共 50 条
Evaluation of screening and commercial methods for detection of methicillin resistance in coagulase-negative staphylococci
被引:25
|作者:
Hussain, Z
Stoakes, L
Lannigan, R
Longo, S
Nancekivell, B
机构:
[1] Univ Western Ontario, London Hlth Sci Ctr, Dept Microbiol, London, ON, Canada
[2] Univ Western Ontario, Dept Microbiol & Immunol, London, ON, Canada
关键词:
D O I:
10.1128/JCM.36.1.273-274.1998
中图分类号:
Q93 [微生物学];
学科分类号:
071005 ;
100705 ;
摘要:
The National Committee for Clinical Laboratory Standards recommends 48 h of incubation by the oxacillin salt agar screen (OSAS) method for the detection of methicillin-resistant coagulase-negative staphylococci (CoNS). An earlier identification of methicillin resistance is desirable. The time to detection of the mecA gene by PCR was compared with the times to detection by OSAS, by the oxacillin disk diffusion (ODD) method, and with MicroScan Gram Positive Combo type 6 panels (MicroScan Inc. Sacramento, Calif.) and Vitek GPS-SA cards (bioMerieux Vitek Inc., Hazelwood, Mo.). The combination of the Vitek card and the ODD method detected 92 of 99 methicillin-resistant strains of CoNS at 24 h; however, 6 mecA-positive strains were phenotypically methicillin susceptible. We conclude that most methicillin-resistant CoNS can;be detected and the results can be reported after overnight incubation by a combination of methods.
引用
收藏
页码:273 / 274
页数:2
相关论文