Rapid sealing of porcine renal blood vessels, ex vivo, using a high power, 1470-nm laser, and laparoscopic prototype

被引:16
|
作者
Hardy, Luke A. [1 ]
Hutchens, Thomas C. [1 ]
Larson, Eric R. [2 ]
Gonzalez, David A. [1 ]
Chang, Chun-Hung [1 ]
Nau, William H. [3 ]
Fried, Nathaniel M. [1 ]
机构
[1] Univ North Carolina Charlotte, Dept Phys & Opt Sci, Charlotte, NC 28223 USA
[2] Medtronic, R&D, Boulder, CO USA
[3] Medtronic, Surg Innovat Grp, Boulder, CO USA
关键词
arteries; blood vessels; burst pressure; coagulation; infrared; laparoscopic; laser; SURGICAL ENERGY DEVICES; VAGINAL HYSTERECTOMY; HARMONIC SCALPEL; RANDOMIZED-TRIAL; LIGASURE; SYSTEM; LIGATION; SPLENECTOMY; SPREAD; CLIPS;
D O I
10.1117/1.JBO.22.5.058002
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Energy-based, radiofrequency (RF) and ultrasonic (US) devices currently provide rapid sealing of blood vessels during laparoscopic procedures. We are exploring infrared lasers as an alternate energy modality for vessel sealing, capable of generating less collateral thermal damage. Previous studies demonstrated feasibility of sealing vessels in an in vivo porcine model using a 1470-nm laser. However, the initial prototype was designed for testing in open surgery and featured tissue clasping and light delivery mechanisms incompatible with laparoscopic surgery. In this study, a laparoscopic prototype similar to devices currently in surgical use was developed, and performance tests were conducted on porcine renal blood vessels, ex vivo. The 5-mm outer-diameter laparoscopic prototype featured a traditional Maryland jaw configuration that enables tissue manipulation and blunt dissection. Laser energy was delivered through a 550-ae m-core-diameter optical fiber with side-delivery from the lower jaw and beam dimensions of 18-mm length x 1.2-mm width. The 1470-nm diode laser delivered 68 W with 3-s activation time, consistent with vessel seal times associated with RF and US-based devices. A total of 69 fresh porcine renal vessels with mean diameter of 3.3 +/- 1.7 mm were tested, ex vivo. Vessels smaller than 5-mm diameter were consistently sealed (48/51) with burst pressures greater than malignant hypertension blood pressure (180 mmHg), averaging 1038 +/- 474 mmHg. Vessels larger than 5 mm were not consistently sealed (6/18), yielding burst pressures of only 174 +/- 221 mmHg. Seal width, thermal damage zone, and thermal spread averaged 1.7 +/- 0.8, 3.4 +/- 0.7, and 1.0 +/- 0.4 mm, respectively. Results demonstrated that the 5-mm optical laparoscopic prototype consistently sealed vessels less than 5-mm diameter with low thermal spread. Further in vivo studies are planned to test the performance across a variety of vessels and tissues. (C) 2017 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:7
相关论文
共 19 条
  • [1] Rapid sealing and cutting of porcine blood vessels, ex vivo, using a high-power, 1470-nm diode laser
    Giglio, Nicholas C.
    Hutchens, Thomas C.
    Perkins, William C.
    Latimer, Cassandra
    Ward, Arlen
    Nau, William H.
    Fried, Nathaniel M.
    JOURNAL OF BIOMEDICAL OPTICS, 2014, 19 (03)
  • [2] Simultaneous sealing and bisection of porcine renal blood vessels, ex vivo, using a continuous-wave, infrared diode laser at 1470 nm
    Saeed, Woheeb M.
    Yoshino, Jude K.
    Traynham, Alexandria J.
    Fried, Nathaniel M.
    LASERS IN MEDICAL SCIENCE, 2024, 39 (01)
  • [3] Tissue Dissection using a 1470-nm Diode Laser and Laparoscopic Prototype
    Chang, Chun-Hung
    Hammerland, John
    Nau, William H.
    Fried, Nathaniel M.
    ENERGY-BASED TREATMENT OF TISSUE AND ASSESSMENT IX, 2017, 10066
  • [4] Rapid Infrared Laser Sealing and Cutting of Porcine Renal Vessels, Ex Vivo
    Giglio, Nicholas C.
    Hutchens, Thomas C.
    Perkins, William C.
    Latimer, Cassandra
    Ward, Arlen
    Nau, William H.
    Fried, Nathaniel M.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS X, 2014, 8926
  • [5] Laparoscopic Prototype for Optical Sealing of Renal Blood Vessels
    Hardy, Luke A.
    Hutchens, Thomas C.
    Larson, Eric R.
    Gonzalez, David A.
    Chang, Chun-Hung
    Nau, William H.
    Fried, Nathaniel M.
    THERAPEUTICS AND DIAGNOSTICS IN UROLOGY: LASERS, ROBOTICS, MINIMALLY INVASIVE, AND ADVANCED BIOMEDICAL DEVICES, 2017, 10038
  • [6] Design of control system for 1470-nm high-power semiconductor laser lipolysis device
    Dong N.-N.
    Cui J.-J.
    Xu J.-G.
    Xu J.
    Cui, Jin-Jiang (cuijj@sibet.ac.cn), 1896, Chinese Academy of Sciences (26): : 1896 - 1903
  • [7] Retroperitoneoscopic partial nephrectomy using a 980/1470-nm dual-diode laser for small exophytic renal tumors
    Jie Dong
    Weifeng Xu
    Guanghua Liu
    Yi Xie
    Yi Qiao
    Zhigang Ji
    Lasers in Medical Science, 2022, 37 : 471 - 477
  • [8] Retroperitoneoscopic partial nephrectomy using a 980/1470-nm dual-diode laser for small exophytic renal tumors
    Dong, Jie
    Xu, Weifeng
    Liu, Guanghua
    Xie, Yi
    Qiao, Yi
    Ji, Zhigang
    LASERS IN MEDICAL SCIENCE, 2022, 37 (01) : 471 - 477
  • [9] Infrared Laser Sealing of Porcine Vascular Tissues using a 1,470nm Diode Laser: Preliminary In Vivo Studies
    Cilip, Christopher M.
    Kerr, Duane
    Latimer, Cassandra A.
    Rosenbury, Sarah B.
    Giglio, Nicholas C.
    Hutchens, Thomas C.
    Nau, William H.
    Fried, Nathaniel M.
    LASERS IN SURGERY AND MEDICINE, 2017, 49 (04) : 366 - 371
  • [10] High-power diode laser at 980 nm for the treatment of benign prostatic hyperplasia: ex vivo investigations on porcine kidneys and human cadaver prostates
    Michael Seitz
    Oliver Reich
    Christian Gratzke
    Boris Schlenker
    Alexander Karl
    Markus Bader
    Wael Khoder
    Florian Fischer
    Christian Stief
    Ronald Sroka
    Lasers in Medical Science, 2009, 24 : 172 - 178