Ion-optics calculations of the LLNL AMS system for biochemical 14C measurements

被引:11
|
作者
Ognibene, TJ [1 ]
Brown, TA [1 ]
Knezovich, JP [1 ]
Roberts, ML [1 ]
Southon, JR [1 ]
Vogel, JS [1 ]
机构
[1] Univ Calif Lawrence Livermore Natl Lab, Ctr Accelerator Mass Spectrometry, Livermore, CA 94551 USA
关键词
ion-optics; computer modeling; accelerator mass spectrometry;
D O I
10.1016/S0168-583X(00)00151-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A dedicated AMS system for biochemical C-14 measurements is being built at the Center for Accelerator Mass Spectrometry. The system is centered around a National Electrostatics Corporation Model 3SDH-1 1-MV Pelletron accelerator and is designed to accept two ion sources. The LLNL 64-sample Cs-sputter ion source with its zoom lens beam line is attached to one port of the electrostatic switching element. To insure efficient coupling of the source to the acceptance of the accelerator, ion-optics calculations of the low-energy injection beam line have been conducted; the results of which were used to determine the layout of the ion-optical components of the beam line. Beam tests of the low-energy injection line show that beam behavior was accurately predicted by the calculations. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:47 / 51
页数:5
相关论文
共 50 条
  • [1] LLNL AMS spectrometer for biochemical 14C measurements.
    Ognibene, TJ
    Bench, GS
    Brown, TA
    Peaslee, GF
    Southon, JR
    Vogel, JS
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2001, 222 : U8 - U8
  • [2] Automated evaluation of 14C AMS measurements
    Puchegger, S
    Rom, W
    Steier, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2000, 172 : 274 - 280
  • [3] The LLNL accelerator mass spectrometry system for biochemical 14C-measurements
    Ognibene, TJ
    Bench, G
    Brown, TA
    Vogel, JS
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2004, 223 : 12 - 15
  • [4] Calculation of the compounded uncertainty of 14C AMS measurements
    Nadeau, Marie-Josee
    Grootes, Pieter M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 294 : 420 - 425
  • [5] AMS measurements of 14C for human bone fossils
    Zhao, Qiang
    Liu, Kexin
    Guo, Zhiyu
    Li, Kun
    Yuan, Sixun
    Wang, Jianjun
    Gao, Shijun
    Lu, Xiangyang
    Li, Bin
    Yuan, Jinglin
    Chen, Chiaerh
    He Jishu/Nuclear Techniques, 20 (10): : 595 - 597
  • [6] Application of AMS 14C measurements to criminal investigations
    Nakamura, T.
    Kojima, S.
    Ohta, T.
    Nishida, M.
    Rakowski, A.
    Ikeda, A.
    Oda, H.
    Niu, E.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2007, 272 (02) : 327 - 332
  • [7] Application of AMS 14C measurements to criminal investigations
    T. Nakamura
    S. Kojima
    T. Ohta
    M. Nishida
    A. Rakowski
    A. Ikeda
    H. Oda
    E. Niu
    Journal of Radioanalytical and Nuclear Chemistry, 2007, 272 : 327 - 332
  • [8] The 14C AMS system at the University of Georgia
    Roberts, ML
    Culp, RA
    Dvoracek, DK
    Hodgins, GWL
    Neary, MP
    Noakes, JE
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2004, 223 : 1 - 4
  • [9] Development of an automatic gas handling system for microscale AMS 14C measurements
    Uhl, Thomas
    Luppold, W.
    Rottenbach, A.
    Scharf, A.
    Kritzler, K.
    Kretschmer, W.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2007, 259 (01): : 303 - 307
  • [10] Improving a gas ion source for 14C AMS
    Fahrni, S. M.
    Wacker, L.
    Synal, H-A
    Szidat, S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 294 : 320 - 327