CBCT-Based Image Guidance for Percutaneous Access: Electromagnetic Navigation Versus 3D Image Fusion with Fluoroscopy Versus Combination of Both Technologies-A Phantom Study

被引:3
|
作者
Tacher, Vania [1 ,2 ,3 ]
Blain, Maxime [1 ]
Herin, Edouard [1 ,2 ]
Vitellius, Manuel [1 ]
Chiaradia, Melanie [1 ]
Oubaya, Nadia [4 ,5 ]
Derbel, Haytham [1 ,2 ,3 ]
Kobeiter, Hicham [1 ,2 ]
机构
[1] CHU Henri Mondor, AP HP, Serv Imagerie Med, 51 Ave Marechal de Lattre de Tassigny, F-94010 Creteil, France
[2] UPEC, F-94010 Creteil, France
[3] INSERM, U955 18, IMRB, Creteil, France
[4] Hop Henri Mondor, AP HP, Serv Sante Publ, Creteil, France
[5] Paris Est Univ, DHU A TVB, IMRB EA 7376 CEpiA Clin Epidemiol & Ageing Unit, UPEC, F-94000 Creteil, France
关键词
3D image guidance; CBCT; Percutaneous intervention; Electromagnetic navigation; Phantom study; Radiation exposure; Accuracy; CONE-BEAM CT; TRANSARTERIAL CHEMOEMBOLIZATION; ANGIOGRAPHY; BIOPSY; FEASIBILITY; ANEURYSM;
D O I
10.1007/s00270-019-02356-w
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose We set out to compare three types of three-dimensional CBCT-based imaging guidance modalities in a phantom study: image fusion with fluoroscopy (IF), electromagnetic navigation (EMN) and the association of both technologies (CEMNIF). Materials and Methods Four targets with a median diameter of 11 mm [first quartile (Q1): 10; third quartile (Q3): 12] with acute angle access (z-axis < 45 degrees) and four targets of 10 mm [8-15] with large angle access (z-axis > 45 degrees) were defined on an abdominal phantom (CIRS, Meditest, Tabuteau, France). Acute angle access targets were punctured using IF, EMN or CEMNIF and large angle access targets with EMN by four operators with various experiences. Efficacy (target reached), accuracy (distance between needle tip and target center), procedure time, radiation exposure and reproducibility were explored and compared. Results All targets were reached (100% efficacy) by all operators. For targets with acute angle access, procedure times (EMN: 265 s [236-360], IF: 292 s [260-345], CEMNIF: 320 s [240-333]) and accuracy (EMN: 3 mm [2-5], IF: 2 mm [1-3], CEMNIF: 3 mm [2-4]) were similar. Radiation exposure (EMN: 0; IF: 708 mGy.cm(2) [599-1128]; CEMNIF: 51 mGy.cm(2) [15-150]; p < 0.001) was significantly higher with IF than with CEMNIF and EMN. For targets with large angle access, procedure times (EMN: 345 s [259-457], CEMNIF: 425 s [340-473]; p = 0.01) and radiation exposure (EMN: 0, CEMIF: 159 mGy.cm(2) [39-316]; p < 0.001) were significantly lower with EMN but with lower accuracy (EMN: 4 mm [4-6] and CEMNIF: 4 mm [3, 4]; p = 0.01). The operator's experience did not impact the tested parameters regardless of the technique. Conclusion In this phantom study, EMN was not limited to acute angle targets. Efficacy and accuracy of puncture for acute angle access targets with EMN, IF or CEMNIF were similar. CEMNIF is more accurate for large angle access targets at the cost of a slightly higher procedure time and radiation exposure.
