Minimum variance beamforming combined with covariance matrix-based adaptive weighting for medical ultrasound imaging

被引:5
|
作者
Wang, Yuanguo [1 ]
Wang, Yadan [1 ]
Liu, Mingzhou [1 ]
Lan, Zhengfeng [2 ]
Zheng, Chichao [2 ]
Peng, Hu [2 ,3 ]
机构
[1] Hefei Univ Technol, Sch Mech Engn, Hefei 230009, Peoples R China
[2] Hefei Univ Technol, Dept Biomed Engn, Hefei 230009, Peoples R China
[3] Hefei Univ Technol, Anhui Prov Key Lab Measuring Theory & Precis Inst, Hefei 230009, Peoples R China
基金
中国博士后科学基金; 中国国家自然科学基金;
关键词
Minimum variance beamforming; Adaptive beamformer; Covariance matrix; Coherence; Standard deviation; LAG SPATIAL COHERENCE; ALGORITHM; GPU;
D O I
10.1186/s12938-022-01007-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: The minimum variance (MV) beamformer can significantly improve the image resolution in ultrasound imaging, but it has limited performance in noise reduction. We recently proposed the covariance matrix-based statistical beamforming (CMSB) for medical ultrasound imaging to reduce sidelobes and incoherent clutter. Methods: In this paper, we aim to improve the imaging performance of the MV beamformer by introducing a new pixel-based adaptive weighting approach based on CMSB, which is named as covariance matrix-based adaptive weighting (CMSAW). The proposed CMSAW estimates the mean-to-standard-deviation ratio (MSR) of a modified covariance matrix reconstructed by adaptive spatial smoothing, rotary averaging, and diagonal reducing. Moreover, adaptive diagonal reducing based on the aperture coherence is introduced in CMSAW to enhance the performance in speckle preservation. Results: The proposed CMSAW-weighted MV (CMSAW-MV) was validated through simulation, phantom experiments, and in vivo studies. The phantom experimental results show that CMSAW-MV obtains resolution improvement of 21.3% and simultaneously achieves average improvements of 96.4% and 71.8% in average contrast and generalized contrast-to-noise ratio (gCNR) for anechoic cyst, respectively, compared with MV. in vivo studies indicate that CMSAW-MV improves the noise reduction performance of MV beamformer. Conclusion: Simulation, experimental, and in vivo results all show that CMSAW-MV can improve resolution and suppress sidelobes and incoherent clutter and noise. These results demonstrate the effectiveness of CMSAW in improving the imaging performance of MV beamformer. Moreover, the proposed CMSAW with a computational complexity of O(N-2) has the potential to be implemented in real time using the graphics processing unit.
引用
收藏
页数:24
相关论文
共 50 条
  • [1] Minimum variance beamforming combined with covariance matrix-based adaptive weighting for medical ultrasound imaging
    Yuanguo Wang
    Yadan Wang
    Mingzhou Liu
    Zhengfeng Lan
    Chichao Zheng
    Hu Peng
    BioMedical Engineering OnLine, 21
  • [2] Minimum Variance Beamforming Combined with Adaptive Coherence Weighting Applied to Medical Ultrasound Imaging
    Asl, Babak Mohammadzadeh
    Mahloojifar, Ali
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2009, 56 (09) : 1923 - 1931
  • [3] Adaptive Diagonal Reducing and Adaptive Weighting Approach to Covariance Matrix-Based Statistical Beamforming for Ultrasound Imaging
    Wang, Yuanguo
    Wang, Yadan
    Pan, Jingwen
    Zheng, Chichao
    Peng, Hu
    2022 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IEEE IUS), 2022,
  • [4] Covariance Matrix-Based Statistical Beamforming for Medical Ultrasound Imaging
    Wang, Yuanguo
    Zheng, Chichao
    Liu, Mingzhou
    Peng, Hu
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2022, 69 (01) : 208 - 221
  • [5] Minimum variance adaptive beamforming applied to medical ultrasound imaging
    Synnevag, JF
    Austeng, A
    Holm, S
    2005 IEEE Ultrasonics Symposium, Vols 1-4, 2005, : 1199 - 1202
  • [6] Adaptive minimum variance beamforming combined with phase coherence imaging for ultrasound imaging
    Xu, Mengling
    Chen, Yimin
    Ding, Mingyue
    Ming Yuchi
    MEDICAL IMAGING 2012: ULTRASONIC IMAGING, TOMOGRAPHY, AND THERAPY, 2012, 8320
  • [7] Adaptive diagonal loaded minimum variance beamforming applied to medical ultrasound imaging
    刘昊霖
    张志宏
    刘东权
    Journal of Central South University, 2015, 22 (05) : 1826 - 1832
  • [8] Adaptive diagonal loaded minimum variance beamforming applied to medical ultrasound imaging
    Liu Hao-lin
    Zhang Zhi-hong
    Liu Dong-quan
    JOURNAL OF CENTRAL SOUTH UNIVERSITY, 2015, 22 (05) : 1826 - 1832
  • [9] Adaptive diagonal loaded minimum variance beamforming applied to medical ultrasound imaging
    Hao-lin Liu
    Zhi-hong Zhang
    Dong-quan Liu
    Journal of Central South University, 2015, 22 : 1826 - 1832
  • [10] Forward-backward Minimum Variance Beamforming Combined with Coherence Weighting Applied to Ultrasound Imaging
    ZHOU Ting
    LI Qiao-liang
    CHEN Xin
    WANG Tian-fu
    CHEN Si-ping