Planar amperometric glucose sensor based on glucose oxidase immobilized by chitosan film on Prussian blue layer

被引:70
|
作者
Zhu, JH [1 ]
Zhu, ZQ
Lai, ZS
Wang, R
Guo, XM
Wu, XQ
Zhang, GX
Zhang, ZR
Zhang, ZR
Wang, YT
Chen, ZY
机构
[1] E China Normal Univ, IT Coll, Shanghai 200062, Peoples R China
[2] Shanghai Teachers Univ, Shanghai 2000234, Peoples R China
[3] Shanghai Inst Med Control, Shanghai 200233, Peoples R China
[4] Fudan Univ, Dept Surg, Huashan Hosp, Shanghai 200040, Peoples R China
来源
SENSORS | 2002年 / 2卷 / 04期
关键词
Prussian Blue; electrocatalysis; chitosan; selective permeability; glucose sensor;
D O I
10.3390/s20400127
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A planar amperometric glucose microsensor based on glucose oxidase immobilized by chitosan film on Prussian Blue layer has been developed. The experimental results show that the optimum detection potential is 50 mV ( versus Ag/AgCl) and the optimum pH is 6.5. Under the selective conditions the sensor exhibits excellent sensitivity of 98 nA/M and a linear range of 0.1-6.0 mM. The apparent Michanelis-Menten constant of the sensor is 21 mM. The response time is less than 60 seconds. No apparent change in the response to glucose was observed during one month. Foremost, the interference of ascorbic and uric acids can be avoided due to selective permeability of chitosan film and electrocatalysis of PB layer to H2O2. The sensor has been applied to detect glucose in human blood serum.
引用
收藏
页码:127 / 136
页数:10
相关论文
共 50 条
  • [1] Glucose Sensor Based on Glucose Oxidase-Lipid LB Film Immobilized in Prussian Blue Layer
    Lee, Dong-Yun
    Kafi, A. K. M.
    Choi, Won-Suk
    Park, Sang-Hyun
    Kwon, Young-Soo
    [J]. JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2008, 8 (09) : 4543 - 4547
  • [2] Amperometric Glucose Biosensor Based on Titanium Electrode Modified with Prussian Blue Layer and Immobilized Glucose Oxidase
    Valiuniene, A.
    Virbickas, P.
    Rekertaite, A.
    Ramanavicius, A.
    [J]. JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2017, 164 (14) : B781 - B784
  • [3] Amperometric glucose biosensors based on layer-by-layer assembly of chitosan and glucose oxidase on the Prussian blue-modified gold electrode
    Bing Yin
    Ruo Yuan
    Yaqing Chai
    Shihong Chen
    Shurui Cao
    Yang Xu
    Ping Fu
    [J]. Biotechnology Letters, 2008, 30 : 317 - 322
  • [4] Amperometric glucose biosensors based on layer-by-layer assembly of chitosan and glucose oxidase on the Prussian blue-modified gold electrode
    Yin, Bing
    Yuan, Ruo
    Chai, Yaqing
    Chen, Shihong
    Cao, Shurui
    Xu, Yang
    Fu, Ping
    [J]. BIOTECHNOLOGY LETTERS, 2008, 30 (02) : 317 - 322
  • [5] The Development of Reagentless Amperometric Glucose Biosensor Based on Gold Nanostructures, Prussian Blue and Glucose Oxidase
    Sakalauskiene, Laura
    Brasiunas, Benediktas
    Popov, Anton
    Kausaite-Minkstimiene, Asta
    Ramanaviciene, Almira
    [J]. BIOSENSORS-BASEL, 2023, 13 (10):
  • [6] Glucose biosensor based on glucose oxidase immobilized in sol–gel chitosan/silica hybrid composite film on Prussian blue modified glass carbon electrode
    Xue-Cai Tan
    Yuan-Xin Tian
    Pei-Xiang Cai
    Xiao-Yong Zou
    [J]. Analytical and Bioanalytical Chemistry, 2005, 381 : 500 - 507
  • [7] A glucose biosensor based on Prussian blue/chitosan hybrid film
    Wang, Xueying
    Gu, Haifang
    Yin, Fan
    Tu, Yifeng
    [J]. BIOSENSORS & BIOELECTRONICS, 2009, 24 (05): : 1527 - 1530
  • [8] Amperometric glucose sensor based on glucose oxidase immobilized in electrochemically generated poly(ethacridine)
    Xu, JJ
    Chen, HY
    [J]. ANALYTICA CHIMICA ACTA, 2000, 423 (01) : 101 - 106
  • [9] Amperometric glucose biosensors based on Prussian Blue- and polyaniline-glucose oxidase modified electrodes
    Garjonyte, R
    Malinauskas, A
    [J]. BIOSENSORS & BIOELECTRONICS, 2000, 15 (9-10): : 445 - 451
  • [10] Glucose biosensor based on glucose oxidase immobilized in sol-gel chitosan/silica hybrid composite film on Prussian blue modified glass carbon electrode
    Tan, XC
    Tian, YX
    Cai, PX
    Zou, XY
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2005, 381 (02) : 500 - 507