Evaluation of the Vitek Ms system for the identification of filamentous fungi

被引:7
|
作者
Zvezdanova, Margarita Estreya [1 ,2 ]
Escribano, Pilar [1 ,2 ]
Guinea, Jesus [1 ,2 ,3 ]
Munoz, Patricia [1 ,2 ,3 ,4 ]
Rodriguez-Temporal, David [1 ,2 ]
Rodriguez-Sanchez, Belen [1 ,2 ]
机构
[1] Hosp Gen Univ Gregorio Maranon, Clin Microbiol & Infect Dis Dept, Madrid 28007, Spain
[2] Inst Invest Sanitaria Gregorio Maranon IiSGM, Madrid 28007, Spain
[3] CIBER Enfermedades Resp CIBERES CB06 06, Madrid 28007, Spain
[4] Univ Complutense Madrid, Sch Med, Med Dept, Madrid 28040, Spain
关键词
DESORPTION IONIZATION-TIME; FLIGHT MASS-SPECTROMETRY; SCEDOSPORIUM; PERFORMANCE;
D O I
10.1093/mmy/myac027
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The implementation of matrix-assisted laser desorption/ionization-time of flight mass spectrometry (MALDI-TOF MS) for the identification of fungal isolates remains challenging and has been limited to experienced laboratories in sample preparation and in-house libraries construction. However, the development of commercial kits for standardized fungal sample preparation and updated reference libraries can fill this gap. This study aimed to evaluate the performance of the commercial VITEK MS Mould Kit (bioMerieux, Marcy l'Etoile, France) and the VITEK MS system (bioMerieux) for identification using a panel of fungal species of clinical interest. Overall, 200 isolates belonging to 13 genera and 43 fungal species were analyzed with the VITEK MS system equipped with the v3.2 IVD database. Overall, 89.0% of the isolates were correctly identified, 41.5 and 43.5% at species and complex level, respectively. For an additional 4.0% of the identifications, correlation at the genus level was reported. The remaining 21 isolates (10.5%) could not be identified among which 85.0% (18/21) were species not claimed in the database. One Syncephalastrum isolate was misidentified as Rhizopus microsporus complex. Specifically, 100% of the Scedosporium/Lomentospora, 97.1% of the Fusarium, 65.7% of the Mucorales and 86.4% of the Aspergillus isolates were correctly identified at the species and complex level. The methodology described allows for an easy implementation of MALDI-TOF MS for routine identification of fungal species in a fast and reliable manner. Although further improvement in the databases is still required, an important number of fungal species can be correctly identified at the species level using this method.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Multicenter Evaluation of the Vitek MS v3.0 System for the Identification of Filamentous Fungi
    Rychert, Jenna
    Slechta, E. Sue
    Barker, Adam P.
    Miranda, Edwin
    Babady, N. Esther
    Tang, Yi-Wei
    Gibas, Connie
    Wiederhold, Nathan
    Sutton, DeAnna
    Hanson, Kimberly E.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2018, 56 (02)
  • [2] Vitek® MS v3.0 System in the Identification of Filamentous Fungi
    Pinheiro, Dolores
    Monteiro, Carolina
    Faria, Miguel A.
    Pinto, Eugenia
    [J]. MYCOPATHOLOGIA, 2019, 184 (05) : 645 - 651
  • [3] Vitek® MS v3.0 System in the Identification of Filamentous Fungi
    Dolores Pinheiro
    Carolina Monteiro
    Miguel A. Faria
    Eugénia Pinto
    [J]. Mycopathologia, 2019, 184 : 645 - 651
  • [4] Performance Evaluation of VITEK MS for the Identification of a Wide Spectrum of Clinically Relevant Filamentous Fungi Using a Korean Collection
    Shin, Ju Hyeon
    Kim, Soo Hyun
    Lee, Dain
    Lee, Seung Yeob
    Chun, Sejong
    Lee, Jun Hyung
    Won, Eun Jeong
    Choi, Hyun Jung
    Choi, Hyun Woo
    Kee, Seung Jung
    Shin, Myung Geun
    Shin, Jong Hee
    [J]. ANNALS OF LABORATORY MEDICINE, 2021, 41 (02) : 214 - 220
  • [5] Clinical Evaluation of VITEK MS PRIME with PICKME Pen for Bacteria and Yeasts, and RUO Database for Filamentous Fungi
    Lee, Hyeyoung
    Koo, Jehyun
    Oh, Junsang
    Cho, Sung-Il
    Lee, Hyunjoo
    Lee, Hyun Ji
    Sung, Gi-Ho
    Kim, Jayoung
    [J]. MICROORGANISMS, 2024, 12 (05)
  • [6] Evaluation of the Vitek MS Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry System for Identification of Clinically Relevant Filamentous Fungi
    McMullen, Allison R.
    Wallace, Meghan A.
    Pincus, David H.
    Wilkey, Kathy
    Burnham, Carey-Ann D.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2016, 54 (08) : 2068 - 2073
  • [7] Evaluation of MALDI-TOF VITEK®MS and VITEK® 2 system for the identification of Staphylococcus saprophyticus
    Martins, Katheryne B.
    Ferreira, Adriano M.
    Mondelli, Alessandro L.
    Rocchetti, Taisa T.
    da Cunha, Maria de L. R. de S.
    [J]. FUTURE MICROBIOLOGY, 2018, 13 (14) : 1603 - 1610
  • [8] Evaluation of bioMerieux VITEK® MS MALDI-TOF MS System for identification of dermatophytes
    Chew, K. L.
    Cheng, J. W. S.
    Achik, R.
    Lee, S. Y.
    Osman, N. H.
    Jureen, R.
    [J]. MEDICAL MYCOLOGY, 2018, 56 : S151 - S151
  • [9] Evaluation of species-specific PCR, Bruker MS, VITEK MS and the VITEK 2 system for the identification of clinical Enterococcus isolates
    Fang, H.
    Ohlsson, A. -K.
    Ullberg, M.
    Ozenci, V.
    [J]. EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2012, 31 (11) : 3073 - 3077
  • [10] Evaluation of species-specific PCR, Bruker MS, VITEK MS and the VITEK 2 system for the identification of clinical Enterococcus isolates
    H. Fang
    A.-K. Ohlsson
    M. Ullberg
    V. Özenci
    [J]. European Journal of Clinical Microbiology & Infectious Diseases, 2012, 31 : 3073 - 3077