摘要
Background: The periprosthetic femoral fracture is one of the most severe complications after total hip arthroplasty and is associated with an increased mortality. The underlying causes and the patient-and implant-specific risk factors of periprosthetic femoral fractures remain insufficiently understood. The aim of this study was to gain a more profound understanding of the underlying fracture mechanisms and to provide experimental datasets for validation of computational models. Methods: Six cadaveric femurs were implanted with straight hip stems (Zweymueller design) and loaded until fracture reproducing the clinically relevant load cases stumbling and sideways fall. Displacements and the strain distribution on the surface of the femurs and implants, as well as the fracture load and implant subsidence were measured. Findings: For the load case stumbling the mean fracture load was 6743 N and two different mechanisms leading to fracture could be identified: high subsidence with low femoral bending and small subsidence with high femoral bending. For the load case sideways fall the mean fracture load was 1757 N and both tested femurs fractured due to a rotation of the hip stem around its own axis. The detailed datasets provided by this study can be used in future computational models. Interpretation: We demonstrated that the underlying fracture mechanisms of periprosthetic femoral fractures can be fundamentally different in the load case stumbling. The seating and exact position of the hip stem in the femur may correlate with implant subsidence and therefore lead to different types of fracture mechanisms resulting in different patient-specific fracture risks.
引用
下载
收藏
页数:9
相关论文
共 50 条
  • [1] Acute Periprosthetic Hip Fractures With Short, Uncemented Femoral Stems
    Staunton, Peter
    Alhojailan, Khalifa
    Desgagne, Charles
    Epure, Laura
    Zukor, David
    Huk, Olga
    Antoniou, John
    JOURNAL OF ARTHROPLASTY, 2024, 39 (09): : S248 - S253
  • [2] Breaking Bad: A Comparative Descriptive Analysis of Periprosthetic Fractures Around Cemented and Uncemented Femoral Stems
    Fenelon, Christopher
    Murphy, Evelyn P.
    Baig, Muhammad N.
    Kearns, Stephen R.
    Murphy, Colin G.
    Curtin, William
    JOURNAL OF ARTHROPLASTY, 2019, 34 (08): : 1783 - 1786
  • [3] Letter to the Editor on "Breaking Bad: A Comparative Descriptive Analysis of Periprosthetic Fractures Around Cemented and Uncemented Femoral Stems"
    Hu, Lintao
    Zhang, Xianzuo
    Zhou, Huajun
    Shang, Xifu
    JOURNAL OF ARTHROPLASTY, 2020, 35 (01): : 301 - 302
  • [4] Higher periprosthetic fracture rate associated with use of modern uncemented stems compared to cemented stems in femoral neck fractures
    Song, Jin Soo A.
    Dillman, Daryl
    Wilson, Dave
    Dunbar, Michael
    Richardson, Glen
    HIP INTERNATIONAL, 2019, 29 (02) : 177 - 183
  • [5] OSTEOPOROSIS AND THE INCIDENCE OF PERIPROSTHETIC FEMORAL FRACTURES AFTER UNCEMENTED PRIMARY TOTAL HIP ARTHROPLASTY
    Ene, R.
    Nica, M.
    Panti, Z.
    Ene, P.
    Cirstoiu, M.
    Cirstoiu, C.
    OSTEOPOROSIS INTERNATIONAL, 2015, 26 : S449 - S449
  • [6] Response to Letter to the Editor on "Breaking Bad: A Comparative Descriptive Analysis of Periprosthetic Fractures Around Cemented and Uncemented Femoral Stems"
    Fenelon, Christopher
    Murphy, Evelyn P.
    Baig, Muhammad N.
    Kearns, Stephen R.
    Murphy, Colin G.
    Curtin, William
    JOURNAL OF ARTHROPLASTY, 2020, 35 (01): : 302 - 302
  • [7] Treatment of periprosthetic femoral fractures with modular stems
    Daniel Hernandez-Vaquero
    Jesus Fernandez-Lombardia
    Jimena Llorens de los Rios
    Ivan Perez-Coto
    Susana Iglesias-Fernandez
    International Orthopaedics, 2015, 39 : 1933 - 1938
  • [8] Increased risk for early periprosthetic fractures after uncemented total hip replacement
    Solgaard, Soren
    Kjersgaard, Anne Grete
    DANISH MEDICAL JOURNAL, 2014, 61 (02):
  • [9] Treatment of periprosthetic femoral fractures with modular stems
    Hernandez-Vaquero, Daniel
    Fernandez-Lombardia, Jesus
    Llorens de los Rios, Jimena
    Perez-Coto, Ivan
    Iglesias-Fernandez, Susana
    INTERNATIONAL ORTHOPAEDICS, 2015, 39 (10) : 1933 - 1938
  • [10] Uncemented Collared Femoral Stems in Total Hip Arthroplasty
    Lemme, Nicholas J.
    McDonald, Christopher L.
    Hamilton, William G.
    Crisco, Joseph J.
    Cohen, Eric M.
    ORTHOPEDICS, 2022, 45 (03) : E122 - +