Comparison of glucose concentrations in canine whole blood, plasma, and serum measured with a veterinary point-of-care glucometer

被引:5
|
作者
Suchowersky, Natalie D. [1 ]
Carlson, Elizabeth A. [1 ]
Lee, Hollie P. [1 ]
Behrend, Ellen N. [1 ]
机构
[1] Auburn Univ, Coll Vet Med, Dept Clin Sci, 1220 Wire Rd, Auburn, AL 36849 USA
关键词
diabetes mellitus; dogs; error grid; portable blood glucometer; AUTOMATED BIOCHEMICAL ANALYZER; METERS; ACCURACY;
D O I
10.1177/10406387211019755
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Previous studies have determined that, compared to whole blood, serum or plasma used in a portable blood glucometer (PBG) may provide more accurate results. We investigated the accuracy of a veterinary PBG (AlphaTRAK 2; Zoetis) for the measurement of glucose concentrations in serum, plasma, and whole blood compared to plasma glucose concentration measured by a biochemical analyzer. Blood samples from 53 client-owned dogs were collected. Lin concordance correlation coefficient (rho(c)) and Bland-Altman plots were used to determine correlation and agreement between the results obtained for the different sample types. Glucose concentration in whole blood measured by the veterinary PBG was more strongly correlated with the glucose concentration measured by the biochemical analyzer (rho(c) = 0.92) compared to plasma and serum glucose concentrations (rho(c) = 0.59 and 0.57, respectively). The mean differences between the glucose concentrations in whole blood, plasma, and serum measured by the veterinary PBG and the glucose concentration determined by the biochemical analyzer were 1.0, 6.3, and 6.7 mmol/L (18, 113, and 121 mg/dL), respectively. Our findings suggest that, when using this veterinary PBG, the accuracy of a glucose measurement obtained is higher when using whole blood compared to plasma or serum. Use of whole blood allows for more correct assessment and diagnosis, which are necessary for appropriate therapeutic intervention.
引用
收藏
页码:695 / 702
页数:8
相关论文
共 50 条
  • [1] Comparison of glucose concentrations in serum, plasma, and blood measured by a point-of-care glucometer with serum glucose concentration measured by an automated biochemical analyzer for canine and feline blood samples
    Lechner, Matthew J.
    Hess, Rebecka S.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2019, 80 (12) : 1074 - 1081
  • [2] Correlation between glucose concentrations in serum, plasma, and whole blood measured by a point-of-care glucometer and serum glucose concentration measured by an automated biochemical analyzer for canine and feline blood samples
    Tauk, Barbara S.
    Drobatz, Kenneth J.
    Wallace, Koranda A.
    Hess, Rebecka S.
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2015, 246 (12): : 1327 - 1333
  • [3] Comparison of a point-of-care glucometer and an automated biochemical analyzer to measure glucose concentrations in whole blood, serum, and plasma from porcine blood samples.
    Gohlke, M. K.
    Nunnelley, W. Z.
    Brandebourg, T. D.
    JOURNAL OF ANIMAL SCIENCE, 2017, 95 : 157 - 158
  • [4] Formulation and validation of a predictive model to correct blood glucose concentrations obtained with a veterinary point-of-care glucometer in hemodiluted and hemoconcentrated canine blood samples
    Lane, Selena L.
    Koenig, Amie
    Brainard, Benjamin M.
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2015, 246 (03): : 307 - 312
  • [5] FORMULATION OF A PREDICTIVE MODEL TO CORRECT BLOOD GLUCOSE CONCENTRATIONS OBTAINED USING A VETERINARY POINT-OF-CARE GLUCOMETER IN ANEMIC AND HEMOCONCENTRATED CANINE SAMPLES
    Lane, S.
    Koenig, A.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2013, 27 (03) : 745 - 745
  • [6] Ammonia concentrations in canine whole blood, EDTA-anticoagulated whole blood, and plasma measured by use of a point-of-care ammonia meter
    Odunayo, Adesola
    Tobias, Karen M.
    Okafor, Chika C.
    Flatland, Bente
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2017, 78 (11) : 1239 - 1244
  • [7] Usefulness of whole blood, plasma, peritoneal fluid, and peritoneal fluid supernatant glucose concentrations obtained by a veterinary point-of-care glucometer to identify septic peritonitis in dogs with peritoneal effusion
    Koenig, Amie
    Verlander, Lindsey Lane
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2015, 247 (09): : 1027 - 1032
  • [8] A veterinary-calibrated point-of-care glucometer accurately measures blood glucose concentration in dogs and cats
    Cammarano, Kellie
    Carney, Patrick C.
    Rose, Hailey E.
    Merkhassine, Michael
    Ippolito, Elizabeth
    Arsenault, Anne
    Plotsker, Noah M.
    Loftus, John P.
    Kowalczyk, Marcin
    Prieto, Jennifer M. B.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2024, 85 (09)
  • [9] Comparison of a Point-of-Care Glucometer and a Laboratory Autoanalyzer for Measurement of Blood Glucose Concentrations in Domestic Pigeons (Columba livia domestica)
    Mohsenzadeh, Mahdieh Sadat
    Zaeemi, Mahdieh
    Razmyar, Jamshid
    Azizzadeh, Mohammad
    JOURNAL OF AVIAN MEDICINE AND SURGERY, 2015, 29 (03) : 181 - 186
  • [10] Comparison of 2 glucose analytical methodologies in immature Kemp's ridley sea turtles: dry chemistry of plasma versus point-of-care glucometer analysis of whole blood
    Perrault, Justin R.
    Arendt, Michael D.
    Schwenter, Jeffrey A.
    Byrd, Julia L.
    Tuxbury, Kathryn A.
    Stacy, Nicole I.
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2021, 33 (03) : 595 - 599