Feasibility of a quality assurance programme of bedside blood glucose testing in a hospital setting: 7 years' experience

被引:11
|
作者
Nobels, F
Beckers, F
Bailleul, E
De Schrijver, P
Sierens, L
Van Crombrugge, P
机构
[1] Onze Lieve Vrouw Hosp, Dept Endocrinol, B-9300 Aalst, Belgium
[2] Onze Lieve Vrouw Hosp, Dept Clin Chem, B-9300 Aalst, Belgium
关键词
analysis; blood glucose; diabetes mellitus; point-of-care systems; quality assurance;
D O I
10.1111/j.1464-5491.2004.01333.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims Description of the long-term use of a quality assurance (QA) programme for bedside capillary blood glucose (CBG) testing in a general hospital. Methods The main points of the programme were selection of instrumentation using a standardized testing procedure, design and implementation of a quality control (QC) procedure, and training and motivation of nurses. The QC procedures consisted of a daily aqueous control and a weekly split-sample control (a measurement on a capillary blood sample using the glucose meter and on a simultaneously drawn venous sample with the laboratory analyser). When the result was out of range, a laboratory technician visited the ward to investigate the problem. All wards received a weekly report. Results The programme was introduced in 1995 and followed up through 2002. The split-sample control was more efficient in detecting clinically relevant errors than the aqueous control. Most errors (91-97%) were operator-related rather than instrument-related. The compliance with the split-sample controls remained high, with 84-91% of weekly controls performed over 7 years. Respectively 91, 95 and 96% of the measurements remained within the range of +/-20% of the laboratory value in the years 2000, 2001 and 2002. Conclusion A QA programme of bedside CBG testing can successfully be implemented. It is feasible to obtain a satisfying level of measurement accuracy and to maintain a high level of compliance with the programme over several years. Split-sample controls are an essential part of the control procedure.
引用
收藏
页码:1288 / 1291
页数:4
相关论文
共 21 条
  • [1] Testing blood glucose at the bedside in a chronic care hospital
    Greendyke, RM
    Gifford, FR
    [J]. LABORATORY MEDICINE, 1997, 28 (01) : 63 - 67
  • [2] LABORATORY RESPONSIBILITIES IN THE QUALITY ASSURANCE OF A HOSPITAL BEDSIDE GLUCOSE MONITORING PROGRAM
    CORDY, PA
    KUBASIK, NP
    DSOUZA, JP
    MAYER, TK
    [J]. CLINICAL CHEMISTRY, 1990, 36 (06) : 1006 - 1007
  • [4] Setting up a quality assurance programme at the medical emergency department of a university hospital: promises and limitations
    Wasserfallen, JB
    Moeschler, O
    Yersin, B
    [J]. SOZIAL-UND PRAVENTIVMEDIZIN, 2001, 46 (02): : 131 - 133
  • [5] EVALUATION OF THE QUALITY OF BEDSIDE MONITORING OF THE BLOOD-GLUCOSE LEVEL IN A TEACHING HOSPITAL
    TING, C
    NANJI, AA
    [J]. CANADIAN MEDICAL ASSOCIATION JOURNAL, 1988, 138 (01) : 23 - 26
  • [6] Setting up a quality system for management of blood glucose monitoring systems at hospital
    Prevosto, JM
    Astier, H
    Renard, C
    Cheminel, V
    Goudable, J
    Chaulet, JF
    [J]. ANNALES DE BIOLOGIE CLINIQUE, 1998, 56 (05) : 606 - 611
  • [7] Use of bedside capillary blood glucose data to measure the quality of in-hospital diabetes care
    Katz, G
    Linekin, PL
    Newman, LG
    Levy, CJ
    Roman, SH
    [J]. DIABETES, 1996, 45 : 533 - 533
  • [8] Improvements in Blood Glucose Re-Testing after Hypoglycemia in a Medical University Hospital Setting
    Hermayer, Kathie L.
    Scheurer, Danielle B.
    Arnold, Pamela C.
    Merrill, Mary-Eliese
    Zhu, Yusheng
    Epps, Juanita A.
    Strickland-Hedgpeth, Angela B.
    Hutto, Amy
    Wilson, Amy R.
    Robertson, Dawn
    Cawley, Patrick J.
    [J]. DIABETES, 2013, 62 : A105 - A105
  • [9] 3-YEAR EXPERIENCE WITH A QUALITY-CONTROL PROGRAM FOR BEDSIDE GLUCOSE TESTING WITHOUT LABORATORY COMPARISONS
    LEPAGE, R
    BESNER, G
    THIBAULT, L
    VINCENT, C
    [J]. CLINICAL CHEMISTRY, 1990, 36 (06) : 1006 - 1006
  • [10] REVISED GUIDELINES FOR COMPATIBILITY TESTING IN HOSPITAL BLOOD-BANKS - IMPORTANT CHANGES FOR QUALITY ASSURANCE AND PROCEDURES
    VOAK, D
    WATERS, AH
    [J]. CLINICAL AND LABORATORY HAEMATOLOGY, 1990, 12 (02): : 235 - 236