Comparison of the Technical and Clinical Performance of the Elecsys® HBsAg II Assay With the Architect®, AxSym®, and Advia® Centaur HBsAg Screening Assays

被引:35
|
作者
Louisirirotchanakul, S. [1 ]
Khupulsup, K. [2 ]
Akraekthalin, S. [3 ]
Chan, K. P. [4 ]
Saw, S. [5 ]
Aw, T. C. [6 ]
Cho, D-H. [7 ]
Shin, M-G. [8 ]
Lim, J. [9 ]
机构
[1] Siriraj Hosp, Bangkok, Thailand
[2] Ramathibodi Hosp, Bangkok, Thailand
[3] Phyathai Hosp, Bangkok, Thailand
[4] Singapore Gen Hosp, Singapore 0316, Singapore
[5] Natl Univ Singapore Hosp, Singapore, Singapore
[6] Changi Gen Hosp, Singapore, Singapore
[7] Cheil Gen Hosp, Seoul, South Korea
[8] Chonnam Natl Univ, Dept Lab Med, Sch Med, Hwasun Hosp, Hwasun, South Korea
[9] Catholic Univ Korea, Dept Lab Med, St Marys Hosp, Seoul, South Korea
关键词
hepatitis B virus; immunoassay; hepatitis B surface antigen mutants; HEPATITIS-B-VIRUS; SURFACE-ANTIGEN; GENETIC-VARIABILITY; DIAGNOSTIC KITS; S-GENE; DETERMINANT; INFECTION; VARIANTS; VACCINATION; PREVALENCE;
D O I
10.1002/jmv.21706
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
South East Asia has some of the highest prevalence rates of hepatitis B virus (HBV) infection (>= 8%) in the world, and the emergence of hepatitis B surface antigen (HBsAg) mutant strains is a growing problem. Assays with the highest levels of sensitivity, including mutant detection, should be used for routine HBsAg screening. In this large multicenter study, the clinical and technical performance of the fully automated Elecsys HBsAg II assay was compared with the Architect, AxSYM, and Advia Centaur HBsAg assays for HBsAg screening. Nine laboratories (three each from Thailand, Korea, and Singapore) compared the Elecsys HBsAg II assay with their routine HBsAg screening assay against a range of stored and routine clinical samples, including recombinant mutants. The Elecsys HBsAg II assay demonstrated equivalent sensitivity and specificity to the Architect HBsAg assay. However, the Elecsys HBsAg II assay recognized a native mutant sample (L94S, L97V, L98V, T123A) that the Architect HBsAg assay failed to detect. The AxSYM and Advia Centaur HBsAg assays appeared less sensitive for the detection of early HBV infection and also failed to detect some of the recombinant mutant strains. There was almost complete agreement between the Elecsys HBsAg II assay and comparator assays with respect to routine serum samples. The results of this study demonstrate that the Elecsys HBsAg II assay is a highly sensitive and specific screening assay for HBsAg and detects reliably the most important and clinically relevant HBV mutants and genotypes. It is suitable for routine HBsAg screening in Asia. J. Med. Virol. 82:755-762, 2010. (C) 2010 Wiley-Liss, Inc.
引用
收藏
页码:755 / 762
页数:8
相关论文
共 18 条
  • [1] A multicenter clinical evaluation of the ADVIA centaur® HBsAg and ADVIA Centaur® HBsAg confirmatory assays.
    Wians, FH
    Schappert, J
    Schiff, E
    Reamer, CR
    Preisel-Simmons, B
    [J]. CLINICAL CHEMISTRY, 2004, 50 (06) : A162 - A162
  • [2] HBsAg blood screening and diagnosis: performance evaluation of the ARCHITECT HBsAg qualitative and ARCHITECT HBsAg qualitative confirmatory assays
    Popp, Christian
    Krams, Doris
    Beckert, Christian
    Buenning, Carsten
    Queiros, Lucinda
    Piro, Loredana
    Luciani, Marina
    Roebbecke, Markus
    Kapprell, Hans-Peter
    [J]. DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2011, 70 (04) : 479 - 485
  • [3] Correlation between the Elecsys HBsAg II assay and the Architect assay for the quantification of hepatitis B surface antigen (HBsAg) in the serum
    Wursthorn, Karsten
    Jaroszewicz, Jerzy
    Zacher, Behrend J.
