Evaluation and validation of tungsten fiducial marker-based image-guided radiotherapy

被引:5
|
作者
Shahid, Wajeehah [1 ]
Mukhtar, Raheel [1 ]
Rizvi, Syed Faheem Askari [2 ]
Shahid, Samiah [3 ]
Iqbal, Muhammad Aamir [4 ]
机构
[1] Univ Lahore, Dept Phys, Lahore 54000, Punjab, Pakistan
[2] Lanzhou Univ, Coll Chem & Chem Engn, Lanzhou 730000, Gansu, Peoples R China
[3] Univ Lahore, Inst Mol Biol & Biotechnol, Lahore 54000, Punjab, Pakistan
[4] Zhejiang Univ, Sch Mat Sci & Engn, Hangzhou, Zhejiang, Peoples R China
来源
关键词
tungsten fiducial marker; eclipse dosimeter software; styrofoam phantom; image-guided radiotherapy (IGRT); MV portal imager;
D O I
10.1088/2057-1976/abf90b
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this research work, a simple homemade cubic phantom was designed to validate the Image-Guided Radiotherapy (IGRT) set up and verified with the help of tungsten fiducial markers (size 2-3 mm) inserted into the cubic phantom. Phantom made up of Styrofoam, was scanned with the help of 16 slice Toshiba CT scanner where each slice was of 1 mm thickness and HU level set to -1000. A radio-opaque contrast medium was rubbed on the phantom to visualize the scanner images. Once the iso-center had been marked on a phantom with the help of in-room positioning laser and the fields (RT-LAT and AP) were applied on the contoured body of the phantom in Varian's ARIA-11 Eclipse dosimeter software, the same position of the phantom was reproduced on Varian's Linear Accelerator DHX. Known shifts of 3.0 to 30.0 mm from the marked iso-center were applied on the phantom by moving the couch in all six directions one by one. On each applied couch shift, an x-ray image of the phantom was acquired with the help of an MV portal imager of Linac in AP and RT-LAT direction. This image was superimposed with a reference image of phantom and shift accuracy calculated by ARIA-11 software was noted down. It turned out that irrespective of the position of the phantom on the couch, the calculated corrected shift and deviation from reference position was always between +/- 1-2 mm which is the required accuracy for IGRT according to International Atomic Energy Agency (IAEA). This process was repeated 40 times and each time, the corrected shift came out to be +/- 1-2 mm. We can conclude that our system is safe and accurate enough to perfectly position the actual patient for IGRT.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Optimization of fiducial marker configurations in image-guided radiotherapy
    Liu, H
    Yu, Y
    Ruo, R
    Okunieff, P
    O'Dell, W
    Schell, M
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1240 - 1241
  • [2] Verification of optimal fiducial marker configurations in image-guided radiotherapy
    Liu, H
    Ruo, R
    Schell, MC
    Okunieff, P
    O'Dell, W
    Yu, Y
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1241 - 1241
  • [3] Evaluation of a Novel Liquid Fiducial Marker, BioXmark®, for Small Animal Image-Guided Radiotherapy Applications
    Brown, Kathryn H.
    Ghita, Mihaela
    Schettino, Giuseppe
    Prise, Kevin M.
    Butterworth, Karl T.
    [J]. CANCERS, 2020, 12 (05)
  • [4] A new fiducial marker for Image-guided radiotherapy of prostate cancer: Clinical experience
    Carl, Jesper
    Nielsen, Jane
    Holmberg, Mats
    Larsen, Erik Hojkjaer
    Fabrin, Knud
    Fisker, Rune V.
    [J]. ACTA ONCOLOGICA, 2008, 47 (07) : 1358 - 1366
  • [5] Novel liquid fiducial tissue marker with potential use in image-guided radiotherapy
    Schaarup-Jensen, Henrik
    Jolck, Rasmus
    Hansen, Anders
    Jensen, Andreas
    Bruun, Linda
    Christensen, Anders
    Scherman, Per Bengtsson
    Kjaer, Andreas
    Clausen, Mads
    Andresen, Thomas
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2015, 250
  • [6] Evaluation of different fiducial markers for image-guided radiotherapy and particle therapy
    Habermehl, Daniel
    Henkner, Katrin
    Ecker, Swantje
    Jaekel, Oliver
    Debus, Juergen
    Combs, Stephanie E.
    [J]. JOURNAL OF RADIATION RESEARCH, 2013, 54 : 61 - 68
  • [7] A novel liquid fiducial marker for image-guided adaptive radiotherapy in bladder cancer.
    De Ridder, M.
    Gebrandy, L. C.
    De Reijke, T. M.
    Hinnen, K. A.
    Hulshof, M. C. C. M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S637 - S637
  • [8] Novel fiducial marker has optimal characteristics for image-guided radiotherapy of abdominal tumours
    Stefanowicz, S.
    Moeller, M. L.
    Thiele, J.
    Jaster, M.
    Hoinkis, C.
    Troost, E. G. C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S916 - S916
  • [9] In vitroandin vivoevaluation of the safety and efficacy of a novel liquid fiducial marker for image-guided radiotherapy
    Sun, Liang-Chao
    Su, Yi
    Ding, Xing-Chen
    Xu, Dong-Shui
    Li, Cheng-Ming
    Wang, Lu
    Li, Wan-Long
    Sun, Xin-Dong
    Yu, Jin-Ming
    Meng, Xue
    [J]. ONCOLOGY LETTERS, 2020, 20 (01) : 569 - 580
  • [10] Fiducial-based image-guided radiotherapy for whole breast irradiation
    Latifi, Kujtim
    Pritz, Jakub
    Zhang, Geoffrey G.
    Moros, Eduardo G.
    Harris, Eleanor E. R.
    [J]. JOURNAL OF RADIATION ONCOLOGY, 2013, 2 (02) : 185 - 190