EVALUATION OF THREE HANDHELD LACTATE METERS FOR MEASURING HEMOLYMPH L-LACTATE IN DECAPOD CRUSTACEANS

被引:4
|
作者
Bakke, Snorre [1 ]
Woll, Astrid K. [1 ]
机构
[1] Moreforsking AS, NO-6021 Alesund, Norway
来源
JOURNAL OF SHELLFISH RESEARCH | 2014年 / 33卷 / 01期
关键词
L-lactate; decapod crustaceans; handheld lactate meters; concordance; correlation; BLOOD LACTATE; PORTABLE LACTATE; BLUE-CRAB; CANCER-PAGURUS; METABOLIC DEPRESSION; PHYSIOLOGICAL STRESS; NEPHROPS-NORVEGICUS; LOCOMOTOR-ACTIVITY; OXYGEN-TRANSPORT; JASUS-EDWARDSII;
D O I
10.2983/035.033.0109
中图分类号
S9 [水产、渔业];
学科分类号
0908 ;
摘要
The aim of this study was to determine the reliability and accuracy of 3 commercially available handheld lactate meters when measuring hemolymph L-lactate in decapod crustaceans. The instruments tested were Accutrend Plus (AP), Lactate Scout+ (LS), and Lactate Plus (LP), and all models were tested in triplicate and compared with a commonly used enzymatic kit for L-lactate quantification (TB-kit). The Norway lobster Nephrops norwegicus and the European lobster Hommarus gammarus were used as model organisms, with emersion as an invoker of hemolymph L-lactate. Between-method comparisons were investigated by determining correlations (Pearson's r) and concordance (concordance correlation coefficient (CCC) and Bland-Altmand's 95% limit of agreement (95% LOA)) between handheld instruments and TB-kit. Within-instrument variation was tested by calculating the coefficient of variation (CV), and the average absolute deviation from the mean (AD). The within-instrument variation was low for all models (CV, 3.8-6.1%; AD, <0.3 mM), showing high reproducibility of technical replicates. A significant correlation with TB-kit was found for all handheld instruments (AP, r = 0.970; LS, r = 0.956; LP, r = 0.950). For both AP and LS, there was a moderate and significant concordance with TB-kit (AP, CCC = 0.945; LS, CCC = 0.940). The mean difference from TB-kit was -1.42 mM (95% LOA, -4.50-1.66 mM) for AP and 1.19 mM(95% LOA, -3.13-5.51 mM) for LP. Emersion experiments with the European lobster showed that AP and LS were in line with TB-kit in terms of detecting differences in L-lactate levels between groups of animals. As a result of the high frequency of error readings from LP, this instrument was not considered reliable when measuring L-lactate in decapod crustaceans. In conclusion the handheld instruments AP and LS are found to be reliable instruments when measuring hemolymph L-lactate in decapod crustaceans. Comparison of results from different methods should be avoided, and in cases when accurate absolute measurement of L-lactate is needed, methods allowing for greater accuracy and resolution should be used.
引用
收藏
页码:69 / 76
页数:8
相关论文
共 50 条
  • [1] Validation of handheld meters to measure blood L-lactate concentration in dairy cows and calves
    Burfeind, O.
    Heuwieser, W.
    JOURNAL OF DAIRY SCIENCE, 2012, 95 (11) : 6449 - 6456
  • [2] Validation of the handheld Lactate-Pro analyzer for measurement of blood L-lactate concentration in cattle
    Buczinski, Sebastien
    Dore, Elizabeth
    Boulay, Guillaume
    Francoz, David
    VETERINARY CLINICAL PATHOLOGY, 2014, 43 (04) : 567 - 572
  • [3] L-LACTATE AFFECTS CO2 TRANSPORT IN CRUSTACEAN HEMOLYMPH
    MORRIS, S
    TAYLOR, AC
    COMPARATIVE BIOCHEMISTRY AND PHYSIOLOGY A-PHYSIOLOGY, 1988, 91 (03): : 523 - 527
  • [4] L-lactate analysis
    不详
    ZUCKERINDUSTRIE, 2005, 130 (03): : 216 - 216
  • [5] l-Lactate treatment by photosynthetic cyanobacteria expressing heterogeneous l-lactate dehydrogenase
    Yuichi Kato
    Kosuke Inabe
    Yuji Haraguchi
    Tatsuya Shimizu
    Akihiko Kondo
    Tomohisa Hasunuma
    Scientific Reports, 13
  • [6] EVALUATION OF L-LACTATE IN DIGESTIVE SECRETIONS OF THE RABBIT
    VERNAY, M
    ANNALES DE ZOOTECHNIE, 1986, 35 (01): : 87 - 89
  • [7] l-Lactate treatment by photosynthetic cyanobacteria expressing heterogeneous l-lactate dehydrogenase
    Kato, Yuichi
    Inabe, Kosuke
    Haraguchi, Yuji
    Shimizu, Tatsuya
    Kondo, Akihiko
    Hasunuma, Tomohisa
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [8] Co-immobilized L-lactate oxidase and L-lactate dehydrogenase on a film mounted on oxygen electrode for highly sensitive L-lactate determination
    Casimiri, V
    Burstein, C
    BIOSENSORS & BIOELECTRONICS, 1996, 11 (08): : 783 - 789
  • [9] D-lactate and L-lactate emergency diagnosis
    Kleine, TO
    Benes, L
    XX WORLD CONGRESS OF PATHOLOGY AND LABORATORY MEDICINE, 1999, : 151 - 155
  • [10] L-lactate transport into rat heart mitochondria and reconstruction of the L-lactate/pyruvate shuttle
    Valenti, D
    de Bari, L
    Atlante, A
    Passarella, S
    BIOCHEMICAL JOURNAL, 2002, 364 (01) : 101 - 104