Reproducibility of an intraoral scanner: A comparison between in-vivo and ex-vivo scans

被引:43
|
作者
Sun, LiJun [1 ]
Lee, Jin-Sol [1 ]
Choo, Hyun-Hee [1 ]
Hwang, Hyeon-Shik [2 ]
Lee, Kyung-Min [2 ]
机构
[1] Chonnam Natl Univ, Sch Dent, Gwangju, South Korea
[2] Chonnam Natl Univ, Sch Dent, Dept Orthodont, 33 Yongbong Ro, Gwangju 61186, South Korea
基金
新加坡国家研究基金会;
关键词
DENTAL IMPRESSIONS; DIGITAL MODELS; RELIABILITY; ACCURACY; VALIDITY;
D O I
10.1016/j.ajodo.2017.09.022
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The purpose of this study was to assess the reproducibility of in-vivo and ex-vivo scans using an intraoral scanner. Methods: Twenty adults with no missing teeth except for third molars were included in the study. Alginate impressions were taken, and plaster models were made from the impressions. Each subject underwent full-arch intraoral scanning twice with a TRIOS scanner (3Shape, Copenhagen, Denmark) at an interval of 2 weeks, and, the plaster models were scanned at the same interval with the same scanner. The first images of each scan were superimposed on the second scanned images using surface-based registration. In each case, the differences between the 2 scanned images were evaluated with color mapping. The reproducibility between the in-vivo and ex-vivo scans was compared using independent t tests and Bland-Altman analysis. Results: The discrepancies between the first and second images were greater in the posterior than in the anterior regions for both the in-vivo and ex-vivo scans. Average surface differences between the first and second images were greater for the in-vivo scans (0.04 mm) than for the ex-vivo scans (0.02 mm). The Bland-Altman plots showed that the reproducibility of both scans was within the limits of agreement. Conclusions: The reproducibility of in-vivo scanning was comparable with ex-vivo scanning, although it showed a slight difference (0.02 mm) compared with ex-vivo scanning.
引用
收藏
页码:305 / 310
页数:6
相关论文
共 50 条
  • [1] COMPARISON OF FUNCTION OF EX-VIVO AND IN-VIVO LIVERS
    CUSCHIERI, A
    BAKER, PR
    SHIELDS, R
    BRITISH JOURNAL OF SURGERY, 1972, 59 (11) : 901 - +
  • [2] Ex-vivo systems for neuromodulation: A comparison of ex-vivo and in-vivo large animal nerve electrophysiology
    Ribeiro, Mafalda
    Andreis, Felipe R.
    Jabban, Leen
    Nielsen, Thomas G. N. dS.
    Smirnov, Sergey, V
    Lutteroth, Christof
    Proulx, Michael J.
    Rocha, Paulo R. F.
    Metcalfe, Benjamin
    JOURNAL OF NEUROSCIENCE METHODS, 2024, 406
  • [3] Comparison of in-vivo and ex-vivo Dielectric Properties of Biological Tissues
    Salahuddin, Saqib
    La Gioia, Alessandra
    Elahi, Muhammad Adnan
    Porter, Emily
    O'Halloran, Martin
    Shahzad, Atif
    2017 INTERNATIONAL CONFERENCE ON ELECTROMAGNETICS IN ADVANCED APPLICATIONS (ICEAA), 2017, : 582 - 585
  • [4] An overview on the veracity of intraoral digital scanning system and utilization of iTero scanner for analyzing orthodontic study models both In-Vivo and Ex-Vivo
    Jabri, M. A.
    Wu, S.
    Pan, Y.
    Wang, L.
    NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2021, 24 (01) : 1 - 7
  • [5] Analysis Of Dosage Effects On The Comparison Of In-Vivo And Fixed Ex-Vivo Airway Measurements From Computed Tomography Scans
    Judisch, A. L.
    Stoyles, N.
    Hammond, E.
    Sloan, C.
    Hoffman, E. A.
    Sieren, J. C.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2015, 191
  • [6] Comparison of Virtual Intersection and Occlusal Contacts between Intraoral and Laboratory Scans: An In-Vivo Study
    Beck, Florian
    Lettner, Stefan
    Cepic, Lana Zupancic
    Schedle, Andreas
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (03)
  • [7] ASSESSMENT OF HEPATOBILIARY FUNCTION IN-VIVO AND EX-VIVO IN THE RAT
    HANDLER, JA
    KOSSOR, DC
    GOLDSTEIN, RS
    JOURNAL OF PHARMACOLOGICAL AND TOXICOLOGICAL METHODS, 1994, 31 (01) : 11 - 19
  • [8] Atypical fibroxanthoma: in-vivo and ex-vivo confocal features
    Pampena, Riccardo
    Peccerillo, Francesca
    Piana, Simonetta
    Paolino, Giovanni
    Mercuri, Santo R.
    Pellacani, Giovanni
    Longo, Caterina
    ITALIAN JOURNAL OF DERMATOLOGY AND VENEREOLOGY, 2021, 156 (06) : 41 - 42
  • [9] Thermal ablation of lung tissue: Comparison of RF and LASER ablation in-vivo and ex-vivo
    Wacker, FK
    Nour, SG
    Knappe, V
    Duerk, JL
    Lewin, JS
    RADIOLOGY, 2002, 225 : 291 - 291
  • [10] Fetoscopic Panorama Reconstruction: Moving from Ex-vivo to In-vivo
    Gaisser, Floris
    Peeters, Suzanne H. P.
    Lenseigne, Boris
    Jonker, Pieter P.
    Oepkes, Dick
    MEDICAL IMAGE UNDERSTANDING AND ANALYSIS (MIUA 2017), 2017, 723 : 581 - 593