Predicted effects of hemoglobin A1c assay precision on a patient population distribution of serial hemoglobin A1c difference values

被引:0
|
作者
Stickle, Douglas F. [1 ]
Seligman, Mark L.
Landmark, James D.
Quon, Michael J.
机构
[1] Univ Nebraska, Med Ctr, Dept Pathol & Microbiol, Omaha, NE 68198 USA
[2] Univ Washington, Dept Stat, Seattle, WA 98195 USA
[3] NIH, Natl Ctr Complementary & Alternat Med, Bethesda, MD 20892 USA
关键词
glycated hemoglobin; hemoglobin A1c; diabetes; precision; point-of-care testing;
D O I
10.1016/j.cca.2006.11.022
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Interpretation of serial measurements of % hemoglobin A1c includes an assessment of differences from preceding values (DHbA1c). We examined predicted effects of different assay precisions on an observed population distribution for DHbA1c. Methods: Primary data were 5260 DHbA1c values from sequential HbA1c measurement pairs obtained within 1 calendar year. Each DHbA1c was replaced by a distribution obtained from sampling each component HbA1c value according to a normal distribution characterized by a fixed coefficient of variation (CV) of either 1%, 3% or 5% (forming data sets A, B and C, respectively). Data sets B and C, with inferior precision, were compared with the reference data set A (highest precision). Results: Using DHbA1c bin widths of 0.5% HbA1c, differences in assay precision caused significant redistribution of numbers within bins. For instance, for CV=5%, there was a 7.2% decrease in the number of results within the DHbA1c bin=(-0.5 to <= 0.0)% compared with the number for CV=1%, and a 6.4% increase in numbers of results for DHbA1c > 0.5. Conclusion: Different HbA1c assay CVs can significantly affect the fraction of patients within different clinical categorizations for DHbA1c and consequently may differently influence patient care recommendations. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:201 / 205
页数:5
相关论文
共 50 条
  • [1] Predicted effects of assay precision on a population distribution of delta values for serial measurements of hemoglobin A1c
    Stickle, D. F.
    Seligman, M. L.
    Landmark, J. D.
    Quon, M. J.
    CLINICAL CHEMISTRY, 2006, 52 (06) : A99 - A100
  • [2] Interpretation of Hemoglobin A1c Values
    Sacks, David B.
    John, W. Garry
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2014, 311 (22): : 2271 - 2272
  • [3] Standardization of hemoglobin A1c values
    Jain, M
    McCall, A
    Riddle, M
    ANNALS OF INTERNAL MEDICINE, 2000, 132 (08) : 676 - 676
  • [4] Hemoglobin A1c Values and CGM
    Blevins, Thomas
    DIABETES TECHNOLOGY & THERAPEUTICS, 2012, 14 (10) : 971 - 971
  • [5] A Patient with Undetectable Hemoglobin A1c
    Chakraborty, Sutirtha
    Gupta, Debkishore
    CLINICAL CHEMISTRY, 2013, 59 (05) : 856 - 857
  • [6] Hemoglobin A1c Values and CGM Response
    Pepper, Gary M.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2012, 14 (10) : 972 - 972
  • [7] Evaluation of an immunoturbidimetric assay for hemoglobin A1c
    Akan, P
    EUROMEDLAB 2003: PROCEEDINGS OF THE 15TH IFCC-FESCC EUROPEAN CONGRESS OF CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2003, : 187 - 193
  • [8] Evaluation of an Enzymatic Hemoglobin A1C Assay
    Majeed, Numan
    Kanani, Fatima
    Zubairi, Adnan Mustafa
    ANNALS OF KING EDWARD MEDICAL UNIVERSITY LAHORE PAKISTAN, 2024, 30 (02): : 184 - 190
  • [9] A Patient With Discordant Hemoglobin A1c Results
    Sterling, Brent
    Griffith, Michelle L.
    Shajani-Yi, Zahra
    CLINICAL CHEMISTRY, 2020, 66 (07) : 990 - 991
  • [10] A Patient with an Unexpectedly Low Hemoglobin A1c
    Dillon, Jessica L.
    Cervinski, Mark A.
    CLINICAL CHEMISTRY, 2018, 64 (08) : 1263 - 1264