引用
收藏
页码:495 / 504
页数:10
相关论文
共 17 条
  • [1] CBCT-Based Image Guidance for Percutaneous Access: Electromagnetic Navigation Versus 3D Image Fusion with Fluoroscopy Versus Combination of Both Technologies—A Phantom Study
    Vania Tacher
    Maxime Blain
    Edouard Hérin
    Manuel Vitellius
    Mélanie Chiaradia
    Nadia Oubaya
    Haytham Derbel
    Hicham Kobeiter
    CardioVascular and Interventional Radiology, 2020, 43 : 495 - 504
  • [2] CBCT-based 3D MRA and angiographic image fusion and MRA image navigation for neuro interventions
    Zhang, Qiang
    Zhang, Zhiqiang
    Yang, Jiakang
    Sun, Qi
    Luo, Yongchun
    Shan, Tonghui
    Zhang, Hao
    Han, Jingfeng
    Liang, Chunyang
    Pan, Wenlong
    Gu, Chuanqi
    Mao, Gengsheng
    Xu, Ruxiang
    MEDICINE, 2016, 95 (32)
  • [3] Prostate localization: Fiducial marker versus cone beam CT (CBCT) 3D image fusion
    Hammoud, R. W.
    Pradhan, D.
    Kim, J.
    Patel, S. H.
    Kowalski, S.
    Guan, H.
    Xu, Y.
    Elshaikh, M.
    Ajlouni, M.
    Movsas, B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S679 - S680
  • [4] A Comparison of Accuracy of Image- versus Hardware-based Tracking Technologies in 3D Fusion in Aortic Endografting
    Rolls, A. E.
    Maurel, B.
    Davis, M.
    Constantinou, J.
    Hamilton, G.
    Mastracci, T. M.
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2016, 52 (03) : 323 - 331
  • [5] Comparison of Smartphone Augmented Reality, Smartglasses Augmented Reality, and 3D CBCT-guided Fluoroscopy Navigation for Percutaneous Needle Insertion: A Phantom Study
    Long, Dilara J.
    Li, Ming
    De Ruiter, Quirina M. B.
    Hecht, Rachel
    Li, Xiaobai
    Varble, Nicole
    Blain, Maxime
    Kassin, Michael T.
    Sharma, Karun V.
    Sarin, Shawn
    Krishnasamy, Venkatesh P.
    Pritchard, William F.
    Karanian, John W.
    Wood, Bradford J.
    Xu, Sheng
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2021, 44 (05) : 774 - 781
  • [6] Comparison of Smartphone Augmented Reality, Smartglasses Augmented Reality, and 3D CBCT-guided Fluoroscopy Navigation for Percutaneous Needle Insertion: A Phantom Study
    Dilara J. Long
    Ming Li
    Quirina M. B. De Ruiter
    Rachel Hecht
    Xiaobai Li
    Nicole Varble
    Maxime Blain
    Michael T. Kassin
    Karun V. Sharma
    Shawn Sarin
    Venkatesh P. Krishnasamy
    William F. Pritchard
    John W. Karanian
    Bradford J. Wood
    Sheng Xu
    CardioVascular and Interventional Radiology, 2021, 44 : 774 - 781
  • [8] Comparison of novel machine vision spinal image guidance system with existing 3D fluoroscopy-based navigation system: a randomized prospective study
    Malham, Gregory M.
    Munday, Nigel R.
    SPINE JOURNAL, 2022, 22 (04): : 561 - 569
  • [9] Comparison of image quality of 3D ultrasound: motorized acquisition versus freehand navigated acquisition, a phantom study
    N. M. Bekedam
    L. H. E. Karssemakers
    M. J. A. van Alphen
    R. L. P. van Veen
    L. E. Smeele
    M. B. Karakullukcu
    International Journal of Computer Assisted Radiology and Surgery, 2023, 18 : 1649 - 1663
  • [10] Comparison of image quality of 3D ultrasound: motorized acquisition versus freehand navigated acquisition, a phantom study
    Bekedam, N. M.
    Karssemakers, L. H. E.
    van Alphen, M. J. A.
    van Veen, R. L. P.
    Smeele, L. E.
    Karakullukcu, M. B.
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2023, 18 (09) : 1649 - 1663