DOSIMETRY OF A SMALL AIR CAVITY FOR CLINICAL ELECTRON BEAMS: A MONTE CARLO STUDY

被引:1
|
作者
Chow, James C. L.
Grigorov, Grigor N.
机构
[1] Princess Margaret Hosp, Radiat Med Program, Toronto, ON M4X 1K9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON, Canada
[3] Univ Waterloo, Dept Phys, Waterloo, ON N2L 3G1, Canada
[4] Ryerson Univ, Dept Phys, Toronto, ON, Canada
[5] Grand River Reg Canc Ctr, Dept Med Phys, Kitchener, ON, Canada
关键词
Monte Carlo simulation; Electron radiotherapy; Dosimetry; Inhomogeneity; Air cavity; LATERAL BUILDUP RATIO; DOSE DISTRIBUTIONS; CHALLENGING COMPLICATION; ORAL MUCOSITIS; NECK TUMORS; RADIOTHERAPY; HEAD; SIMULATION; INTERFACE; ALGORITHM;
D O I
10.1016/j.meddos.2009.03.004
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
This study investigated dosimetric changes in a water phantom when a small air cavity was presented at the central axis of a clinical electron beam. We used 6-, 9-, and 16-MeV electron beams with a 10 x 10 cm(2) applicator and cutout produced by a Varian 21 EX linear accelerator. Percentage depth doses (PDDs) for different depths (0.5-7 cm), thicknesses (2-10 mm), and widths (1-5 cm) of air cavities were calculated using Monte Carlo simulations (EGSnrc code) validated by film measurements. By comparing PDDs of phantoms with and without the air cavity, it was found that when the depth or thickness of cavity was changed, the PDD curve below the cavity was shifted with a distance equal to the thickness of the cavity. However, when the width of the air cavity was changed, both the PDD curve and its slope within and below the cavity were changed. A larger width of the air cavity resulted in a shallower PDD curve within the cavity. The slope of the PDD curve below the cavity tended towards a value as the width of the air cavity was increased to 3-5 cm for the 6-, 9-, and 16-MeV electron beams. The dependence of the depth dose on the width of the air cavity is a result of the contribution of the electron side scattering in the water surrounding the cavity. The change in depth dose resulting from the presence of an air cavity can cause discrepancies between the calculated and actual dose during radiotherapy, unless the effects of the air cavity are properly characterized during treatment planning. From the dosimetry data in this study, neglecting an air cavity of 1-cm thickness in the build-up region of a 6-MeV electron beam resulted in a delivered dose 10-12% larger than the original prescription. Delivered doses 3% and 6% higher than the prescribed dose were observed when doses were prescribed at Rim for a 16-MeV electron beam. These results were obtained by neglecting air cavities with thicknesses equal to 2 and 4 mm, respectively, at a depth of 5 cm. (C) 2010 American Association of Medical Dosimetrists.
引用
下载
收藏
页码:92 / 100
页数:9
相关论文
共 50 条
  • [1] Dosimetric Effect of a Small Air Cavity for Clinical Electron Beams: A Monte Carlo Study
    Chow, J.
    Grigorov, G.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [2] Monte Carlo physical dosimetry for small photon beams
    Perucha, M
    Sánchez-Doblado, F
    Rincón, M
    Leal, A
    Núnez, L
    Arráns, R
    Carrasco, E
    Sánchez-Calzado, JA
    Errazquin, L
    ADVANCED MONTE CARLO FOR RADIATION PHYSICS, PARTICLE TRANSPORT SIMULATION AND APPLICATIONS, 2001, : 555 - 559
  • [3] A Monte Carlo approach for small electron beam dosimetry
    Rincón, M
    Sánchez-Doblado, F
    Perucha, M
    Leal, A
    Arráns, R
    Carrasco, E
    Sánchez-Calzado, JA
    Errazquin, L
    RADIOTHERAPY AND ONCOLOGY, 2001, 58 (02) : 179 - 185
  • [4] Evaluation of air cavity effect for electron beams dose calculations using Monte Carlo method
    Moreno, M. Zarza
    Leal, A.
    Teixeira, N.
    Jesus, A. P.
    Mora, G.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S226 - S226
  • [5] Monte-Carlo-Based Cavity Correction Factors for Ion Chambers in Clinical Electron Beams
    Ono, T.
    Araki, F.
    Yoshiyama, F.
    MEDICAL PHYSICS, 2010, 37 (06) : 3281 - +
  • [6] Dosimetry of Electron Beams at Extremely Extended SSD: Monte Carlo Calculations and Measurements
    Xu, W.
    Zhang, R.
    Xu, M.
    Roeske, J.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Monte Carlo Dose Calculation of Small Field Electron Beams
    Wu, Q.
    Rodrigues, A.
    Sawkey, D.
    Yin, F.
    MEDICAL PHYSICS, 2014, 41 (06) : 307 - 307
  • [8] Monte Carlo study of correction factors for the use of plastic phantoms in clinical electron dosimetry
    Araki, Fujio
    MEDICAL PHYSICS, 2007, 34 (11) : 4368 - 4377
  • [9] Mucosal dosimetry on unflattened photon beams: a Monte Carlo phantom study
    Chow, James C. L.
    Owrangi, Amir M.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (01):
  • [10] Experimental and Monte Carlo investigations of the energy dependence of alanine/EPR dosimetry system for clinical electron and photon beams
    Zeng, GG
    McEwen, MR
    Rogers, DWO
    Klassen, NV
    MEDICAL PHYSICS, 2003, 30 (06) : 1426 - 1426