Adaptive spatial calibration of a 3D ultrasound system

被引:12
|
作者
Hartov, Alex [1 ]
Paulsen, Keith [1 ]
Ji, Songbai [1 ]
Fontaine, Kathryn [1 ]
Furon, Marie-Laure [2 ]
Borsic, Andrea [1 ]
Roberts, David [3 ]
机构
[1] Dartmouth Coll, Hanover, NH 03766 USA
[2] ISAE ENSICA, F-31400 Toulouse, France
[3] Dartmouth Hitchcock Med Ctr, Lebanon, NH 03766 USA
关键词
biomedical ultrasonics; calibration; medical image processing; tracking; MULTIMODALITY IMAGE-REGISTRATION; FREEHAND ULTRASOUND; GUIDED NEUROSURGERY; TRACKING; ULTRASONOGRAPHY; COMPENSATION; EXPERIENCES; GUIDANCE; PROBE; CT;
D O I
10.1118/1.3373520
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methods: The method of calibration is based on images of a fixed plane of unknown location with respect to the 3D tracking system. This approach has, for advantage, to eliminate the measurement of the plane location as a source of error. The devised method is sufficiently general and adaptable to calibrate scanheads for 2D images and 3D volume sets using the same approach. The basic algorithm for both types of scanheads is the same and can be run unattended fully automatically once the data are collected. The approach was devised by seeking the simplest and most robust solutions for each of the steps required. These are the identification of the plane intersection within the images or volumes and the optimization method used to compute a calibration transformation matrix. The authors use adaptive algorithms in these two steps to eliminate data that would otherwise prevent the convergence of the procedure, which contributes to the robustness of the method. Results: The authors have run tests amounting to 57 runs of the calibration on two a scanhead that produce 3D imaging volumes, at all the available scales. The authors evaluated the system on two criteria: Robustness and accuracy. The program converged to useful values unattended for every one of the tests (100%). Its accuracy, based on the measured location of a reference plane, was estimated to be 0.7 +/- 0.6 mm for all tests combined. Conclusions: The system presented is robust and allows unattended computations of the calibration parameters required for freehand tracked ultrasound based on either 2D or 3D imaging systems.
引用
收藏
页码:2121 / 2130
页数:10
相关论文
共 50 条
  • [1] Calibration of 3D ultrasound to an electromagnetic tracking system
    Lang, Andrew
    Parthasarathy, Vijay
    Jain, Ameet
    [J]. MEDICAL IMAGING 2011: ULTRASONIC IMAGING, TOMOGRAPHY, AND THERAPY, 2011, 7968
  • [2] Spatial calibration of a 2D/3D ultrasound using a tracked needle
    Vasconcelos, Francisco
    Peebles, Donald
    Ourselin, Sebastien
    Stoyanov, Danail
    [J]. INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2016, 11 (06) : 1091 - 1099
  • [3] Spatial calibration of a 2D/3D ultrasound using a tracked needle
    Francisco Vasconcelos
    Donald Peebles
    Sebastien Ourselin
    Danail Stoyanov
    [J]. International Journal of Computer Assisted Radiology and Surgery, 2016, 11 : 1091 - 1099
  • [4] Spatial Calibration for 3D Freehand Ultrasound via Independent General Motions
    Cai, Qianqian
    Lu, Jian-yu
    Peng, Chang
    Prieto, Juan C.
    Rosenbaum, Alan J.
    Stringer, Jeffrey S. A.
    Jiang, Xiaoning
    [J]. PROCEEDINGS OF THE 2020 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2020,
  • [5] Actuator-Assisted Calibration of Freehand 3D Ultrasound System
    Koo, Terry K.
    Silvia, Nathaniel
    [J]. JOURNAL OF HEALTHCARE ENGINEERING, 2018, 2018
  • [6] Temporal and Spatial Calibration of a freehand 3D ultrasound reconstructions system by using an N-wire phantom
    Mathieu, Bielmann
    Claire, Closset
    Lorenzo, Leija
    Arturo, Vera
    [J]. 2015 12TH INTERNATIONAL CONFERENCE ON ELECTRICAL ENGINEERING, COMPUTING SCIENCE AND AUTOMATIC CONTROL (CCE 2015), 2015,
  • [7] A method for the calibration of 3D ultrasound transducers
    Hastenteufel, M
    Mottl-Link, S
    Wolf, I
    de Simone, R
    Meinzer, HP
    [J]. MEDICAL IMAGING 2003: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2003, 5029 : 231 - 238
  • [8] 3D ultrasound computed tomography system calibration using a neural network
    Wang, Hongjian
    Shen, Ning
    Lei, Xiaoxu
    Gemmeke, Hartmut
    Zapf, Michael
    Yu, Shoujian
    Xia, Xiaoling
    [J]. MEDICAL IMAGING 2022: ULTRASONIC IMAGING AND TOMOGRAPHY, 2022, 12038
  • [9] Automatic calibration of a robotized 3D ultrasound imaging system by visual servoing
    Krupa, Alexandre
    [J]. 2006 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), VOLS 1-10, 2006, : 4136 - 4141
  • [10] A temporal calibration method for freehand 3D ultrasound system: a preliminary result
    Nakamoto, M
    Sato, Y
    Nakada, K
    Nakajima, Y
    Konishi, K
    Hashizume, M
    Tamura, S
    [J]. CARS 2003: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2003, 1256 : 1365 - 1365