Evaluation of the Abbott Cell Dyn 4000 automated fluorescent reticulocyte measurements:: comparison with manual, FACScan and Sysmex R1000 methods

被引:18
|
作者
d'Onofrio, G
Kim, YR
Schulze, S
Lorentz, T
Dörner, K
Goossens, W
Zini, G
Tommasi, M
Kendall, R
Scott, CS
机构
[1] Abbott Diagnost GmbH, D-65205 Wiesbaden, Germany
[2] Catholic Univ Hosp, Haematol Serv, Rome, Italy
[3] Abbott Diagnost, Santa Clara, CA USA
[4] Stadt Krankenhaus Zentrallab, Kiel, Germany
[5] UZ Gasthuisberg, Dept Haematol, Louvain, Belgium
来源
CLINICAL AND LABORATORY HAEMATOLOGY | 1997年 / 19卷 / 04期
关键词
reticulocytes; fluorescence; methods; CD4000;
D O I
10.1046/j.1365-2257.1997.00078.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Abbott Cell Dyn 4000 (CD4000) is the first haematology analyser in which fully-automated reticulocyte measurements can be routinely determined by fluorescence as part of the full blood count. This communication reports the first evaluation of this method which was undertaken by three independent reference laboratories in Belgium, Germany and Italy. A total of 695 different samples was entered into the study which was designed to compare CD4000 reticulocyte information (enumeration and qualitative maturational data) with results determined in parallel with the existing manual (supravital staining) reference procedure, and two semi-automated fluorescent assays (Becton Dickinson FACScan and Sysmex R1000 instruments). These studies revealed good agreement between the CD4000 and the manual procedure, with no inter-method bias. Comparison of CD4000 and FACScan reticulocyte measurements, however, indicated a distinct tendency for the FACScan to give higher reticulocyte estimates than the CD4000. Finally, the comparison of the CD4000 with the Sysmex. R1000 showed excellent agreement in the range 0-6% reticulocytes, although there was some inter-method bias in the higher range (> 15%). Analysis of agreement levels between the methods using specific 'clinical decision points' confirmed the tendency for overestimation by the FACScan, in that 58% of the samples with a reticulocytopenia of <0.5% as defined by the CD4000 gave FACScan results within the normal range (0.5-1.8%). In contrast, there was absolute agreement between the CD4000 and the Sysmex R1000 for all reticulocytopenias. Comparison (195 samples) of instrument fluorescent reticulocyte maturation profiles demonstrated an exponential relationship (r = 0.78) between CD4000 IRF and R1000 HFR (highly fluorescent reticulocyte fraction) values. The suggestion that the CD4000 IRF values includes some of the MFR as well the HFR reticulocyte fraction was confirmed as the correlation between the CD4000 IRF and the Sysmex R1000 MFR plus HFR percentages was linear (r = 0.82). This study confirms a high performance level for the CD4000 automated fluorescent reticulocyte method.
引用
收藏
页码:253 / 260
页数:8
相关论文
共 19 条
  • [1] Automated reticulocyte analysis with immature reticulocyte fraction (IRF): Comparison of Abbott Cell Dyn 4000, Sysmex R-3000, two thiazole orange flow cytometric, and manual visual methods.
    Davis, BH
    Bigelow, NC
    VanHove, L
    BLOOD, 1996, 88 (10) : 2767 - 2767
  • [2] WBC differentials, flagging performance and reticulocyte quantification on the Abbott Cell-DYN 4000 in comparison to the Coulter STKS and the reticulocyte analyzer sysmex R-1000.
    Luthe, H
    Binder, L
    Martetschläger, I
    Oellerich, M
    CLINICAL CHEMISTRY, 1999, 45 (06) : A62 - A62
  • [3] EVALUATION OF THE SYSMEX R-1000 - AN AUTOMATED RETICULOCYTE ANALYZER
    TICHELLI, A
    GRATWOHL, A
    DRIESSEN, A
    MATHYS, S
    PFEFFERKORN, E
    REGENASS, A
    SCHUMACHER, P
    STEBLER, C
    WERNLI, M
    NISSEN, C
    SPECK, B
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1990, 93 (01) : 70 - 78
  • [4] Automated reticulocyte counting - an evaluation of GEN-S, Cell-Dyn 3500 and Cell-Dyn 4000.
    Yu, PH
    So, CC
    Wong, KF
    Lee, KC
    Chow, CS
    Yip, LK
    Ho, HK
    Suen, M
    CLINICAL AND LABORATORY HAEMATOLOGY, 1999, 21 (02): : 145 - 147
  • [5] Laboratory evaluation of the miles H center dot 3 automated reticulocyte counter - A comparative study with manual reference method and Sysmex R-1000
    Buttarello, M
    Bulian, P
    Venudo, A
    Rizzotti, P
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 1995, 119 (12) : 1141 - 1148
  • [6] Evaluation of reticulocyte count with Cell-Dyn 3500(R): Comparison with a flow cytometry technique
    Alvarez, C
    Chabert, C
    Quillet, P
    BaufineDucrocq, H
    ANNALES DE BIOLOGIE CLINIQUE, 1997, 55 (03) : 215 - 221
  • [7] Manual and automated methods for the determination of leukocyte counts at extreme low levels: Comparative evaluation of the Nageotte chamber and the Abbott cell dyn 3500 analyser
    Muller, TH
    Doscher, A
    Schunter, F
    Scott, CS
    TRANSFUSION SCIENCE, 1997, 18 (04): : 505 - 515
  • [8] Automated reticulocyte counting and immature reticulocyte fraction measurement - Comparison of ABX PENTRA 120 Retic, Sysmex R-2000, flow cytometry, and manual counts
    Lacombe, F
    Lacoste, L
    Vial, JP
    Briais, A
    Reiffers, J
    Boisseau, MR
    Bernard, P
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1999, 112 (05) : 677 - 686
  • [9] Comparison of the reticulocyte mode of the Abx Pentra 120 Retic, Coulter® General-S™, Sysmex® SE 9500, Abbott CD 4000 and Bayer Advia® 120 haematology analysers in a simultaneous evaluation
    Van den Bossche, J
    Devreese, K
    Malfait, R
    Van de Vyvere, M
    De Schouwer, P
    CLINICAL AND LABORATORY HAEMATOLOGY, 2001, 23 (06): : 355 - 360
  • [10] Comparison of Manual and Automated Grading Methods for Corneal Endothelial Cell Density Measurements by Specular Microscopy
    Modak, Cristina
    Huang, Jianyan
    Maram, Jyotsna
    Marion, Ken
    Obregon, Luisa
    Niemeyer, Moritz
    Sadda, Srinivas R.
    Chopra, Vikas
    Lee, Olivia L.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)