Real-Time Visualization of a Focused Ultrasound Beam Using Ultrasonic Backscatter

被引:5
|
作者
Thies, Miles [1 ]
Oelze, Michael L. [1 ]
机构
[1] Univ Illinois, Beckman Inst Adv Sci & Technol, Dept Elect & Comp Engn, Urbana, IL 61820 USA
关键词
Medical treatment; Imaging; Real-time systems; Ultrasonic imaging; Monitoring; Visualization; Acoustics; Beam visualization; focused ultrasound (FUS); real-time; therapy monitoring; COHERENCE FACTOR; CAVITATION; HIFU; THERAPY; THALAMOTOMY; SCATTERING; SURGERY; FIELDS; ARRAY;
D O I
10.1109/TUFFC.2020.3035784
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Focused ultrasound (FUS) therapies induce therapeutic effects in localized tissues using either temperature elevations or mechanical stresses caused by an ultrasound wave. During an FUS therapy, it is crucial to continuously monitor the position of the FUS beam in order to correct for tissue motion and keep the focus within the target region. Toward the goal of achieving real-time monitoring for FUS therapies, we have developed a method for the real-time visualization of an FUS beam using ultrasonic backscatter. The intensity field of an FUS beam was reconstructed using backscatter from an FUS pulse received by an imaging array and then overlaid onto a B-mode image captured using the same imaging array. The FUS beam visualization allows one to monitor the position and extent of the FUS beam in the context of the surrounding medium. Variations in the scattering properties of the medium were corrected in the FUS beam reconstruction by normalizing based on the echogenicity of the coaligned B-mode image. On average, normalizing by echogenicity reduced the mean square error between FUS beam reconstructions in nonhomogeneous regions of a phantom and baseline homogeneous regions by 21.61. FUS beam visualizations were achieved, using a single diagnostic imaging array as both an FUS source and an imaging probe, in a tissue-mimicking phantom and a rat tumor in vivo with a frame rate of 25-30 frames/s.
引用
收藏
页码:1213 / 1223
页数:11
相关论文
共 50 条
  • [1] Real-Time Visualization of a Focused Ultrasound Beam Using Ultrasonic Backscatter for Monitoring of Mechanical-Based Therapies
    Thies, Miles
    Oelze, Michael L.
    [J]. PROCEEDINGS OF THE 2020 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2020,
  • [2] Real-time visualization of high-intensity focused ultrasound treatment using ultrasound imaging
    Vaezy, S
    Shi, XG
    Martin, RW
    Chi, E
    Nelson, PI
    Bailey, MR
    Crum, LA
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (01): : 33 - 42
  • [3] Monitoring external beam radiotherapy using real-time beam visualization
    Jenkins, Cesare H.
    Naczynski, Dominik J.
    Yu, Shu-Jung S.
    Xing, Lei
    [J]. MEDICAL PHYSICS, 2015, 42 (01) : 5 - 13
  • [4] Ultrasonic backscatter from elongated grains using line focused ultrasound
    Kube, Christopher M.
    Arguelles, Andrea P.
    Turner, Joseph A.
    [J]. ULTRASONICS, 2018, 82 : 79 - 83
  • [5] ULTRASONIC TISSUE CHARACTERIZATION BY ANALYSIS OF INTEGRATED BACKSCATTER IN REAL-TIME
    THOMAS, LJ
    WICKLINE, SA
    PEREZ, JE
    MILLER, JG
    [J]. CIRCULATION, 1984, 70 (04) : 405 - 405
  • [6] Optimized Schlieren imaging for real-time visualization of high-intensity focused ultrasound waves
    Nawrath, Helena
    Holl, Andreas
    Bedarf, Lennart
    Mai, Denis
    Alfke, Heiko
    Groebner, Jens
    [J]. BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2024,
  • [7] Real-time ultrasound-guided high-intensity focused ultrasound therapy with enhanced visualization and reduced treatment time
    Sikdar, S
    Zderic, V
    Managuli, R
    Vaezy, S
    Kim, Y
    [J]. 4th International Symposium on Therapeutic Ultrasound, 2005, 754 : 47 - 51
  • [8] QUANTITATIVE REAL-TIME IMAGING OF MYOCARDIUM BASED ON ULTRASONIC INTEGRATED BACKSCATTER
    THOMAS, LJ
    BARZILAI, B
    PEREZ, JE
    SOBEL, BE
    WICKLINE, SA
    MILLER, JG
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1989, 36 (04) : 466 - 470
  • [9] THE USE OF REAL-TIME ULTRASOUND TO ENHANCE FETOSCOPIC VISUALIZATION
    SCHWARTZ, DB
    ZWEIBEL, WJ
    ZAGZEBSKI, J
    ARBOGAST, RL
    DONOVAN, D
    [J]. JOURNAL OF CLINICAL ULTRASOUND, 1983, 11 (03) : 161 - 164
  • [10] QUANTITATIVE REAL-TIME SCHLIEREN SYSTEM FOR ULTRASOUND VISUALIZATION
    PORTER, B