The value of the coefficient of variation in assessing repeat variation in ECG measurements

被引:15
|
作者
McLaughlin, SC
Aitchison, TC
Macfarlane, PW
机构
[1] Univ Glasgow, Dept Med Cardiol, Glasgow, Lanark, Scotland
[2] Univ Glasgow, Dept Stat, Glasgow, Lanark, Scotland
关键词
electrocardiogram; computer analysis; repeatability;
D O I
10.1053/euhj.1997.0741
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims The coefficient of variation is a popular measure for describing the amount of repeat variability present in ECG measurements from recording to recording. However, it can be misleading. The aim of the present study was to assess repeat variation (reclassification) in computer measured ECG criteria, i.e. positive to negative or vice versa, and compare this with the coefficient of variability. Methods and Results Two ECGs were obtained from each of 295 patients, one day apart, and separately from a further 364 patients, several minutes apart. All patients were considered to be in a stable condition. Estimates of the coefficients of variation were obtained for a number of ECG parameters used in the diagnosis of left ventricular hypertrophy. Corresponding reclassification rates of relevant ECG criteria were also calculated. Large coefficients of variation were observed in voltage parameters, e.g. R in V-5 (20% for day-to-day recordings and 6% for minute-to-minute recordings) while the corresponding reclassification rates were 8% and 0% respectively. The repeat variation in the diagnosis of left ventricular hypertrophy was up to 5% for day-to-day recordings and up to 3% for minute-to-minute recordings based on several different criteria. Conclusion A large coefficient of variation in a particular variable does not necessarily correspond to a high reclassification rate. A batter measure of the impact of ECG variability for a particular measurement is obtained from its reclassification rate. In turn, this may have a minimal effect on the overall diagnosis of a particular abnormality.
引用
收藏
页码:342 / 351
页数:10
相关论文
共 50 条
  • [1] Limitations of Assessing Repeatability of with Coefficient of Variation
    White, Elizabeth
    Choi, Dongseok
    Tan, Ou
    Huang, David
    Jia, Yali
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2024, 65 (07)
  • [2] Use of coefficient of variation in assessing variability of quantitative assays
    Reed, GF
    Lynn, F
    Meade, BD
    CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 2002, 9 (06) : 1235 - 1239
  • [3] Assessing Pain Using the Variation Coefficient of Pupillary Diameter
    Charier, David J.
    Zantour, Daniel
    Pichot, Vincent
    Chouchou, Florian
    Barthelemy, Jean-Claude M.
    Roche, Frederic
    Molliext, Serge B.
    JOURNAL OF PAIN, 2017, 18 (11): : 1346 - 1353
  • [4] ACCURACY AND BEAT-TO-BEAT VARIATION IN ECG COMPUTER MEASUREMENTS
    BOROVSKY, D
    ZYWIETZ, C
    ADVANCES IN CARDIOLOGY, 1976, 16 : 176 - 181
  • [5] THE COEFFICIENT OF VARIATION
    KESTEVEN, GL
    NATURE, 1946, 158 (4015) : 520 - 521
  • [6] Entropy and coefficient of variation (CV) as tools for assessing power quality
    Estrada, J. H.
    Cano-Plata, E. A.
    Younes-Velosa, C.
    Cortes, C. L.
    INGENIERIA E INVESTIGACION, 2011, 31 : 45 - 50
  • [8] A METHOD FOR ASSESSING LOCAL VARIATION AMONG SCATTERED MEASUREMENTS
    PHILIP, GM
    WATSON, DF
    MATHEMATICAL GEOLOGY, 1986, 18 (08): : 759 - 764
  • [9] Calculating the Coefficient of Variation from Duplicate Measurements: A New Method
    Asberg, Arne
    Bolann, Bjorn Johan
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2023, 8 (01): : 220 - 222
  • [10] DEFINING THE TARGET VALUE OF COEFFICIENT OF VARIATION IN CHINESE DIABETES PATIENTS
    Mo, Y.
    Ma, X.
    Lu, J.
    Zhang, L.
    Lu, W.
    Zhu, W.
    Bao, Y.
    Zhou, J.
    Jia, W.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2020, 22 : A215 - A215