    Darnedde, Martina
    Raupach, Regina
    Mederacke, Ingmar
    Cornberg, Markus
    Manns, Michael P.
    Wedemeyer, Heiner
    [J]. JOURNAL OF CLINICAL VIROLOGY, 2011, 50 (04) : 292 - 296
  • [4] EVALUATION OF ARCHITECT HBSAG QUALITATIVE II ASSAY COMPARED WITH ARCHITECT HBSAG QUALITATIVE ASSAY FOR BLOOD DONATION SCREENING
    Oota, S.
    Intharasongkroh, D.
    Chaiwong, K.
    Sakuldamrongpanich, T.
    Phikulsod, S.
    [J]. VOX SANGUINIS, 2013, 105 : 176 - 177
  • [5] Analytical Performance of the Sysmex HISCL HBsAg Assay and Comparison with the Roche Elecsys HBsAg II Quant Assay in the Quantification of Hepatitis B Surface Antigen
    Park, Joonhong
    Bae, Taewon
    Cho, Yonggon
    Kim, Dalsik
    Lee, Jaehyeon
    [J]. MEDICINA-LITHUANIA, 2021, 57 (12):
  • [6] QUANTITATIVE ASSESSMENT OF SERUM HBSAG LEVEL USING THE ELECSYS® HBSAG II SCREENING ASSAY: RESULTS OF A FEASIBILITY STUDY
    Bonino, Ferruccio
    Ofenloch, Beatus
    Melchior, Walter
    Upmeier, Barbara
    Roessler, Dieter
    van der Helm, Wim
    [J]. HEPATOLOGY, 2008, 48 (04) : 726A - 726A
  • [7] Comparison of the elecsys HBsAg II assay and the architect assay for quantification of hepatitis B surface antigen in chronic hepatitis B patients
    Liao, Chen-Chih
    Hsu, Chao-Wei
    Gu, Po-Wen
    Yeh, Chau-Ting
    Lin, Shi-Ming
    Chiu, Cheng-Tang
    [J]. BIOMEDICAL JOURNAL, 2015, 38 (03) : 250 - 256
  • [8] Comparison between Elecsys HBsAg II and Architect HBsAg QT Assays for Quantification of Hepatitis B Surface Antigen among Patients Coinfected with HIV and Hepatitis B Virus
    Maylin, Sarah
    Boyd, Anders
    Delaugerre, Constance
    Zoulim, Fabien
    Lavocat, Fabien
    Simon, Francois
    Girard, Pierre-Marie
    Lacombe, Karine
    [J]. CLINICAL AND VACCINE IMMUNOLOGY, 2012, 19 (02) : 242 - 248
  • [9] Comparative study: Analytical and clinical performance of advia centaur troponin I - Ultra assay and elecsys troponin T assay
    Khoo, R.
    Kandiban, P.
    Saw, S.
    Sethi, S.
    [J]. CLINICAL CHEMISTRY, 2008, 54 (06) : A87 - A87
  • [10] Multicentre evaluation of the Elecsys® hepatitis B surface antigen II assay for detection of HBsAg in comparison with other commercially available assays
    Ji-Dong Jia
    Ma Hong
    Lai Wei
    Xin-Xin Zhang
    Yuan-Li Mao
    Lan-Lan Wang
    Zhi-Liang Gao
    Jin-Lin Hou
    Jun Zhang
    [J]. Medical Microbiology and Immunology, 2009, 198 : 263 - 269