 MIBG SPECT in neuroblastoma
    Pirson, AS
    Krug, B
    Tuerlinckx, D
    Lacrosse, M
    Luyx, D
    Vander Borght, T
    [J]. CLINICAL NUCLEAR MEDICINE, 2005, 30 (02) : 100 - 101
  • [4] Characterization of adrenal activity in I-123 MIBG SPECT for pediatric neuroblastoma
    Kim, Sunhee
    Abdalla, Wael
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [5] SPECT/CT Prevents False Positive Interpretations of I-123 MIBG Scans in Children with Neuroblastoma
    Bar-Sever, Z.
    Ash, S.
    Yaniv, I.
    Kornreich, L.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 : S363 - S363
  • [6] SPECT-CT of I-123 in a head phantom
    Galt, James
    Aarsvold, John
    Fitz, James
    Margeanu, David
    Sabiston, Elizabeth
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [7] Quantitative I-123 mIBG SPECT in differentiating abnormal and normal mIBG myocardial uptake
    Ji Chen
    Russell D. Folks
    Liudmila Verdes
    Daya N. Manatunga
    Arnold F. Jacobson
    Ernest V. Garcia
    [J]. Journal of Nuclear Cardiology, 2012, 19 (1) : 92 - 99
  • [8] FDG PET-CT and I-123 MIBG in the evaluation of neuroblastoma
    Win, Sandra
    Goris, Michael
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [9] Pseudoneuroblastoma demonstrated by I-123 MIBG SPECT
    Okuyama, C
    Ushijima, Y
    Nakamura, T
    Nishimura, T
    [J]. CLINICAL NUCLEAR MEDICINE, 2001, 26 (05) : 474 - 475
  • [10] Post surgical 123I-MIBG SPECT/CT in neuroblastoma
    Wartski, Myriam
    Jehanno, Nina
    Cellier, Cecile
    Brisse, Herve
    Alberini, Jean Louis
    Petras, Slavomir
    Schleiermacher, Gudrun
    Michon, Jean